Mark Armstrong1 and Dan Xiao1
1University of Windsor, Windsor, ON, Canada
Synopsis
Keywords: Pulse Sequence Design, New Trajectories & Spatial Encoding Methods
Pi Echo Planar Imaging (PEPI) has a significantly
reduced gradient duty cycle compared to FSE. It has a great potential for low field MRI where concomitant fields
are significant. However, its application is limited due to the requirement of a near perfect refocusing pulse to avoid coherence pathway artifacts.
In this work an optimized phase cycling scheme is proposed which minimizes
coherence pathway contributions to the signal. This method has been validated in
simulations and experiments showing a reduced sensitivity to flip angle
compared to the conventional XY-16 method. Artifact-free 2D PEPI was achieved.
Introduction
Fast Spin Echo (FSE)
experiments provide good quality images in a rapid acquisition. For high
resolution imaging, high gradient duty cycle is required in FSE, which may lead
to significant concomitant field effects in low field imaging. The Pi Echo Planar
Imaging (PEPI)1 sequence employs unbalanced gradients, allowing phase accumulation
over the entire echo train, so that the gradient duty cycle can be drastically
reduced. A shorter echo spacing may also be achieved. PEPI requires near ideal
180° refocusing pulses to eliminate the coherence pathway artifacts, which has
limited its application to 3D imaging of small subjects confined to the
homogeneous region of the RF coil. Sufficient quality 2D PEPI could not be
obtained due to the imperfect slice profile. In this work, an optimized phase
cycling scheme is proposed to minimize coherence pathway artifacts in the PEPI
experiments with non-ideal refocusing RF pulses and to enable 2D PEPI imaging.Methods
The phase and
amplitude of coherence pathways are calculated. The cost function in the
optimization is $$$ \sum{_i}\sum{_j}\vert{A_j \sum{_k} e^{i(\varphi_j\cdot \Phi_k - \theta{reciever,i})}}\vert$$$, where i, j, and k are indices of the echo number, coherence
pathway, and repetition
of scan, respectively. Φk is a vector containing the phases of all the
RF pulses in the kth repetition, and φj is a vector storing the phase dependence of
the jth coherence
pathway on the RF phase. θreceiver,i is the receiver phase at the ith echo time
as determined by the phase of the desired coherence pathway. Aj is the amplitude of coherence pathway. The optimization was performed for each combination of the targeted echo train length and number of averages.
The
schemes were validated with simulation and PEPI experiments on a resolution
phantom and a mouse head. The pulse sequence diagram for the 2D PEPI scheme is
shown in Fig. 1, where a small amplitude and/or short duration phase encoding gradient
is applied before every second echo. All experiments were performed at 1T. For the resolution phantom, the FOV was 25mm
x 25mm with a 0.4mm x 0.4mm resolution and 1mm slice thickness. A 1350ms TR, 4ms TE, and 16 ETL with 4 averages were used. For
the mouse head image, the FOV was 20mm x 20mm with a 0.16mm x 0.16mm resolution
and 1mm slice thickness. A 975ms TR and 5ms TE with a 16 ETL was used. A scheme
optimized with 10 averages was repeated 4 times for a total of 40 averages.Results and Discussion
An
example of the optimized phase cycling scheme is shown in Fig. 2. The optimized
phases appear irregular. The conventional XY-162 phase cycling scheme is plotted
for comparison.
PEPI
images were simulated with a 170° refocussing angle as shown in Fig. 3. Severe coherence pathway artifacts are observed in the phase encoding direction in each
of the unaveraged images. After averaging, the artifacts are mostly removed, demonstrating
the effectiveness of the phase cycling scheme.
The
optimized phase cycling scheme was applied in 2D slice selective PEPI
experiments as in Fig. 4. In Fig. 4a-c, the proposed scheme is shown to be
quite insensitive to changes in flip angle, with only low intensity coherence
pathway artifacts at a nominal 160°
refocussing. The conventional XY-16 method had coherence pathway artifacts even
at a nominal 180°, as shown in Fig.
4d. It was much more sensitive to non-ideal refocusing. The image
quality severely degraded as the flip angle decreased, as in Fig. 4e and Fig. 4f.
The scheme was applied to a 2D slice
selective PEPI on a mouse head as shown in Fig. 5. The conventional XY-16 method
had significant coherence pathway artifacts, distorting a large portion of the
image as shown in Fig. 5b. The optimized phase cycling scheme provided a
relatively artifact free image in Fig 5a, which had a similar quality to an FSE
image (Fig. 5c). For the same resolution, the FSE required a maximum gradient
amplitude of 164 mT/m compared to the 47 mT/m in PEPI. The FSE had a 49 times
higher gradient duty cycle compared to PEPI in the phase encoding dimension.
There are slight differences in the contrast, and the SNR appeared lower in
PEPI. This is mostly because FSE used the relevant coherence pathways, while the
phase cycling cancelled all the coherence pathways except for the primary spin
echoes.
Signal averaging may generally be necessary
for high resolution and/or low field imaging. The proposed method is advantageous
in these applications that no extra scan time is required.Conclusion
A phase cycling
scheme has been proposed which minimizes the contribution of unwanted coherence
pathways for PEPI experiments. The artifact reduction removed the perfect
refocusing condition previously required by PEPI. This allows high quality slice
selective 2D measurement that can be implemented in clinical and preclinical
scanners to achieve high resolution imaging with low gradient amplitude and much reduced gradient duty cycle.Acknowledgements
MA thanks NSERC Canada for a CGS scholarship. DX thanks NSERC Canada for a Discovery grant.References
1. Xiao, D., & Balcom, B. J. (2015). "π Echo-Planar Imaging
with concomitant field compensation for porous media MRI" Journal of
Magnetic Resonance, 260, 38–45.
2. Gullion, T., Baker, B., &
Conradi, M., (1990) "New, compensated Carr-Purcell sequences" Journal of
Magnetic Resonance, 89, 479-484.