Jae-Youn Keum1, Jeong Hee Yoon2, Michael Garwood3, and Jang-Yeon Park1,4
1Department of Biomedical Engineering, Sungkyunkwan University, Suwon, Korea, Republic of, 2Department of Radiology, Seoul National University Hospital and College of Medicine, Seoul, Korea, Republic of, 3Department of Radiology, Center for Magnetic Resonance Research, Minneapolis, MN, United States, 4Department of Intelligent Precision Healthcare Convergence, Sungkyunkwan University, Suwon, Korea, Republic of
Synopsis
Keywords: Pulse Sequence Design, Diffusion/other diffusion imaging techniques, Hyperbolic-Secant Pulse
Blood
flow artifacts sometimes occur in 2D spin-echo sequence despite its intrinsic flow-suppression effect. Pre-saturation
pulses have widely been used for flow-suppression, but it has several
disadvantages such as requirement of additional RF pulses and being not
effective at suppressing relatively slow blood flow signal. In this study, we
propose a more effective flow-suppressed spin-echo sequence using
hyperbolic-secant (HS) pulses for π/2-excitation and π-refocusing.
The proposed method was applied to 2D spin-echo diffusion EPI for liver imaging.
Introduction
Diffusion-weighted
imaging (DWI) is widely used in clinical applications due to its high
sensitivity to Brownian motion of water molecules. Liver DWI using a spin-echo
diffusion EPI sequence is one of the popular clinical applications of DWI. However,
liver DWI suffers from cardiac and respiratory motions, particularly in left
lobe of the liver1. To reduce motion-induced artifacts,
motion-compensated diffusion-weighted gradients with gradient moment nulling
(e.g. bipolar diffusion-weighted gradients) have been used, and some optimization
techniques have also been recently proposed to reduce the minimum TE2-4.
However, motion-compensated gradients can make moving blood spins to be in-phase,
so bright blood signals are often detected in DWI. Several methods have been
proposed to eliminate these bright blood signals. Van et al. applied monopolar
diffusion-weighted gradient along the anterior-posterior direction with bipolar
diffusion-weighted gradients applied in other directions5, and Zhang
et al. proposed a motion-robust and blood-suppressed DWI sequence with a moderate
first-moment motion sensitive (M1) value4. However, since these
techniques use moderate M1 values, they provide only partial blood suppression
and motion compensation. In this study, we propose a flow-suppressed 2D
spin-echo EPI sequence that uses hyperbolic-secant (HS) pulses for both π/2-excitation and π-refocusing, with high B1-insensitivity due to the adiabatic
property of the π refocusing HS pulse. The performance of this sequence was
demonstrated by theory, simulation, and liver imaging.Methods
Ro et
al. proposed a flow-suppression technique using tailored π/2 and π RF pulses
that could produce opposing quadratic phases in a spin-echo sequence6.
On the other hand, Park et al. demonstrated that when HS pulses are used for
both π/2-excitation and π-refocusing in a spin-echo sequence, opposing
quadratic-function-like non-linear phase distributions would be generated and could
be canceled in specific conditions7. We used one of these conditions
to achieve flow suppression, i.e., Tp,1 = Tp,2 , BW1 = 2BW2 , G1 = 2G2, where Tp is a pulse
duration, BW is a pulse bandwidth, G is a
slice-selection gradient. Subscripts 1 and 2 denote π/2-excitation and π-refocusing,
respectively (Fig. 1). First, numerical simulations were performed to
investigate flow suppression as a function of flow velocity and pulse
parameters, using HS pulses with Tp,1 = Tp,2 = 5.12 ms, BW1/2π = 2BW2/2π = 4.00 kHz. Then, human liver images were acquired with the
proposed method using a body coil on a 3T scanner (Prisma, Siemens, Erlangen,
Germany). Scan parameters were as follows: TR/TE = 1200/80 ms, FOV = 280×280 mm2,
matrix size = 128×128, slice thickness = 6 mm, b(average) = 50(2) s/ mm2,
readout-BW = 1502 Hz/px, acceleration factor = 2, and fat suppression = Spectral Attenuated Inversion-Recovery (SPAIR)8. HS pulses
with Tp,1 = Tp,2 = 5.12 ms, BW1/2π = 2BW2/2π = 4.00 kHz. For
comparison, same experiments were performed using sinc pulses. In all human experiments,
prospective respiratory gating was used.Results
Figure
2a shows the simulation results that non-linear phase profiles are fully
compensated for static spins. In contrast, for moving spins, a nearly linear
phase distribution is formed due to the shifted quadratic phase distribution
with opposite polarity (Fig. 2b), resulting in signal loss within a slice (Fig.
3a). Signal intensity for a given velocity decreased as pulse duration and
bandwidth increased (Fig. 3b, flow velocity = 10 cm/s). In Fig. 4, thanks to
the adiabatic π refocusing HS pulse, human liver images acquired with HS pulses
showed better image quality than sinc pulses in terms of SNR, especially in the
abdominal periphery. Moreover, moving blood signals were suppressed well, especially
in the abdominal aorta and right branch of the portal vein (yellow arrows).Discussions
The
proposed flow-suppression spin-echo diffusion EPI sequence using HS pulses was not
only effective to suppress blood flow in the liver, especially prominent in the
abdominal aorta, but also able to provide better image quality due to less
sensitivity to B1 inhomogeneity. An additional advantage is that the
parameters can be optimized to target a specific velocity of blood flow. When
it is targeted to suppress slow blood flow (e.g., 2.5cm/s), it can also be more robust to B0
inhomogeneity effects due to relatively high RF pulse bandwidth (e.g., BW1/2π = 2BW2/2π = 6.13 kHz). From a practical viewpoint, the proposed method does not require any
additional RF pulses or gradients, except for replacing the RF pulses with HS
pulses. A limitation of the proposed method at this stage is that flow
suppression is achieved only in the slice-selection direction.Conclusions
A flow-suppressed
2D spin-echo EPI sequence with high B1-insensitivity was proposed
using hyperbolic secant pulses and successfully demonstrated by simulations and
human liver imaging. Although we presented the method in spin-echo diffusion
EPI, the technique can also be applied to any types of spin-echo sequences.Acknowledgements
This
work was supported by the National Research Foundation of Korea (NRF) grant
funded by the Korea government (MSIT): NRF-2020R1A2B5B02002676, NRF-2021R1A4A5032806.References
1.
Murphy P, Wolfson T, Gamst A, Sirlin C, Bydder M. Error model for reduction of
cardiac and respiratory motion effects in quantitative liver DW-MRI. J Magn
Reson Imaging. 2013;70:1460–1469.
2. Aliotta E, Wu HH, Ennis DB. Convex optimized diffusion
encoding (CODE) gradient waveforms for minimum echo time and bulk
motion-compensated diffusion-weighted MRI. Magn Reson Med. 2017;77:717–729.
3. Peña Nogales O, Zhang Y, Wang X, et al. Optimized Diffusion‐Weighting
Gradient Waveform Design (ODGD) formulation for motion compensation and
concomitant gradient nulling. Magn Reson Med. 2019;81:989–1003.
4. Zhang Y, Peña‐Nogales Ó, Holmes JH, Hernando D.
Motion‐robust and blood‐suppressed M1‐optimized diffusion MR imaging of the
liver. Magn Reson Med. 2019;00:1–10.
5. Van AT, Cervantes B, Ogino T, et al. Partial
velocity‐compensated diffusion encoding for combined motion compensation and
residual vessel signal suppression in liver DWI. In Proceedings of the 26th
Annual Meeting of ISMRM, Paris, France; 2018.
6. Ro, Y. M, Cho, Z. H. A novel flow-suppression technique
using tailored RF pulses. Magn Reson Med. 1993;29(5):660–666.
7. Park J-Y et al, Spin-echo MRI using π/2 and π hyperbolic
secant pulses. Magn Reson Med. 2006;55:848-857.
8. Lauenstein TC,
Sharma P, Hughes T, Heberlein K, Tudorascu D, Martin DR. Evaluation of
optimized inversion-recovery fat-suppression techniques for T2-weighted
abdominal MR imaging. J Magn Reson Imaging 2008;27(6):1448–1454.