Maxim Terekhov1, Ibrahim A. Elabyad1, David Lohr1, and Laura M. Schreiber1
1Chair of Molecular and Cellular Imaging, Comprehensive Heart Failure Center, University Hospital Würzburg, Wuerzburg, Germany
Synopsis
Keywords: RF Arrays & Systems, Safety
The incomplete consensus regarding SAR safety in ultra-high-field
MRI leads to a large variety of
exclusion criteria for volunteer cohorts in the different centers, especially regarding
the acceptance of conductive implants. Therefore, SAR analysis of new
commercial transceiver arrays contributes to supplying local IRBs with information
regarding the safety of human subjects. This study aimed to analyze SAR for a
prototype of a new 8Tx/16Rx array for 7T cardiac MRI. The main goal was to
analyze the array safety for human subjects with a different type of passive
implant located outside of the array borders.
Introduction
MRI at ultra-high
B0 field (UHF) is an emerging technique to improve the
signal-to-noise ratio (SNR) compared to clinical MRI scanners (1.5-3T).
However, the application of UHF MRI faces B1+ homogeneity
problems, especially when applied in the thorax and abdominal body regions.
Parallel transmit (pTx) technology has been implemented to overcome this
problem. However, pTX RF excitation turns the analysis of the electromagnetic
(EM) energy deposition in the body (required for subject safety) into a complex
problem. So far, the overall problems concerning SAR safety (including
simulations, soft- and hardware control, and efficient procedures of finding
the appropriate B1+-shimming) did not allow vendors to approach the FDA/CE
for certification of 7T MRI of the thorax, and abdomen, or pelvis using transceiver
arrays. All UHF MRI studies with such arrays, in particular for cardiac
applications, need to be approved by an institutional review board (IRB). Passive
conductive implants are usually listed as exclusion criteria due to the tissue
heating following RF-array irradiation, which limits the number of potentially
suitable subjects for UHF MRI to those without an implant, which makes
difficult studies in large cohorts [1]. The incomplete consensus regarding safety leads to a
large variety of exclusion criteria for volunteer cohorts in the different
centers, especially regarding the acceptance of conductive implants that are remote
from the array of conductors. Additional analysis on each new commercial
transceiver array is an important contribution, supplying local IRBs with
reliable information to allow for informed decisions regarding the safety of human
subjects. Therefore, this study aimed to make a consideration of the SAR safety
for a prototype of a new 8Tx/16Rx array for 7T cardiac MRI with elements
geometry implemented later in a commercial array[2]. The main goal was to analyze the array safety when
used in human subjects with a different type of passive implant located outside
of the array borders. Method
For SAR safety analysis the EM-simulations were performed with CST-Microwave-Studio (CST-MWS,
Dassault Systeme) using “Duke” and “Ella” human models (ITI’S Foundation). The
local averaged 10g SAR was evaluated using the IEEE/IEC-62704-1. The 16 loop
elements of the array were excited with the same normalized input accepted
power (Pin =1W) using
three optimized static phase-only shim settings computed in the work [[3]].
The simulations of peak SAR were made for (1) a dental
retainer (DR) 75mm in length and placed in the lower jaw, (2) a T-shape spiral
intra-uterine device (IUD), and (3) a hip joint implant (HJI). The distance
from the nearest array edge for the IUD and a hip joint implant is 24 cm and 28
cm respectively. The IUD and HJI were simulated in the “Ella”
model and a DR in both models. For all simulations, the array was positioned
such that the centers of both anterior and posterior parts (Figure 1a) were
located on the body's vertical axes and matched the center of the heart. Additionally,
for the Duke model with the allocated DR, the anterior array was simulated
shifted to the right-and-head direction (on 50mm), and straight to the head
(on 75mm) as shown on Figure 1(b). Results
Figure 2 presents the simulation results for SAR in the presence of a dental
retainer in the bottom jaw of the “Duke” model. The peak SAR at 1 W
accepted power in the region of the DR does not exceed 0.005 W/kg, whereas
as expected the peak SAR under the array is higher by up to two orders of
magnitude. The results show that even a 75 mm shift towards the head keeps
the SAR in the region of the retainer on average by one order of magnitude
lower than the maximal SAR within the array conductor perimeter. Figure 2b with
results for diagonal shift confirms this observation. The maximal predicted SAR
in the tissue surrounding the retainer does not exceed 10% of the peak value
predicted within the array.
Figure 3 (a) shows the position of the implants and the results of the SAR
simulations in the “Ella” model. The SAR decreases in all the directions
in-plane with the array. The peak SAR decreases at least by one order of
magnitude within 10 cm from the array perimeter (b,c) Beyond this
distance, at 15-20 cm the peak local SAR value is 2..3 orders of magnitude
smaller even in the presence of a massive conductive body such as a
stainless-steel HJI.Discussion
The simulation results confirm that the local SAR
values in all the directions co-planar to the array plane and are negligible
compared to the peak values predicted within the array borders. In our opinion,
these results can be generalized for considerations regarding SAR safety and
massive conductive implants located outside of the array’s perimeter. Conclusion
Based on our results we suggest, for the analyzed
array geometry, that the presence of passive electrically conductive implants
located beyond 10 cm from the conductor perimeter of the array should not lead
to detectable RF-induced tissue heating around the implants if their size and
geometry are similar to an IUD or a dental implant. The subjects with such
implants may be included in 7T cardiac MRI studies using the considered array
type. Acknowledgements
Part of the study obtained financial support
from the German Ministry of Education and Research (BMBF) under grants #
01EO1004 & 01EO1504. L.M. Schreiber
receives research support from Siemens Healthineers. The position of D. Lohr is partially paid for that research
support. We thank Dr. Titus Lanz and
Carsten Kögler ( Rapid Biomedical) , Dr. Robin Heidemann and Dr. Adriane
Groeger (Siemens Healthineers) for the encouraging and insightful discussions. References
1. Reiter, T., et al., On the way to routine cardiac MRI at 7 Tesla
- a pilot study on consecutive 84 examinations. PLoS One, 2021. 16(7): p. e0252797.
2. Terekhov, M.L., D. Reiter, T,
Elabyad, I.A., Hock, M., Schreiber, L.M. New
commercial 8Tx/16RX array for Clinical
Application in 7T Cardiac MRI: initial experience in Healthy Volunteers. in ISMRM Annual Meeting. 2021. Virtual
Meeting.
3. Terekhov, M.,
Elabyad, IA, Resmer, F., Lanz, T., Reiter, T., Lohr, D., Schlöttelburg,
W., Schreiber, L.M. Customized B1+-Shaping using Multi-Channel Transceiver Array Prototype
for 7T Cardiac MRI with Central Elements Symmetry. in International Society for
Magnetic Resonance in Medicine Annual Meeting. 2020. Virtual
Meeting.