Vincent Gras1, Edouard Chazel1, Nicolas Boulant1, Alexis Amadon1, and Michel Luong1,2
1CEA, CNRS, BAOBAB, Neurospin, University of Paris Saclay, Gif sur Yvette, France, 2IRFU, Paris Saclay University, Gif-sur-Yvette, France
Synopsis
Keywords: Safety, Brain Connectivity
Efficient
monitoring of the specific absorption rate (SAR) in parallel-transmission MRI
is performed by evaluating the RF power deposition on specific SAR matrices
called Virtual Observation Points (VOPs). But given the various sources of
uncertainties that occur in practice (head placement, anatomy, precision of the
body model to name a few), a challenging task remains to determine a suitable
safety margin to be applied on the VOPs to offer a tradeoff between safety and scanning performance. In this work, as a first step towards this, we propose a workflow
to tackle this problem for brain imaging at 11.7 Tesla.
Introduction
Parallel
transmission (pTX) is a versatile solution to enable UHF MRI of the human body,
where radiofrequency (RF) field inhomogeneity appears very challenging. Today,
monitoring of the specific absorption rate (SAR) in pTX essentially relies on precise
electromagnetic simulations (SAR model) able to return the local SAR
distribution inside the body in response to any applied pTX RF waveform1-3. SAR monotoring is then performed by evaluating the RF power deposition on specific SAR matrices called Virtual Observation Points (VOPs)3, obtained by compression of the SAR model3-6. In that framework, given the various
sources of uncertainties that occur in practice, a challenging task remains to
determine a suitable safety margin to be applied on the VOPs to offer a tradeoff
between safety and scanning performance7-9. In this work, as a first step
towards this, we propose a workflow to tackle this problem for a 11.7 Tesla head imaging setup.Theory
$$$\newcommand{\Q}{\mathbf{\mathrm{Q}}}\newcommand{\Qset}{\mathcal{Q}}\newcommand{\x}{\mathbf{x}}\newcommand{\r}{\mathbf{r}}\newcommand{\M}{\mathcal{M}}\newcommand{\sar}{\mathrm{SAR}}\newcommand{\sed}{\mathrm{SED}}\newcommand{\C}{\mathrm{R}_{\mathcal{Q}^*}(\mathbf{\mathrm{Q}})}\newcommand{\X}{\mathbf{\mathrm{X}}}\newcommand{\Y}{\mathbf{\mathrm{Y}}}\newcommand{\H}{\mathcal{H}_+^{N_c}}\newcommand{\tr}{\mathrm{trace}}$$$Provided
that the electric field distribution per Watt of forward RF power is known
for every transmitter, the local SAR at a given position in the body can be expressed mathematically as the
evaluation of a complex Hermitian matrix $$$\Q(\r)$$$ on the
vector $$$\x(t)=\left(x_i(t)\right)_{1\le i\le{N_c}}$$$ formed by
the complex RF voltages applied to the channels $$$1{\cdots}N_c$$$ of the TX array, i.e. $$$\sar(\Q(\r),\x(t))=\x(t)^H\Q(\r)\x(t)$$$ ($$$\cdot^H$$$ for the Hermitian conjugate). In what follows, given a body model $$$\M$$$, $$$\Qset_\M:=\left\{\Q(\r);\r\in\M\right\}$$$ is called the SAR model on $$$\M$$$.
An unconditionally safe VOP model is a set of positive definite matrices $$$\Qset^*$$$ that guarantees that for any model $$$\M$$$ and for any pTX RF waveform $$$\x(t)$$$:$$\forall\Q\in\Qset_\M,\exists\Q^*\in\Qset^*\:\mathrm{such\:that}:\:\forall t;\sed(\Q,\x,t)\le\sed(\Q^*,\x,t)$$where:
$$\sed(\Q,\x,t)=\int_{0}^{t}\sar(\Q,\x(t))dt'\label{cond_sed}$$
This condition implies that for any RF shim $$$\x$$$:
$$\exists\Q^*\in\Qset^*\:\mathrm{such\:that}:\:\sar(\Q,\x)\le\sar(\Q^*,\x)\label{cond_sar}$$
We can show that:
- The above condition is satisfied on $$$\Q$$$ if the so-called safety criterion $$$\C$$$ defined as:
$$\C:={\max}_{\x}\left({\min}_{\Q^*\in\Qset^*}\left(\frac{\x^H\Q\x}{\x^H\Q^*\x}\right)\right)$$
does not exceed 1;
2. the above defined criterion can be computed by convex programming:
$$\C=\max_{\x}\left(\x^H\Q\x\right)\:\mathrm{subject\:to}:\:\x^H\Q^*\x\le1\:\forall\:\Q^*\in\Qset^*$$.
To summarize, given a VOP set $$$\Qset^*$$$, the criterion $$$\C$$$ provides a mathematically proven worst-case across all possible RF shims. Importantly, this worst-case does not necessarily correspond to an eigenvalue of $$$\Q$$$.
Material and methods
Safety criterion maps were computed numerically on a collection of simulation data on a 16TX head coil array model tuned at 500 MHz
10 (Iseult coil for 11.7T brain MRI) (see Figure 1). Two numerical head models (male and female)
comprising nine different tissues were used for this study.
Electromagnetic simulations were performed with HFSS. A set of 16 transformations (see Table
1) was applied to each model in order to mimic a variation of the placement of
the head, its size and its electric properties. From each simulation, the theoretical
10g-average SAR matrix was computed voxel-wise .
A VOP set was computed from the two "nominal" simulations using a VOP compression parameter of 1/kg.
For each simulation, the safety criterion was computed voxel by voxel using a sequential quadratic programing algorithm provided
in the Optimization Toolbox of Matlab. We computed on each R-map:
- The
maximum (Rmax)
- The
mass of tissue in gram where R>1 (m1)
- The
mass of tissue in gram where R>1.5 (m1.5)
Results
After
compression, the number of VOPs was 150. A graphical representation of these
VOP is provided in Figure 2. The R-maps for all 16 transformations and both body models are shown in Figure 3 (sagittal view). Box
plots of the safety criterion across voxels are reported in Figure 4. It appears
that the most influencing transformations are the geometrical scaling S*0.85,
SX0.85, SY0.85 and SY1.15 (Rmax ~ 1.4 or above), the translation
TZ-2cm (Rmax ~ 1.3). The rotations did not lead to a Rmax greater than 1.3.
Finally, the variations in the dielectric properties had the smallest influence
(Rmax < 1.1). A value less than 1 guarantees that a given Q matrix is
upper-bounded by the VOP set. In
Figure 4, we report the Rmax, m1 and the m1.5 metrics for both body models.
We observe
that m1.5 reaches in the worst case 100 gram.Discussion and conclusion
Given the high
dimensionality offered by pTX, testing random excitations hardly approaches the worst-case scenario hinted by computation of the R-map. Interestingly also, Rmax can directly be used as the (worst
case) necessary safety margin, introduced in the literature for the use of VOPs
in practice7-9. The safety criterion method has been applied on two sets of 17
electromagnetic simulations to quantify the effect of experimental uncertainties
(head position, head shape, tissue electric properties) on local SAR monitoring
using VOPs. We found here that a multiplicative safety factor of 1.5 on the VOP
model was likely to be sufficient to confidently operate with the generated VOP
model, a factor of 2 being the absolute worst case in this test study.
This factor is consistent with the safety factor normally applied for 7T brain
MRI9, suggesting that the increase in the Larmor frequency does not lead to a dramatic
increase in the interindividual variability of the SAR as compared to 7T. Naturally,
to gain confidence on the safety factor estimate, it would be beneficial to add
new body models transformation combinations. The mass of tissue being potentially SAR
underestimated with a VOP model (m1) also provides useful information
for risk assessment.
Acknowledgements
No acknowledgement found.References
1.
S.
Wolf, D. Diehl, M. Gebhardt, J. Mallow, et O. Speck, « SAR simulations for
high-field MRI: How much detail, effort, and accuracy is needed? », Magnetic
Resonance in Medicine, vol. 69, no 4, Art. no 4, 2013, doi: 10.1002/mrm.24329.
2.
Graesslin,
I. et al. (2012) ‘A specific absorption rate prediction concept for parallel
transmission MR’, Magn Reson Med, 68(5), pp. 1664–1674. doi:10.1002/mrm.24138.
3.
Eichfelder,
G. and Gebhardt, M. (2011) ‘Local specific absorption rate control for parallel
transmission by virtual observation points’, Magn Reson Med, 66(5), pp. 1468–1476.
doi:10.1002/mrm.22927.
4.
J.
Lee, M. Gebhardt, L. L. Wald, et E. Adalsteinsson, « Local SAR in parallel
transmission pulse design », Magn Reson Med, vol. 67, no 6, Art. no 6, 2012,
doi: 10.1002/mrm.23140.
5.
Kuehne,
A. et al (2017). ‘Massively accelerated VOP compression for population-scale RF
safety models’, In proceedings of the 25th ISMRM conference, Honolulu, HI,
United States. Available:
https://cds.ismrm.org/protected/17MPresentations/abstracts/0478.html.
6.
S.
Orzada, T. M. Fiedler, H. H. Quick, et M. E. Ladd, « Local SAR compression
algorithm with improved compression, speed, and flexibility », Magn. Reson.
Med., vol. 86, no 1, p. 561‑568, juill. 2021, doi:
10.1002/mrm.28739.
7.
M.
de Greef, O. Ipek, A. J. E. Raaijmakers, J. Crezee, et C. A. T. van den Berg, «
Specific absorption rate intersubject variability in 7T parallel transmit MRI
of the head », Magn Reson Med, vol. 69, no 5, Art. no 5, 2013, doi:
10.1002/mrm.24378.
8.
Y.
Shao, P. Zeng, et S. Wang, « Statistical simulation of SAR variability with
geometric and tissue property changes by using the unscented transform:
Statistical SAR », Magnetic Resonance in Medicine, vol. 73, no 6, Art. no 6,
juin 2015, doi: 10.1002/mrm.25367.
9.
M.
Le Garrec, V. Gras, M.-F. Hang, G. Ferrand, M. Luong, et N. Boulant, «
Probabilistic analysis of the specific absorption rate intersubject variability
safety factor in parallel transmission MRI », Magnetic Resonance in Medicine,
sept. 2016, doi: 10.1002/mrm.26468.
10.
Luong,
M. et al. ‘A Compact 16Tx-32Rx Geometrically Decoupled Phased Array for 11.7T
MRI’, In proceedings of the ISMRM-ESMRMB & ISMRT 31st Annual Meeting,
Abstract 707 (June 2022).
https://index.mirasmart.com/ISMRM2022/PDFfiles/0707.html.