Noriyuki Tawara1 and Daiki Tamada2
1Department of Radiological Sciences, Faculty of Health Sciences, Japan Healthcare University, Sapporo, Japan, 2Department of Radiology, Yamanashi University, Yamanashi, Japan
Synopsis
Keywords: Visualization, Visualization
The objective of
this study was to reproduce the generation process of Steady-State and FLASH
Band in MRI phenomena that cannot be reproduced by the actual equipment because
of the restrictions imposed by venders, in order to directly confirm their
relationship with pulse sequences. MRI simulation can reproduce Bloch equation
faithfully, and thus it is possible to reproduce the relation between pulse
sequences and the phenomena related to MRI as numerical data.
Purpose
Assuming
ideal conditions, MR images similar to those of the actual scanners can be
reproduced computationally using the Bloch equation1, which is
widely a used simulation method. On the other hand, it has been considered
difficult to simulate realistic MR images within the practical computational time
since solving the equation is computationally expensive. However, recent advancement
in computational power enables complicated simulation of MR imaging.
MR images using Steady-State
can reduce the image scanning time, but the early imaging methods generated the
image artifact as FLASH Band2, which had to be controlled. Although
a solution for FLASH Band has already been established, it is not possible to
reproduce the mechanism in actual equipment due to the limitations of vendors.
The purpose of this study was to generate the Steady-State phenomenon (Steady-State)
and FLASH Band, which is challenging to reproduce with the actual equipment
due to constraints of its system interfaces, using an MRI simulation and to demonstrate
the relationship among Steady-State, FLASH Band and the pulse sequence.Methods
Two numerical phantoms with the matriz size of 256×256 were
prepared, and a) MR signal and b) MR image data were obtained by MRI
simulation. The phantom consists a) of 1-pixel with T1 of 1000 ms
and T2 of 100 ms placed at the center. The other phantom has for b) co-axial
cylindrical regions with different T1/T2(Figure 1). The proton density (PD) for
both phantoms was set at 1000. MRI simulations were implemented by BlochSolver3 (MRIsimulations Inc.,
Tokyo, Japan), which is the Bloch-Torrey equation-based MRI simulator. Two gradient echo sequences (GrE) a-1) without and a-2) with
spoiler gradient were used for the phantom a), as shown in the pulse sequence
diagrams in Figure 2. An RF phase increment angle of 117˚4 was used
for the RF pulses. All conditions except gradient and increment angle were
common and TR=40 ms, TE=6 ms, FA=30, the number of phase-encoding (PE)=256, and
slice thickness=5 mm. The number of subvoxels in the x, y, z directions were 32,
1, 1, respectively. Two GRASS sequences with b-1) without and b-2) with PE
rewinder, and b-3) spoiled GRASS (SPGR) were used, as shown in the pulse
sequence diagrams in Figure 3. All conditions except gradient and RF increment
angle were common and TR 25 ms, TE 10 ms, FA 30, and slice thickness 5 mm. The
number of subvoxels in the x, y, z directions were 8, 1, 8, respectively. The
calculations for MRI simulation were performed on a laptop PC (OS: Windows 10)
equipped with GPU (GeForce GTX 1650, 1024 CUDA core, NVIDIA, Santa Clara, CA). Data
processing acquired by MRI simulations was performed using Interactive Data
Language software (IDL; Harris Geospatial Solutions KK, Boulder, CO, USA).Results and Discussion
Figure 4 shows FID
signals (imaginary part) and MR signals with and without the addition of spoiler
gradient. Although the signal evolution does not completely reach Steady-State
as the number of excitation increases because the spoiler gradient is not used
in Figure 4a, FID signal is shown to be stable even when the RF number is
increased due to spoiler gradient loading in Figure 4b. Similarly for MR signal,
signal oscillation is observed in MR signal because the spoiler gradient is not
used in Figure 4c, but MR signal is shown to be stable in Figure 4d due to the
spoiler gradient. Figure 5a shows MR images acquired using GRASS without PE
rewinding. Since TR is shorter than T2 of the tissue in the phantom,
FLASH band occurs along the y-direction, which is PE direction. Figure 5b shows
MR images acquired using GRASS with PE rewinding. FLASH band artifacts is
improved, but they have T2-weighted contrast due to the influence of
the remaining transverse magnetization. Figure 5c shows MR images obtained by SPGR.
Transverse magnetization is completely spoiled by RF spoiling, and FLASH band
is also improved while retaining its characteristics as T1-weighted
image. For MR signals, we were able to reproduce the result consistent with
previous studies, which is Steady-State by setting phase angles of RF pulse and
spoiler gradient at random for each 117˚4.
In addition, achieving Steady-State using RF spoiling involves higher order
echoes generated by short TR RF pulses. The number of subvoxels must be high
because the higher order echoes can be correctly represented by adding up all
the isochromats in the pixel, and the correct MR signal must be obtained. Furthermore,
it was confirmed that Steady-State, which is widely utilized in MR imaging, can
be realized by using RF spoiling and gradient spoiling. When only gradient
spoiling is used, although Steady-State is reached, T2-weighted images are
obtained due to residual transverse magnetization. Therefore, it was confirmed
that to obtain T1-weighted images while reaching the Steady-State, RF spoiling
and gradient spoiling must be used together as in SPGR.Conclusion
Reproduction
regarding Steady-State and FLASH Band and the resolution of FLASH Band, and
confirmation of the relationship between those phenomena and pulse sequences
could be clarified by MRI simulation. This study suggested that there is a high
possibility that the relationship between pulse sequences and various phenomena
related to MRI can be clarified as numerical data by MRI simulation.
Acknowledgements
We
gratefully acknowledge Dr. Katsumi Kose (MRIsimulations Inc.) for his valuable advice.References
1. Bloch F. Nuclear
Induction. Phys Rev. 1946; 70(7-8): 460-474.
2. Crawley AP. Elimination of transverse
coherences in FLASH MRI. Magn Reson Med. 1988; 8(3): 248-260.
3. Kose R, Kose K. BlochSolver: A
GPU-optimized fast 3D MRI simulator for experimentally compatible pulse
sequences. J Magn Reson. 2017; 281: 51-65.
4. Zur Y, et al. Spoiling of
transverse magnetization in steady-state sequences. Magn Reson Med. 1991; 21(2):
251-263.