Egor Berezko1, Georgiy Solomakha1, Klaus Scheffler2,3, and Nikolai Avdievich1
1High-field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 2Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 3Department for Biomedical Magnetic Resonance, University of Tübingen, Tübingen, Germany
Synopsis
Keywords: High-Field MRI, Safety
Home-built RF UHF MRI coils have to be carefully evaluated for
safety issues. This is performed by simulating the coil model, which often
include thin conductors, loaded by a human voxel model. Currently such models
are available mainly for time-domain solvers. Fine mesh over conductors and the
large voxel model may lead to time-consuming simulations. Therefore, for
smaller head-sized coils, the model has to be cut and the proper mesh size
chosen. In this work,
we numerically investigated the dependence of B1+ and SAR
for four different head coils on fine local meshing and the size of the voxel
model.
Introduction
Ultra-high field (UHF, 7T and above) MR imaging is an
extensively developing field. Recently two commercial 7T MRI scanner were
cleared for limited clinical use. For imaging at such high field strengths,
strong inhomogeneity occurs due to the fact of the wavelength shortening inside
the human body. This also greatly increases local tissue heating usually
evaluated by calculating a specific absorption rate (SAR). Therefore, home-built
RF coils are always to be carefully verified for safety issues 1. This is
commonly done by simulating the accurate model of the coil loaded by a
realistic human model. Currently, such models are available mainly for
time-domain solvers. Since experimental verification of simulated SAR is not
feasible, the modeling must be carefully validated, which also includes
the evaluation of all potential sources of errors.
Commonly elements of RF coils (arrays), i.e. loops, dipoles, are
constructed using thin conductors, e.g. strips or wires, (1.5-mm wire 2,3), one needs to
apply a high density of computational mesh to acquire correct port impedances,
current density, and consequently correct field distribution. In addition, the
presence of the full-body voxel model makes the total simulation model very
large. These may lead to unreasonably time-consuming simulations. Therefore,
the choice of the mesh size and proper cutting of the voxel models for
simulations of smaller RF coils, e.g. head size, become an important problem.
In this work, we numerically evaluated the dependences of
impedances, RF field, and peak SAR for several RF array models on the property
of the local mesh and voxel model size.Methods
Firstly,
we evaluated the effect of the amount of mesh cells per cross-section of a coil’s
conductor on Z-parameters (i.e. coil impedance). For that, a pair of 21-cm straight
dipoles loaded by a homogeneous (εr=58.3, σ=0.64 S/m 3)
cylindrical phantom of 170-mm in diameter and 200-mm in length was simulated
with CST Studio Suite 2021, using time-domain solver. The dipoles were
separated by 45-(Fig.1A) and 90-degree angles and computed with different local
mesh densities: 2, 4, 8, 10, 12, and 14 mesh cells per cross-section. Then, we
simulated an 8-element 400-MHz folded-end dipoles array 1 using four sizes of
local mesh cells, i.e. 0.5, 0.7, 1, and
2-mm. In these simulations, we evaluated pSAR10g (peak local SAR
averaged over 10g of tissue) and B1+ for the most common
two CP-modes, excited by the progressive phase shift feed of -45 (CP1) and -90
(CP2) degrees.
Finally, we evaluated the effect of the voxel model size on SAR
and B1+ values. In these simulations, we used the “Duke”
voxel model cut at three different lengths, i.e. short, long, and extra-long
(Fig.2). pSAR10g and ‹B1+› (averaged over
13-cm transversal slab) were numerically calculated for four eight-element
arrays shown in Fig.2, i.e. 300-MHz folded-end dipole array 2, 400-MHz folded-end
dipole array 4, and two (300 and 400-MHz) almost identical in geometry loop
arrays 5.Results
It is
shown in Fig.1 (B, C), that the resonant frequency of the same equally tuned dipoles
shows a shift, if one dipole is placed on the axis, and the second one is
rotated by 45 degrees. This effect could be compared with Fig.1 (D), where the
dipoles are placed directly on the x and y axes with the subsequent global mesh
refinement.
Next, different local
mesh sizes for the shortest voxel model were computed 0.5, 0.7, 1, and 2-mm.
Table 1 shows a small deviation between the conventional local voxel size of 0.7-mm. Table 1 also shows no specific correlation between the total amount of mesh
cells, which depends mostly on the voxel model’s length. Simulated B1+
for both CP-modes and every model shown in Fig.3 (a-h). Fig.4 (a-h) depicts
SAR maps, excited by the CP-modes.Discussion and conclusion
Simulation
results of two dipole and phantom models using a time-domain solver (Fig. 1 B,
C) shows a stable shift of the dipole's impedance. However, since each coil
will be tuned and matched, such impedance changes are not critical but can lead
to an unrequired phase shift for the signal. The pSAR evaluated for different
local mesh properties deviates weakly even with the voxel model's wrong size
(i.e., cropped at the neck). ‹B1+› deviates less than 1%
for each local mesh size.
In this
work, we demonstrated numerically how simulation parameters change with the
voxel model size and local mesh properties. Despite a slight variation in
dipole impedance with local mesh properties discovered, the mesh or voxel model
size did not significantly change ‹B1+› and SAR estimates.
pSAR was expected to increase with the shortening of absorbing tissue due to
the higher power deposition, but simulations did not prove it. Furthermore, the
pSAR simulation results appeared to be stable enough.Acknowledgements
No acknowledgement found.References
1. Hoffmann
J, Henning A, Giapitzakis IA, Scheffler K, Shajan G, Pohmann R, Avdievich NI.
Safety testing and operational procedures for self-developed radiofrequency
coils. NMR Biomed 2016;29(9):1131-1144.
2.
Avdievich, N.I., Solomakha, G., Ruhm, L., Nikulin, A.V., Magill, A.W. and
Scheffler, K., 2021. Folded‐end dipole transceiver array for human whole‐brain
imaging at 7 T. NMR in Biomedicine, 34(8), p.e4541.’
3. Shajan G, Kozlov M, Hoffmann J, Turner R, Scheffler K,
and Pohmann R. A 16-channel dual-row transmit array
in combination with a 31-element receive array for human brain imaging at 9.4
T. Magn Reson Med 2014;71:870–879.
4. Avdievich
NI, Solomakha G, Ruhm L, Henning A, Scheffler K. Unshielded bent folded-end
dipole 9.4 T human head transceiver array decoupled using modified passive
dipoles. Magn Reson Med 2021;86(1):581-597.
5. Avdievich
NI, Giapitzakis IA, Pfrommer A, Henning A. Decoupling of a tight-fit
transceiver phased array for human brain imaging at 9.4T: Loop overlapping
rediscovered. Magn Reson Med 2018;79(2):1200-1211.