Molin Zhang1, Georgy Guryev1, Nick Arango1, Jason Stockmann2,3, Jacob White1, and Elfar Adalsteinsson1,4,5
1EECS, MIT, Cambridge, MA, United States, 2Athinoula A. Martinos Center for Biomedical Imaging, Charlestown, MA, United States, 3Department of Radiology, Harvard Medical School, Charlestown, MA, United States, 4Harvard-MIT Health Sciences and Technology, MIT, Cambridge, MA, United States, 55 Institute for Medical Engineering and Science, MIT, Cambridge, MA, United States
Synopsis
Keywords: Pulse Sequence Design, RF Pulse Design & Fields
We explored
the utility of shim array fields in the mitigation of in-plane B1+ inhomogeneity
for a slice-selective excitation at 7T and achieved better magnitude
of transverse magnetization compared with birdcage transmit using conventional
single-channel RF excitation. Approximately 2:1 range of B1+ was
mitigated, at the cost of increased peak RF power, shim array current demands
up to 20 A, and minor out-of-slice sidelobes.
Introduction
At 7T, within-slice
B1+ inhomogeneity compromises SNR and contrast in diagnostic imaging with
conventional single-channel RF systems. To address this, parallel transmission (pTx)
[1,2,3,4] methods have been demonstrated to successfully mitigate flip angle
variations due to B1+ inhomogeneity, but with additional constraints on SAR
management over single-channel RF excitations.
Recent
developments of independently driven multi-coil (MC) shim arrays enrich the
design space by providing spatially non-linear DB0 fields [5,6]. Such arrays have been
proposed and demonstrated to improve B0 shimming as well as selective
excitation [7] designs. A critical tool for spatially-selective RF designs with
shim arrays is a time-discrete Bloch simulator with auto-differentiation [8],
which can be used to optimize both the RF pulse envelope and shim currents. Such
design methods have yet to be applied to B1+ mitigation at high.
Here
we explore, in simulation, the potential utility of a shim array augmenting a
standard single-channel RF system for improved slice-selective excitation with B1+ inhomogeneity mitigation at 7T. The RF pulse is limited
to a 4-ms duration in the presence of a constant slice-selective gradient, and
a set of optimized currents to drive a large shim array of 72 coils. The B1+
inhomogeneity derives from simulations of a birdcage excitation of an adult
brain at 7T. Preliminary results demonstrate successful mitigation of
within-slice B1+ inhomogeneity with a range of approximately 2:1,
peak-to-trough, with shim array currents under 20A, and out-of-slice ripple
that’s limited to anterior peripheral regions. Methods
As studied in previous works [7], the time-discrete Bloch
simulator approximates the Bloch simulation as a sequence of rotation operators.
Each rotation operator is determined by the total B0 (B0 + DB0) fields in
the z direction and RF (B1x and B1y) in transverse orientation. The design is
configured to simultaneously optimize the RF envelope for a 90-degree
excitation as well as shim currents, given a constant slice selective gradient
according to the cost function below,
$$\underset{I \in \mathbb{R}^{{n_{t}} \times {n_{\text {coil}}}}, b \in \mathbb{C}^{n_t}}{\operatorname{argmin}} \mathcal{L}=f\left(\left|\boldsymbol{M}_{x y}(I, b)\right|,\left|\boldsymbol{M}_D\right|\right)+\lambda_1 \mathcal{R}_1(I)$$
$$s.t. \begin{array}{ll} & \|b\|_{\infty} \leq b_{\max } \\& \|I\|_{\infty, \infty} \leq I_{\max } \\& \left\|\frac{d I}{d t}\right\|_{\infty, \infty} \leq s_{\max }\end{array}$$
Here, $$$f$$$ is the least-square error on magnitude; $$$I$$$ is the coil current; $$$b$$$ holds the coefficients
of the birdcage transmit fields (RF=b*birdcage field); $$$n_t$$$ is the
number of time point; $$$n_{coil}$$$ is the number of shim coils in the shim
array; $$$M_D$$$ is the desired slice profile; $$$M_{xy}$$$ is the actual
excitation profile generated by the birdcage fields and the time-varying ΔB0
fields under the time-discrete Bloch simulator; regularizes shim array currents; $$$b_{max}$$$ is
the maximum coefficients; $$$I_{max}$$$ is the maximum current constraint for
single coil and $$$s_{max}$$$ is the bandwidth constraint for the coil
current.
We applied a stochastic offset strategy to refine the slice
profile in the interval of the fixed spatial points for the optimization. For
each iteration, we added a random offset to the coordinates of the points that
are evaluated and optimized with the auto-differentiation. This approach
promotes better adherence of the desired slice profile throughout the volume as
opposed to being limited to a set of fixed points. Results and Discussion
The simulated adult brain data has dimension of 40*40*50 voxels
with the resolution of 5 mm. As shown in Fig.1, we heuristically place 72 coils
with 4 cm diameter at the surface of a cylinder surrounding the head. The quadrature-driven
birdcage fields for the adult brain were simulated using MARIE [9,10] in the CP
mode.
The objective function is limited to only the magnitude of the transverse
magnetization, which for each voxel is derived by a sum over 80 equally spaced
sub-voxel.
Fig. 2 B,C show the transverse magnetization of the slice of
interest after applying SLR coefficients to the birdcage fields using
time-discrete Bloch simulator, demonstrating an approximately 2:1 ratio of max to
min values of the magnitude of field within the slice.
Compared with the within-slice nonuniform flip angle with the SLR waveform,
the optimized RF coefficients along shim currents driving a multi-coil array,
result in a more uniform transverse magnetization magnitude across the slice. Figure
3 shows the optimized coefficients waveform and phase as well as the coil
current. The peak RF during the excitation process is 0.047 mT. Seen from
figure 4, this method improves the magnetization profile of the slice of
interest and reduces RMSE.
Adjacent slices, above and below the selected axial one, demonstrate
a small amount of through-slice sidelobes, which are most prominent near the
anterior periphery of the brain, see Fig.5. In future work, such sidelobes may
be addressed by adjusting weights in cost function for different spatial regions.
Future work will also implement rewinding of phase within the selected slices
by application of shim array fields and gradient after the RF waveform.Conclusion
In this work,
we explored the potential usage of shim array fields in the mitigation of B1+
inhomogeneity for a slice excitation at 7T and achieved better magnitude of transverse
magnetization compared with birdcage transmit using conventional single-channel
RF excitation.Acknowledgements
This research was supported by
NIH R01 EB017337, U01 HD087211, R01 HD100009, and R01 EB006847. References
1. Katscher, Ulrich, et al. "Transmit sense."
Magnetic Resonance in Medicine: An Official Journal of the International
Society for Magnetic Resonance in Medicine 49.1 (2003): 144-150.
2. Zhu, Yudong.
"Parallel excitation with an array of transmit coils." Magnetic
Resonance in Medicine: An Official Journal of the International Society for
Magnetic Resonance in Medicine 51.4 (2004): 775-784.
3. Ibrahim, Tamer S., et al.
"Effect of RF coil excitation on field inhomogeneity at ultra high fields:
a field optimized TEM resonator." Magnetic resonance imaging 19.10 (2001):
1339-1347.
4. Setsompop, Kawin, et al.
"Slice‐selective RF pulses for in vivo B inhomogeneity mitigation at 7
tesla using parallel RF excitation with a 16‐element coil." Magnetic
Resonance in Medicine: An Official Journal of the International Society for
Magnetic Resonance in Medicine 60.6 (2008): 1422-1432.
5. Stockmann, Jason P., et
al. "A 32‐channel combined RF and B0 shim array for 3T brain imaging."
Magnetic resonance in medicine 75.1 (2016): 441-451.
6. Truong TK, Darnell D, Song
AW. Integrated RF/shim coil array for parallel reception and localized B0
shimming in the human brain. Neuroimage 2014;103:235–240.
7. Zhang, Molin, et al. "Selective RF excitation
designs enabled by time‐varying spatially non‐linear Δ B 0 fields with
applications in fetal MRI." Magnetic Resonance in Medicine 87.5 (2022):
2161-2177.
8. Luo, Tianrui, et al.
"Joint Design of RF and gradient waveforms via auto-differentiation for 3D
tailored excitation in MRI." IEEE Transactions on Medical Imaging 40.12
(2021): 3305-3314.
9. Villena, Jorge Fernández,
et al. "Fast electromagnetic analysis of MRI transmit RF coils based on
accelerated integral equation methods." IEEE Transactions on Biomedical
Engineering 63.11 (2016): 2250-2261.
10. Guryev, G. D., et al. "Fast
full-wave patient-specific field simulations in seconds with MARIE 2.0"
ISMRM 2022.