Katsumi Kose1, Ryoichi Kose1, Daiki Tamada2, and Utaroh Motosugi3
1MRIsimulations Inc., Tokyo, Japan, 2University of Yamanashi, Chuo, Japan, 3Kofu Kyoritsu Hospital, Kofu, Japan
Synopsis
Keywords: Software Tools, Motion Correction, motion simulation
To
investigate the possibility and limitation of MRI simulations based on the
Lagrange description, MRI experiments and simulations of moving objects were
performed. As a result, we were able to obtain simulated MR images that almost reproduced
the experimental results within practical computation time. However, it was
found that the T2 coherence effect should be reduced for moving objects. It was
also found that more precise simulation is necessary to reproduce the detailed motion
artifacts.
Introduction
Bloch
equations are widely used to numerically reproduce MRI phenomena1-8.
In the Euler description of the Bloch simulation, the nuclear spins are fixed
to a coordinate system, while in the Lagrange description, the positions of the
nuclear spins are followed throughout the pulse sequence. The Euler
description is widely used in quantitative MRI, and the Lagrange description is
mainly used in simulation of fluid flow in MRI. In this study, Bloch equations
based on the Lagrange description are applied to moving objects to clarify the
possibility and limitations of MRI simulations.Materials and methods
Figure
1 shows the phantoms used in this study. As shown in Fig. 2(a), the rotation
experiment was performed by rotating the phantom with a DC servo motor through
an acrylic pipe. The MRI system used in the experiment was a digital MRI system
with a 1.5 Tesla superconducting magnet with a 280 mm bore. The pulse sequence
was a GRE sequence with TR=20ms and TE=8ms. 20 consecutive images were acquired
at 3 second time-intervals at a constant rotation speed. As shown in Fig. 2(b),
the translation experiments were performed using a driving system consisting of
a non-magnetic ultrasonic motor and plastic gears. The water phantom was placed
on a wooden dolly and driven with two 1.2 m long wooden cranks connected to the
driving system. The amplitude of the translational motion was 30 mm, and the
period was 3.8 seconds to simulate respiratory motion. The MRI system used in
this experiment was a 3T clinical MRI (SIGNA Premier). The pulse sequences were
axial and coronal 2D GRE sequences with TR=4.224ms, TE=2.1ms, and 8mm slice
thickness. 40 consecutive images were acquired at 0.54 second time-intervals.
Bloch
simulation of the rotation was performed using the Lagrange description, in
which the position of the proton rotates at a constant angular velocity, and at
each position, the proton receives RF and gradient pulses according to the
pulse sequence and simultaneously undergoes nuclear magnetic relaxation. The
time increments for the calculation were 20 microseconds, and the Bloch
equations were solved numerically at each time increment to calculate the MR
signal. The number of spins used for the calculation was 65,536 and the total
number of the time steps for one image was 138,000. Bloch simulation of the translational
motion was performed by the Lagrange description, in which the numerical
phantom moved sinusoidally in the horizontal plane along the z-axis. The time
interval for the calculation was set to 8 microseconds, the same as the signal
sampling interval, and the Bloch equations were solved. The number of spins
used for the calculation was 1,4833,328 and the total number of the time steps
for 8 images was 538,624. All calculations were performed using a single core
Intel CPU (4 GHz clock).Results
Figure
3 shows cross-sections of the rotating phantom acquired with the experiment and
the Bloch simulation. The calculation time for one image was about 4 minutes. Except
for the inhomogeneity of the RF magnetic field effect, motion artifacts are
reproduced almost exactly. Figure 4 shows axial images of the moving phantom
acquired with the experiment and the Bloch simulation. T2 values of 85 ms and 5
ms were used for the Bloch simulation. The calculation time for 8 images were
about 16 hours. The blurred image of the container in the experiment is not
reproduced in the simulation, but the position of the containers is accurately
reproduced. Artifacts were seen in the images calculated with T2 of 85 ms,
whereas they were almost nonexistent in the images calculated with T2 of 5 ms. Figure
5 shows coronal images of the moving phantom acquired with the experiment and
the Bloch simulation, and the absolute values of the point spread function (PSF)
calculated along the direction of the translational motion. The calculation
time for 8 images were about 3 hours. The PSF clearly demonstrates the presence
of the artifacts shown in the simulated images.Discussion
Bloch
simulation based on the Lagrange description allowed us to obtain simulated
images that can be compared with the experimental images for moving objects. In
the simulation of rotation, the artifacts were reproduced almost accurately. On
the other hand, in the simulation of translation, it was possible to reproduce
the position of the translational object, but it was difficult to reproduce the
detailed artifacts. The reason why artifacts could not be reproduced may be due
to insufficient experimental setting of the moving object and/or imperfection
of the imaging system (eddy current effects, etc.). The artifacts associated
with the inhomogeneity of the magnetic susceptibility distribution may be caused
by a special magnetic field distribution in the sample. Therefore, more precise
calculations will be required to reproduce them. In the Lagrange description,
unlike the Euler description, the positions of the spins vary with time, which
reduces the effect of T2 coherence. Therefore, in the Bloch simulation, it is
necessary to use a shorter T2 value than the measured T2.
In
conclusion, the Lagrange description can reproduce MR images of moving objects,
but more careful calculations are needed to reproduce the detailed artifacts.Acknowledgements
No acknowledgement found.References
[1]
Benoit-Cattin H, Collewet G, Belaroussi B, Saint-Jalmes H, Odet C. The SIMRI
project: a versatile and
interactive
MRI simulator. Journal of Magnetic Resonance. 2005; 173:97–115.
[2]
Stoecker T, Vahedipour K, Pflugfelder D, Shah NJ. High-performance computing
MRI simulations.
Magnetic
Resonance in Medicine. 2010; 64:186–193.
[3]
Xanthis CG, Venetis IE, Chalkias AV, Aletras AH. MRISIMUL: A GPU-Based Parallel
Approach to MRI
Simulations.
IEEE Transactions on Medical Imaging. 2014; 33:607–617.
[4]
Kose R, Kose K. BlochSolver: A GPU-optimized fast 3D MRI simulator for
experimentally compatible pulse sequences. J Magn Reson 2017;281:51–65.
[5]
Liu F, Velikina JV, Block WF, Kijowski R, Samsonov AA. Fast realistic MRI
simulations based on generalized multi-pool exchange tissue model. IEEE Trans
Med Imaging 2017; 36:527–37.
[6]
Marshall I. Computational simulations and experimental studies of 3D
phase-contrast imaging of fluid flow in carotid bifurcation geometries. J Magn
Reson Imaging. 2010; 31:928-34.
[7]
Xanthis CG, Venetis IE, Aletras AH. High performance MRI simulations of motion
on multi-GPU systems. J Cardiovasc Magn Reson. 2014; 16:48.
[8]
Fortin A, Salmon S, Baruthio J, Delbany M, Durand E. Flow MRI simulation in
complex 3D geometries: Application to the cerebral venous network. Magn Reson
Med. 2018; 80:1655-1665.