Mostafa Berangi1,2,3, Helmar Waiczies2, and Thoralf Niendorf1,2,3
1Berlin Ultrahigh Field Facility (B.U.F.F.), Berlin, Germany, 2MRI.TOOLS GmbH, Berlin, Germany, 3Charité – Universitätsmedizin Berlin, Berlin, Germany
Synopsis
Keywords: Safety, Parallel Transmit & Multiband
Fixation of
bone fractures with screws is common clinical practical including follow-up monitoring
of the implantation site and of the healing process. MRI-based monitoring can
be compromised by tissue heating and transmission field distortion. To overcome
this challenge, this work uses parallel transmission using an optimal
transmission vector. The proposed approach leverages a single EMF simulation
without the implant, while the effect of implant electromagnetic scattered fields are
estimated using a transfer matrix. Optimal shim vectors are computed by an
optimization process using a multilevel genetic algorithm. Our findings
demonstrate the feasibility and performance of the transfer function driven
approach.
Introduction
Bone fracture
fixation using screws is common clinical practice.1 Biodegradable screws2 are currently clinically available, and
they substantially improve patient comfort by eliminating the need for implant
removal surgery. After implantation, the healing process at the implant site,
and the corrosion status of the biodegradable implants are closely monitored3. MRI can be used to study the
implant-tissue interface and the degradation status of biodegradable implants4. However, the electrically conductive
nature of implants presents challenges for MRI, including: 1) MR safety risks due
to concentrated energy deposition in tissue adjacent to the implant 5; 2) B1+ field
distortions resulting from the implant-induced scattered radiofrequency (RF)
field. These constraints can be greatly reduced or eliminated by parallel
transmission of multichannel RF coil arrays. Recognizing this opportunity, recent
reports have demonstrated the feasibility of safe MRI of small implants at 7.0
T.6,7 We propose a generalized approach, to translate
these findings to any patient, with arbitrary implant geometry, location and
positioning, and to any RF coil configuration. This approach leverages a single
EMF simulation without the implant, estimating the implant electromagnetic scattered fields using an implant-specific transfer matrix. Material and method
EMF distribution of an 8-channel
modular loop-dipole RF transceiver8 placed around
a cylindrical phantom (r=85mm; L=300mm) mimicking the electrical
properties of muscle tissues was simulated at 297.2 MHz (Fig.1) (CST Studio
Suite 2022, CST MWS). Next, the transfer matrix9 of a metallic rod (r=2mm, L=70mm)
mimicking an implant was calculated. The transfer matrix defines the current
distribution along the implant resulting from incidence of a localized and purely
tangential electric field along a certain spatial length on the implant. This
localized excitation was simulated in CST by importing the EMFs from two
counter-propagating plane wave in a 100×100×5mm3 domain as source
fields into another simulation with the implant. The current distribution along
the implant in a scenario where the implant is placed at any arbitrary location
and orientation inside the phantom can be estimated in the first step using the
transfer matrix and incident tangential electric field of the RF transceiver.
The B1+,
tangential E-field and specific absorption rate (SAR) derived from the first
step, along with the transfer matrix deduced from the second step were incorporated
into an optimization step using a multi-objective genetic algorithm (GA) that
aims to maximize a uniform B1+ in a cylindrical (r=20mm;
L=110mm) region of interest (ROI), while minimizing SAR and implant-induced
current throughout the simulation domain. The result of the optimization is a
(GA-driven) excitation vector that provides a strong and uniform B1+
in the ROI, while avoiding SAR increase at the implantation site, and
minimizing SAR throughout the phantom. The performance of this optimized
excitation vector was evaluated in another simulation in which an implant was
inserted into the phantom at the exact location for which the optimization was
performed. For benchmarking this excitation vector was compared to the CP mode.Result
An example
current distribution along the implant is shown in Figure2 together with the
corresponding transfer function. B1+ and point SAR maps obtained
for GA-driven and CP excitations derived from the simulations w/wo the implant
are shown in Figure 3. For transmission field shimming using the CP algorithm, a
strong bow-shaped B1+ void was formed close to the
implant. GA transmission field shimming eliminated B1+
artefacts, and hence facilitated the acquisition of uniform images close to the
implant. The SAR reduction of the GA approach over the CP approach is
highlighted in Figure4. Notwithstanding the strong and uniform B1+
field in the implant region, the GA-driven excitation vector derived from the simulations
w/wo the implant in conjunction with the transfer function successfully
eliminated implant SAR. For comparison, CP-based excitation showed substantial
SAR hot spots at the tips of the implant. The performance gain of the transfer function-based
GA approach benefits MRI of the implant site. Discussion
CP excitation induces
B1 artifacts near the implant and excessive local RF power
deposition. These constraints are detrimental, or even prohibitive for MRI of implant
sites. This work demonstrates that pTX, in conjunction with transfer function-driven
genetic algorithms for transmission field shaping, ensures MR safety and
transmission field uniformity suitable for MRI-aided monitoring of tissue
healing of implantation sites, including MRI characterization of the
degradation state of biodegradable orthopedic implants. While the impact of the
RF transmit array on the efficiency of GA-based excitation vector optimization
is acknowledged, it stands to reason that the approach proposed and evaluated here
is compatible with any RF array with an arbitrary number of transmit channels
to facilitate safe and B1+ distortion-free MRI of
implants. The proposed transfer function-driven
approach is not limited to scenarios where there is only one implant in the simulation
domain, but supports more than one implant, if coupling between implants is
negligible. Conclusion
Our findings help
to meet MR safety and transmission field uniformity requirements of MRI-guided
implant site monitoring. Notwithstanding that this proof-of-principle study was
demonstrated at 7.0 T, the proposed
approach provides a technological basis for MRI of orthopedic and other
conducting implants at clinical field strengths.Acknowledgements
This project is
supported by the H. 2020 European Training Network of the European Union
(MgSafe, grant agr. No. 811226)References
1. Slone, R. M., Heare, M. M., vander Griend, R. A. &
Montgomery, W. J. Orthopedic fixation devices. https://doi.org/10.1148/radiographics.11.5.1947319
11, 823–847 (1991).
2. Luthringer, B. J. C., Feyerabend, F. & Römer, R. W.
Magnesium-Based Implants: A Mini-Review. Magnesium Research Preprint at
https://doi.org/10.1684/mrh.2015.0375 (2014).
3. Myrissa, A. et al. In vitro and in vivo comparison
of binary Mg alloys and pure Mg. Materials Science and Engineering C
(2016) doi:10.1016/j.msec.2015.12.064.
4. Paramitha, D. et al. Monitoring degradation products
and metal ions in vivo. Monitoring and Evaluation of Biomaterials and
their Performance In Vivo 19–44 (2017)
doi:10.1016/B978-0-08-100603-0.00002-X.
5. Winter, L., Seifert, F., Zilberti, L. & Murbach, M.
MRI-Related Heating of Implants and Devices : A Review Physics of Implant
Heating. 1–20 (2020) doi:10.1002/jmri.27194.
6. Mostafa Berangi, Andre Kuehne, Helmar Waiczies & and
Thoralf Niendorf. Excitation vector optimization for safe parallel
transmission MRI of passively conducting implants in the presence of motion.
in (2022).
7. Mostafa Berangi, Andre Kuehne & Thoralf Niendorf. Safe
7T MRI of tissues neighboring Mg-based biodegradable implants using parallel
transmission. in (2021).
8. Ertürk, M. A., Raaijmakers, A. J. E., Adriany, G., Uğurbil,
K. & Metzger, G. J. A 16-channel combined loop-dipole transceiver array
for 7 Tesla body MRI. Magn Reson Med (2017) doi:10.1002/mrm.26153.
9. Tokaya, J. P., Raaijmakers, A. J. E., Luijten, P. R.,
Sbrizzi, A. & van den Berg, C. A. T. MRI-based transfer function
determination through the transfer matrix by jointly fitting the incident and
scattered field. Magn Reson Med 83, 1081–1095 (2020).