Jianmin ZHENG1, Jianxiu Lian2, Hong Wang1, Leilei Li1, Yingjuan Chang1, Guorui Hou1, Gang Lin1, Yang Li1, Yang Li1, and Minwen Zheng1
1Xijing Hospital, Air Force Medical University, Xi 'an, China, 2Philips Healthcare, Beijing, China, Beijing, China
Synopsis
Keywords: Blood vessels, Blood vessels, black blood, venous sinus thrombosis
Cerebral venous
sinus thrombosis (CVST) is one of ischemic cerebrovascular diseases with low
morbidity and high mortality. Three-dimensional volumetric isotropic turbo spin
echo acquisition(3D-VISTA)can be adopted, which has high resolution and most blood signals
will be suppressed. In addition, a motion-sensitized and fault-driven
equilibrium(MSDE) technology were
applied for more thorough blood suppression. 3D-VISTA showed SNR and CNR were significantly
higher than when compared with enhanced three-dimensional TFE sequence, which
could more accurately identify CVST. 3D-VISTA sequence can be used to improve
the recognition rate of craniocerebral veins, which could provide valuable evidence
for accurate identification of clinical craniocerebral venous thrombosis.
Introduction
Craniocerebral
venous thrombosis is a disease that tend to be easily misdiagnosed and
mistreated in clinical practice. Digital subtraction cerebral angiography (DSA)
is the gold standard for the diagnosis of CVST, but invasive operation and risk
of increased intracranial pressure limit its widespread application . The
enhanced magnetization preparation rapid gradient echo sequence is commonly
used to diagnose cerebral venous thrombosis. But there is some interference in
the recognition of thrombus due to the contrast agent filling the venous blood
tube after enhancement. The traditional 3D-TFE enhanced sequence cannot be
completely relied upon by the radiologist due to the risk of missed diagnosis
in some small venous thrombi and the volumetric effect of contrast enhancement [1][2].Three-dimensional
volumetric isotropic turbo spin echo acquisition with motion-sensitized and
fault-driven equilibrium (3D-VISTA-MSDE) sequence can suppress blood signals. Therefore, the objective of this study was to
investigate whether 3D-VISTA-MSDE could improve the accurate identification of
cerebral venous thrombosis.Methods
14 patients who had been clinically
suspected to have craniocerebral venous thrombosis were recruited from May to
October 2022. Magnetic resonance imaging (MRI) were performed on a 3.0T Philips
scanner (Ingenia CX, Best, the Netherlands) with a 32-channel head coil. Patients
underwent contrast enhanced with 3D-TFE and 3D-VISTA sequences, Detailed
sequence parameters are shown in Table 1. Two senior radiologists (LY, 13 years with experience and WMQ 15 years with experience) evaluated the images of
patients with cerebral thrombosis. Signal to noise ratio (SNR)
and contrast noise ratio (CNR) were calculated for evaluating image quality[3]. For each patient, the frequency of the two groups
with definite cerebral venous thrombosis was calculated according to the upper
arachnoid sinus, lower arachnoid sinus, straight sinus, sinuses, left and right
transverse sinus, and left and right sigmoid sinus. Paired Chi-square test
was performed on the statistical data with SPSS statistical software (IBM SPSS
Statistics 21 ). P<0.05 was regarded as significantly difference.Results
13 patients (54.6 ±8.2 years old, 10 males) were enrolled
while one patient was excluded because obvious motion artifacts affect image
observation in this study. Detailed information of patients was shown in Table 2. SNR and CNR of 3D-VISTA sequence were
significantly higher than that of 3D-TFE (P=0.033, P<0.001) in Table 3 . The kappa value of κ was 0.373 (p <0.001)between 3D-VISTA and 3D-TFE in Table4.The
representative case showed in figure 1.Discussion
The current study revealed that 3D-VISTA sequence
can significantly improve the detection of cerebral venous thrombosis, which
had better performance of SNR and CNR when compared with traditional sequence. The
diagnostic significance of 3D-VISTA sequence for cerebral venous thrombosis was
higher than that of 3D-TFE sequence. 3D-VISTA sequence had complete blood
inhibition due to blood empty-flow effect and the pair of motion sensitive
gradients (similar to the bipolar gradients of diffusion sequence) [4]. Through motion sensitive gradients, tissue
signals of flow (movement) are attenuated for further inhibit blood [5]. The limitations was the small number of cases and
the lack of multi-sequence joint contrast diagnosis. Conclusion
Enhanced 3D-VISTA sequence can
significantly improve clinical recognition of craniocerebral venous thrombosis.Acknowledgements
No
acknowledgement.References
[1] Niu, PP, Yu, Y, Guo, ZN, et al.
Diagnosis of non-acute cerebral venous thrombosis with 3D T1-weighted black
blood sequence at 3T. J NEUROL SCI. 2016; 367 46-50. doi:
10.1016/j.jns.2016.05.052
[2]Yang, X, Wu, F, Liu, Y, et al.
Diagnostic performance of MR black-blood thrombus imaging for cerebral venous
thrombosis in real-world clinical practice. EUR RADIOL. 2021; 32 (3):
2041-2049. doi: 10.1007/s00330-021-08286-x
[3] Wintersperger, BJ, Runge, VM, Biswas, J, et
al. Brain tumor enhancement in MR imaging at 3 Tesla: comparison of SNR and CNR
gain using TSE and GRE techniques. INVEST RADIOL. 2007; 42 (8): 558-63. doi:
10.1097/RLI.0b013e31803e8b3f
[4]Yang, Q, Fan, Z, Bi, X, et al. Early
detection and quantification of cerebral venous thrombosis by Magnetic
Resonance Black Blood Thrombus Imaging (MRBTI) J CARDIOVASC MAGN R. 2016; 18
(S1): doi: 10.1186/1532-429x-18-s1-p16
[5]Hakim, A, Kurmann, C, Pospieszny, K, et
al. Diagnostic Accuracy of High-Resolution 3D T2-SPACE in Detecting Cerebral
Venous Sinus Thrombosis. AM J NEURORADIOL. 2022; 43 (6): 881-886. doi:
10.3174/ajnr.A7530