Di Hu1, Yanqiu Lv1, Dandan Zheng2, Geli Hu2, Peng Sun2, and Yun Peng1
1The Department of Radiology,, Beijing Children's Hospital, Capital Medical University, National Center for Children's Health, Beijing, China, 2Clinical & Technical Support, Philips Healthcare, Beijing, China
Synopsis
Keywords: Data Analysis, Reproductive
We measured imaging reproducibility in
structural, resting state fMRI and diffusion tensor scans across different time
points and scanners in healthy volunteers. First, we assessed
structural imaging variability by calculating volume for seven subcortical
structures. Second, we evaluated across-scanner and across-time reliability of
rsfMRI by assessing temporal signal-to-noise ratio of five networks. Finally,
we assessed variability in diffusion metrics across scanners and time points. Our
results provide statistical validations for longitudinal work on multiple
systems, especially for structural study. The influence of different equipment in
rsfMRI and DTI related research may be considered, especially DMN and GCC
analysis involved.
Introduction:
Longitudinal MRI studies are becoming
increasingly important to study structural and functional changes in the brain
following pathophysiological conditions or therapeutic treatment. As the long
time duration and multi-center approach were involved more and more in clinical
research, the combining images acquired at different time points and different
scanners require comparable and stable MR imaging measurements over time and
across sites. In this study, we wanted to go beyond this, by measuring imaging
reproducibility in structural, resting state fMRI and diffusion tensor scans
across different time points and scanners in healthy volunteers. First,
we assessed structural imaging variability across different scanners and time
points by calculating volume for seven subcortical structures which are of
particular interest in neurodegenerative diseases. Second, we
evaluated across-scanner and across-time reliability of rsfMRI by assessing
temporal signal-to-noise ratio (tSNR) and the generation of five networks in
healthy volunteers. Finally, we assessed variability in diffusion metrics
across scanners and time points. Materials and Methods:
Subjects were six healthy adults (3 males
and 3 females) with an age range of 23 – 38 years old. The scan is
carried out on different scanners on the same day and on the same scanner every
other week to evaluate the consistency of different equipment and the stability
of the same equipment. For each subject the scans on both scanners were
conducted at approximately the same time of day. Subjects were advised to
maintain the same caffeine intake on scan days and same sleep schedule the
nights before. They were advised not to exercise on the day prior to scanning
and on the day of scanning before the scan.
MRI scans were performed
on two 3 T MRI scanner: Philips Achieva 3.0T with an 8-channel
head coil and Philips Ingenia CX 3.0T with a 15 channel dStream T/R head coil. High-resolution
3D-T1 weighted images were acquired following the ADNI protocol developed for
multi-center intervendor acquisitions [1]. rsfMRI data were acquired using
whole brain T2*-weighted gradient-echo EPI, sensitive to blood oxygenation
level-dependent (BOLD) contrast. Diffusion-weighted imaging was performed with
a spin-echo echo-planar-imaging (SE-EPI) sequence using 64 different diffusion
directions and a b value of 1000 s/mm2.
For 3D T1, the volume for the following 7
ROIs (Thalamus (right), Putamen (right), Hippocampus (right), Amygdala (right),
brain stem, white matter (WM) and cerebrospinal fluid (CSF)) were estimated. For
rsfMRI, tSNR within ROIs were estimated as the ratio of mean signal from all
the voxels within the ROIs divided by the standard deviation across time [2].
We analyses the tSNR within five functional networks, the gray matter (GM), default-mode
(DMN), ventral attention net (VAN), visual
net(VIS) and the somatomotor(SM) network. For DTI, the mean and standard
deviation of FA value across different conditions for genu of corpus callosum(GCC),
body of corpus callosum(BCC), splenium of corpus callosum(SCC) and the total were
calculated. One-way ANOVA was used to test group mean differences. ICC was used
to measure test-retest reliability. A statistical threshold of p<0.05 was
considered significant.Results:
For 3D T1, the ICC values of the volume of
the seven ROIs indicating that there are no significant differences between the
two time points in two scanners compared to the differences between the
subjects, as shown in Table 1. And the intrascanner coefficients in different
brain regions showed excellent reproducibility, especially in the right amygdala, as shown in
Table 2. For rsfMRI, the mean and standard error of tSNR and the ICC across the
two time points for the five networks were shown in Figure 1. And the intrascanner
coefficients in different networks were shown in Table 3. The reproducibility of tSNR were from moderate to good, with the worst performance
in DMN. For DTI, the ICC of FA values across the two
time points for the corpus callosum regions were demonstrated in Table 4. It
illustrated that the time reproducibility of FA values in corpus callosum was good in
both scanners, with the ICC value of GCC was lower. Besides, the intrascanner
coefficients of FA in corpus callosum, BCC and SCC was 0.92,0.98 and 0.81,
respectively. However, that value in GCC was 0.45.Discussion and Conclusion:
This study reported a group of metrics used
to quantify the reproducibility for standard neuroimaging data (structural,
BOLD and DTI) collected across two identical 3.0 T scanners and at two time
points. Our results show the metrics proposed are highly reproducible for
volumetric data either between same scanner at two time points or between the
two different scanners. The rsfMRI showed stable temporal SNR of subject’s
networks across time points except DMN. And the reproducibility was moderate cross scanners. There
was low variation of FA measures between scans for both
intra- and inter-scanner rescanning in corpus callosum. While, when further
analysis, the
reproducibility of FA in GCC was poor across scanners which may be affected by
the registration and segmentation. These findings
provide statistical validations for longitudinal work on multiple systems,
especially for structural study. The influence of different equipment in rsfMRI
and DTI related research may be considered when analyzing data, especially DMN
and GCC analysis involved. Acknowledgements
No acknowledgement found.References
[1] Jack CR Jr, Bernstein MA, Fox NC et al. The Alzheimer’s Disease
Neuroimaging Initiative (ADNI): MRI methods. J Magn Reson Imaging
,2008,27:685–691. https://doi.org/10.1002/jmri.21049
[2] Parrish TB, Gitelman DR, LaBar KS, Mesulam MM. Impact of
signal-to-noise on functional MRI. Magn Reson Med, 2000, 44: 925–9