Jun Zhao1,2, Yupeng Cao1, Weinan Tang3, Quelu Chen4, Wentao Liu1, and Dong Han1
1Key Laboratory of Biological Effects and Safety, National Center for Nanoscience and Technology, beijing, China, 2School of Future Technology, University of Chinese Academy of Sciences, Beijing, China, 3Wandong Medical Inc, Beijing, China, Beijing, China, 4Department of Radiology, Wenzhou Central Hospital, Affiliated Dingli Clinical Institute of Wenzhou Medical University, Wenzhou, China
Synopsis
Keywords: Pulse Sequence Design, Lung, hard pulse encoding, variable TE, frequency modulation, spiral
Due to its short T
2*
value and low proton density, MR imaging of the lung is quite challenging. Recently,
ultra-short echo time (UTE) techniques, such as stack-of-spiral, were used to
conduct lung MRI with good image quality. Here, based on stack-of-spiral, a frequency modulation hard pulse encoding
(HAPEN) 3D UTE
was proposed to optimize the echo time of different phase encoding steps in the slice
direction. As a result, HAPEN can achieve shorter TE and get better SNR in
human lung MRI compared to the traditional stack-of-spiral UTE sequence.
Introduction
Because of its free of ionizing radiation, MRI was thought to be a good technique
for lung imaging of patients required longitudinal follow-ups [1].
However, due to its short T2* value and low proton density, MR
imaging of the lung is still challenging. Recently, new techniques have been
developed to improve image quality and reduce scan time. Several studies have
demonstrated that the 3D UTE technique with the stack-of-spiral acquisition enabled
high-resolution morphological lung imaging and was less sensitive to motion
effect than typical radial acquisition [2]. However,
the time efficiency of the stack-of-spiral acquisition was still limited due to
a large number of phase encoding steps in the slice direction [3].
In addition, a refocusing gradient was required if a slice-selective rf pulse was
used, which needed a longer RF pulse than non-selective excitation [4].
Those properties limited the
echo time of this technique.
Here, a frequency modulation hard pulse encoding (HAPEN) 3D stack-of-spiral technique
was proposed to optimize the echo time of different phase encoding steps in the
slice direction. The optimization of TE was
implemented in the following ways. First, a hard pulse was used for signal
excitation, and the hard pulse will
move step by step during a
fixed maximum Gz to speed up slice encoding and minimize T2* decay. Second,
the in-plane spiral trajectories were sampled immediately after the end of each
hard pulse excitation.
Third, frequency modulations
are used to broaden the excitation range in the case of a large gradient.
Both simulation and in-vivo imaging were performed to compare the HAPEN and
traditional stack-of-spiral technique [5].Methods
This study was approved
by the National Center for Nanoscience and Technology Ethics Board, and written
informed consent was given by all study participants. All studies were
performed on an i_Space 1.5 T MRI scanner (Wandong Medical Inc, Beijing, China)
with an 8-channel anterior body coil in combination with an 8-channel posterior
spine coil. A sponge phantom was used to verify the ability to image the short
T2 component of the HAPEN sequence. The
cleaning sponge was cut to a dimension of 75 x 65 x 25mm3 and filled
with water volumes varying from about 30%.
One healthy volunteer was imaged with a HAPEN sequence and traditional stack-of-spiral.
The echo time was 0.07ms and 0.2ms, respectively. The spatial resolution for
both sequences was 1.2x1.2x7.0mm3. HAPEN for the human study was
acquired for FOV=160 x 160mm2, TR = 50ms TE =
0.07ms, flip angle = 8°, slice thickness=7mm, spiral
arms = 96, duration of readouts = 3.5ms. All images were processed using MATLAB.Results
The
details of the HAPEN 3D UTE sequence are illustrated in Figure 1. Under the same hardware conditions,
the HAPEN sequence has a faster slice encoding speed and shorter echo time than
the traditional stack-of-spiral sequence.
Figure
2 shows that the excitation range is more extensive and uniform after adding
frequency modulation.
The results of sponge phantom imaging are
demonstrated in Figure 3. As it was shown, the capability of the proposed HAPEN
sequence to image short T2 component was verified. The fitted T2*
value of the sponge was 2.209ms.
As
depicted in Figure 4, higher SNR and better image quality of HAPEN sequence were
shown in in-vivo lung imaging compared to the traditional stack-of-spiral.Discussion and conclusion
This
study proposed a HAPEN sequence for a rapid encoding technique of high-resolution
UTE lung imaging using clinical MRI scanners. The major advantage of the HAPEN
sequence over traditional stack-of-spiral acquisition is its time-efficient
slice encoding to optimize the echo time of different phase encoding. These
features make the HAPEN sequence suitable for imaging tissues with short T2
relaxation times. Future studies will focus on the clinical application of this
new technique in patients with pulmonary disease.Acknowledgements
This work was supported by
National Natural Science Foundation of China (NO.61971151) and Wandong Medical.References
1. Wild
JM, Marshall H, Bock M, et al. MRI of the lung (1/3): methods. Insights Imaging.
2012; 3: 345–53.
2. Willmering
MM, Robison RK, Wang H, Pipe JG, Woods JC. Implementation of the FLORET UTE
sequence for lung imaging. Magn Reson Med. 2019; 82: 1091-1100.
3. Gai
ND, Malayeri A, Agarwal H, Evers R, Bluemke D (2016) Evaluation of optimized
breath-hold and free-breathing 3D ultrashort echo time contrast agent-free MRI
of the human lung. J Magn Reson Imaging 43:1230–1238.
4. Qian
Y, Boada FE. Acquisition-weighted stack of spirals for fast high-resolution
three-dimensional ultra-short echo time MR imaging. Magn Reson Med. 2008; 60: 135-145.
5. F.
Liu, J.V. Velikina, W.F. Block, R. Kijowski, A.A. Samsonov. Fast Realistic
MRI Simulations Based on Generalized Multi-Pool Exchange Tissue Model. IEEE
Transactions on Medical Imaging. 2016; 36(2): 527-537.