Ehsan Kazemivalipour1,2, Markus W. May3,4, Jason P. Stockmann1,2, Robert L. Barry1,2, Boris Keil5,6, Lawrence L. Wald1,2,7, and Bastien Guerin1,2
1A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Erwin L. Hahn Institute for MRI, University Duisburg-Essen, Essen, Germany, 4High-Field and Hybrid MR Imaging, University Hospital Essen, Essen, Germany, 5Institute of Medical Physics and Radiation Protection, Department of Life Science Engineering, Mittelhessen University of Applied Sciences, Giessen, Germany, 6Center for Mind, Brain and Behavior (CMBB), Philipps-University Marburg, Marburg, Germany, 7Harvard-MIT Division of Health Sciences Technology, Cambridge, MA, United States
Synopsis
Keywords: RF Pulse Design & Fields, RF Pulse Design & Fields
We
introduce a VOP compression scheme using “stacked
E-fields” followed by the
removal of linear redundancies with a positive semi-definite convex
optimization problem. This method provides a lower VOP-count at a constant SAR
overestimation factor compared to three other approaches when applied to 7
Tesla 8-channel and 16-channel arrays loaded with three body models, each
placed at three z-positions. The “stacked
E-field” method yielded a compression
factor of 0.0124% and 0.0039% for the 8-channel and 16-channel
arrays, respectively, and 61% and 6% fewer VOPs than the standard VOP approach
that does not remove linear redundancies.
Introduction
Parallel
transmission (pTx) creates complex E-field interferences between transmit
channels that require monitoring in real-time during the MRI sequence. This is
typically done by simulation of multiple body models and positions and
compression of the resulting Q-matrices1 using the virtual observation
points (VOPs) algorithm2. Despite relatively high compression factors, VOP sets
routinely have hundreds of control matrices which can prove challenging for
rapid (millisecond timescale) specific absorption rate (SAR) monitoring. Similarly,
patient-specific pulse optimization including SAR metrics requires a reduced
VOP count while keeping the SAR overestimation factor as small as possible. In
this study, we compared four VOP compression approaches that each produce
different numbers of VOPs for the same SAR overestimation factor and describe
an optimal method.Methods
We simulated a 7T 8-channel helmet pTx head array3 and a 16-channel pTx head/neck array4 (Figure 1).
Each coil was loaded with three body models (MGH model5,6, VHP male7, and VHP female8), which all contain significant cerebrospinal
fluid compartments. Numerical electromagnetic (EM) simulations were performed
with ANSYS Electronics (ANSYS Inc., Canonsburg, PA). The co-simulation approach
was employed for efficient tuning, matching and decoupling (matching <-20dB
and coupling <-10dB). Each body model was modeled at z=0 (eyes at iso-center),
z=-10mm,
and z=+10mm, to provide robustness of the virtual observation point set with
respect to subject position and motion.
The E-fields and B1+-maps of each channel
were exported on 2-mm isotropic grids covering the head, neck, and shoulders. 10g-averaged
SAR Q-matrices were computed
and compressed using the VOP algorithm2 with a maximum SAR
overestimation factor of 5×10-4 W/kg/V2 for the
8-channel array and 7.5×10-4 W/kg/V2 for the 16-channel
array.
We compare four VOP approaches: 1) computation of VOPs for
each body model/position followed by concatenation of the VOP files, 2)
concatenation of E-fields for all body models/positions, followed by VOP
compression. Methods 3) and 4) are the same as 1) and 2), but are followed by a
“generalized VOP” (gVOP) step proposed by Lee et al9 which is based on the
observation that the original VOP approach creates VOPs dominated (in the semi-definite
positive sense) by linear combinations of all other VOP matrices in the set.
For those VOPs (Sj), it is possible to find real numbers λ1,…,λN,
such that
$$\begin{array}{*{20}{c}}{{\rm{gVOP\; \; condition}}:}&{\begin{array}{*{20}{c}}{\sum\limits_{i =
1,..,\,N\atop\,\,\,\,i\ne j}{{\lambda_i}{{\bf{S}}_i}} \succ {{\bf{S}}_j}{\rm{ }}}\\{where\,\sum\limits_{i
= 1,..,\,N\atop\,\,\,\,i\ne j}{{\lambda_i}}=1+\varepsilon}\end{array}}&{}&{[1],}\end{array}$$
where ε is the maximum SAR overestimation
factor expressed as a percentage of the worst possible local-SAR for a unit
pulse. Matrices Sj satisfying condition [1] should
be removed from the set. We used the semi-definite MATLAB program SeDuMi10 interfaced with YALMIP11 to solve the gVOP condition.Results
Figures 2A-C show
the number of VOPs for all coils, body models and positions (method#1), as well
the number of VOPs obtained by concatenation of E-fields for all body model
positions, followed by VOP compression (method#2). Method#1 yielded 4337 and 1933
VOPs for the 8-channel and 16-channel coils, respectively, whereas method#2
yielded 1292 and 817 VOPs for those coils; a 70% and 58% reduction in the
number of VOPs for the same SAR overestimation factor. Therefore, the VOP
algorithm should always be run on concatenated E-fields, and not on individual model
E-fields followed by concatenation of the resulting VOP sets. Figures 2B-D show
that the maximum eigenvalue (worst-case local-SAR) of individual
models/positions (method#1) is equal to that of the concatenated E-field + VOP (method#2),
showing that, as expected, the VOP algorithm captures worst-case SAR
combinations without information loss across body models and positions.
Figures 3A-4A show the number of VOPs associated methods#1-4
for the 8-channel and 16-channel arrays. For both arrays, method#4 (“concatenated
E-field + VOPs + gVOP”) yielded the smallest number of the VOPs at constant SAR
overestimation. The impact of the gVOP step was much more pronounced for the 8-channel
than the 16-channel array, although the reason for that is still under
investigation. Figures 3B-4B show that all VOP sets yields SAR overestimations
within the prescribed limit, which validates the different methods (they only
differ by their compression ratio).
Figure 5 shows that the impact of the gVOP step
increases at low SAR overestimation factors. This signifies that VOP sets with
stringent epsilon SAR overestimation values have a greater tendency to create
redundant control matrices, and therefore benefit more from VOP pruning by
solving Eq. [1]. This phenomenon was much more pronounced for the 8-channel
than the 16-channel array.Discussion
MRI
manufacturers limit the size of VOP files used by the safety watchdog to
guarantee that SAR monitoring is fast enough12. There is a tradeoff between the
number of VOPs and SAR estimation accuracy: small files may lead to large SAR
overestimation, which is undesirable as this yields SAR-inefficient RF pulses
and protocols. We found that the best compression strategy applies the VOP
compression after concatenation of the E-fields of different body models/positions.
Even then, however, a significant number of linear redundancies remain in the
set, which we were able to remove by solving the generalized VOP condition [1].
Doing so further reduced the VOP sets by 61% for the 8-channel coil and 6% for
the 16-channel coil while keeping the SAR overestimation factor constant.Acknowledgements
No acknowledgement found.References
1. Guerin B, Gebhardt M, Cauley S,
et al. Local specific absorption rate (SAR), global SAR, transmitter power, and
excitation accuracy trade-offs in low flip-angle parallel transmit pulse
design. Magn Reson Med. 2014;71(4):1446-1457.
2. Eichfelder G,
Gebhardt M. Local specific absorption rate control for parallel transmission by
virtual observation points. Magn Reson Med. 2011;66(5):1468-1476.
3. Kazemivalipour E,
Wald LL, Guerin B. Comparison of simulated parallel transmit head arrays at 7T
using excitation uniformity, local SAR, and global SAR. Proc Intl Soc Mag Reson
Med 30 2022:5048.
4. May MW, Hansen SJD,
Mahmutovic M, et al. A patient-friendly 16-channel transmit/64-channel receive
coil array for combined head-neck MRI at 7 Tesla. Magn Reson Med.
2022;88(3):1419-1433.
5. Massire A, Cloos MA,
Luong M, et al. Thermal simulations in the human head for high field MRI using
parallel transmission. J Magn Reson Imaging. 2012;35(6):1312-1321.
6. Makris N, Angelone L,
Tulloch S, et al. MRI-based anatomical model of the human head for specific
absorption rate mapping. Med Biol Eng Comput. 2008;46(12):1239-1251.
7. Noetscher GM. The
CAD-Compatible VHP-Male Computational Phantom. In Makarov SN, Noetscher GM,
Nummenmaa A, (Eds). Brain and Human Body Modeling 2020: Computational Human
Models Presented at EMBC 2019 and the BRAIN Initiative(R) 2019 Meeting. Cham
(CH) 2021:309-323.
8. Makarov SN, Noetscher
GM, Yanamadala J, et al. Virtual Human Models for Electromagnetic Studies and
Their Applications. IEEE Rev Biomed Eng. 2017;10:95-121.
9. Lee J, Gebhardt M,
Wald LL, et al. Local SAR in parallel transmission pulse design. Magn Reson
Med. 2012;67(6):1566-1578.
10. Sturm JF. Using
SeDuMi 1.02, A Matlab toolbox for optimization over symmetric cones.
Optimization Methods and Software. 1999:625-653.
11. Lofberg J. Yalmip :
a toolbox for modeling and optimization in MATLAB. In: . IEEE International
Conference on Robotics and Automation 2004:284–289.
12. Jin J, Weber E, Destruel A, et al. An open
8-channel parallel transmission coil for static and dynamic 7T MRI of the knee
and ankle joints at multiple postures. Magn Reson Med. 2018;79(3):1804-1816.