Masahiro Takizawa1, Takashi Nishihara1, and Chikako Moriwake1
1FUJIFILM Healthcare Corporation, 2-1, Shintoyofuta, Japan
Synopsis
Keywords: Pulse Sequence Design, Lung
Cylinder-shaped
pre-saturation pulse train is modified to achieve non subtract scheme for
non-contrast enhanced spatially-selective and time-resolved vessel imaging. The
target vessel is selected by cylinder-shaped pre-saturation, and the dynamics
of blood flow in the target vessel is observed by changing the number of applied
pre-saturation pulses. The developed pulse train was demonstrated to visualize dynamics
of a target pulmonary vessel in the lung.
Introduction
Pulmonary
artery coil embolization (PACE) is often used to treat pulmonary arteriovenous
malformations (PAVM). Before PACE, CT or contract enhanced (CE) time-resolved
MRA is used to specify the target vessel. To label the blood flow, a few non-CE
MRA method using IR or saturation pulse are developed. However, it is difficult
to visualize the target vessel using conventional labelling pulses since the
blood vessels have complex structure. We demonstrated non-CE selective lung
imaging using a cylinder-shaped pre-saturation pulse (cylinder pulse) [1-2].
This method uses subtraction between labeled and non-labeled images to
visualize the target vessel, therefore misregistration caused by patient motion
becomes serious problem. In this study, we modified the sequence design to
achieve non subtract scheme.Methods
Pulse sequence diagram
Figure 1 shows the sequence diagram of the Cyl.-PSat pulse and the modified
pulse train. To suppress residual vessel signals in the target area, a conventional
saturation (Conv.-Sat) pulse, interval time between the Conv.-Sat pulse and
main scan were introduced to the 3D FSE sequence. In this study, Cyl.-PSat
pulse train following the Conv.-Sat pulse was used to suppress blood flow in
the target vessel and the number of cylinders can be increased up to 48 to
observe flow dynamics. Furthermore, Fatsat pulse was inserted between Cyl.-Psat
and main scan to suppress artifacts from body wall motion.
[Phantom experiment]
Experiment was conducted on a 3T whole body MRI system. A phantom
with small constant flow was scanned to observe relationship between the flow
distance and the Cyl.-PSat count. In this phantom, the target signal became
null at the inversion time of 640 msec. So the Cyl.-PSat count were set to 4,
8, 12, 16, 20, 24, 28, and 32.
[Volunteer study]
The
study was approved by the internal review board of FUJIFILM Healthcare Corp. The
Cyl.-PSat was set for the pulmonary artery for the volunteer as shown in fig. 2.
To visualize only inflow blood signal, the Conv.-Sat pulse was set on entire
region of right lung. To set blood signal null in the target area at the point
of excitation for main scan, FA of the Conv.-Sat pulse was set to 180 degrees. The
interval time between the Conv.-Sat pulse and main scan was investigated using
maximum number of Cyl-PSat count before experiment. And to achieve continuous
suppression of inflow blood signal, the duration of Cyl.-PSat pulse was minimized
(20msec). The sequence parameters were as follows; FOV 350mm, TR/TE
4000/41.4msec, slice# 144, thickness 2mm, matrix 256×140, and respiratory gating.Results
[Phantom experiment]
Figure 3 shows the result of the phantom experiment. By decreasing
the Cyl.-PSat count, the water saturation length was increased, and the water
gradually disappeared in the image.
[Volunteer study]
It was
found that all blood flow signal were null when 850msec was used as the
inversion time. As a result, the maximum number of the Cyl.-PSat with 850msec
of the inversion time was 40. To visualize the flow dynamic, the Cyl.-PSat count
were selected as 0 (no Cyl.-PSat), 10, 15, 20, 30, 35, and 40. Maximum
intensity projection was used to show the whole flow dynamics of target vessel.
The length of the visualized artery becomes longer when the Cyl.-PSat count decreases
(Fig. 4).Discussion
The
phantom study and a volunteer study clearly demonstrated that the modified
pulse train can visualize the flow dynamics of the spatially selected vessel. For
the non-subtract scheme, it is important to adjust the interval time between
the Conv.-Sat pulse and main scan to set blood signal null in the target area
at the point of excitation for main scan. As for the vessel visualization, even
no Cyl.-PSat case does not show veins in healthy volunteer. In the actual PAVM case,
the blood flow in the target vein might be observed in the earlier phase than that
in the normal vein. Therefore, it is expected that an interval time that is less
than 1000 msec might be useful.Conclusion
Cylinder-shaped
pre-saturation pulse train was modified to achieve non-subtracted blood flow
imaging. It enabled changing the drawing length of the target vessel by
changing the Cyl.-PSat count, which reflects the flow dynamics in the target
vessel.Acknowledgements
No acknowledgement found.References
[1]
T. Nishihara, et al. Selective MRA for Portal Venography Using Beam Saturation
Pulse, 22nd ISMRM (2014), p. 1501.
[2] M. Takizawa, et al. Modified non-contrast enhanced spatially-selective
and time-resolved vessel imaging by using cylinder-shaped pre-saturation pulse
train in the Lung, ISMRM (2022), p. 4173.