Teresa Nolte1, Masami Yoneyama2, Chiara Morsch1, Alexandra Barabasch1, Maximilian Schulze-Hagen1, Johannes M. Peeters3, Christiane Kuhl4, and Shuo Zhang5
1Diagnostic and Interventional Radiology, Uniklinik RWTH Aachen University, Aachen, Germany, 2Philips Japan, Tokyo, Japan, 3Philips Healthcare, Best, Netherlands, 4Diagnostic and Interventional Radiology, University Hospital RWTH Aachen, Aachen, Germany, 5Philips GmbH Market DACH, Hamburg, Germany
Synopsis
Keywords: Image Reconstruction, Diffusion/other diffusion imaging techniques, Noise
Further
acceleration of diffusion MRI in clinical examinations is desired but challenging
mainly due to low signal and associated potential bias in the quantitative
apparent diffusion coefficient
(ADC) values. Artificial intelligence-based denoising and image
reconstruction may provide a solution to address this challenge. We investigate
and compare different image reconstruction methods, including conventional
parallel imaging, compressed sensing, and a deep learning-based technique, in
ADC accuracy and precision using a diffusion phantom with illustration of the
principle in numeric simulation.
Introduction
Diffusion-weighted
MRI (DWI) plays an important role in clinical diagnosis, particularly in oncology.
Further acceleration of DWI is clinically desired but challenging, mainly due
to its low signal-to-noise ratio (SNR) especially at high b-values, and potential
bias in the apparent diffusion coefficient
(ADC) values associated with this. Typically, a sufficiently high number of
b-factor averages needs to be acquired to prevent noise-related bias and loss of
precision of ADC measurements1. Recently, deep learning-based
reconstruction techniques have emerged, promising further acceleration of
time-consuming sequences like DWI, while improving ADC robustness2,3.
The purpose of this simulation-corroborated phantom study was to validate a
deep-learning based image reconstruction method for DWI with comparison to
conventional acceleration techniques. Methods
To
investigate effects of noise on ADC bias and precision, nominal
diffusion-weighted signals
$$$M(b)=M_0·e^{-b·ADC}$$$ (Eq.1)
were
simulated for a range of nominal ADC values (ADCnom) between 0.3·10-3
and 2.1·10-3 mm²/s, setting b=(50,400,1000)s/mm² and M0=1. Noisy signals with
signal-to-noise ratios (SNR) between 2 and 20 were obtained by adding white
Gaussian noise of different standard deviation (SD), where SNR=M0/SD1. Per ADCnom and SNR level, 200 noisy signals were
fitted with Eq.1. Bias and precision were evaluated as mean and SD of the
fitted ADC values.
For the actual
phantom study, a diffusion phantom (CaliberMRI, Boulder, CO, USA) was measured
in a 3.0T clinical MRI system (3T Elition X, Philips, Best, The Netherlands). A
conventional single-shot echo-planar-imaging (EPI) DWI sequence with equidistant
k-space undersampling for acceleration was employed. Key scan parameters
include FOV=240mm, acq&rec matrix=128x128, slice thickness=1mm, FA=90°, TR=3500ms,
TE=64ms, b-values=50,400 and 1000 s/mm².
Measurements
were performed in two setups. Setup 1 investigated the effect of SNR reduction
on ADC. First, the numbers of b-factor averages for b=50-400-1000 s/mm2 were
reduced from 1-2-4 to 1-1-1. Corresponding scan durations were 1min17s and 35s.
Second, the acceleration factor was increased from 3 to 6, while the other
parameters were fixed with b-value averages 1-1-1. Conventional SENSitivity
Encoding (SENSE) reconstruction was employed.
Setup 2
investigated the effect of reconstruction technique on ADC. To that purpose, two
measurements were acquired with scan acceleration factors of 3 and 6,
respectively; again, b-value averages 1-1-1 and other parameters were unchanged.
Three reconstruction methods were applied, i.e., conventional SENSitivity
Encoding (SENSE), compressed SENSE (C-SENSE) and compressed SENSE Artificial Intelligence
(C-SENSE AI), all integrated to the vendor provided online reconstruction
architecture. While C-SENSE employed compressed sensing principle including
coil sensitivity information4,5, C-SENSE AI integrated a convolutional
neural network (CNN) into the C-SENSE reconstruction based on the recently
introduced Adaptive-CS-Net6, as a software prototype.
For both
setups, mean ADC and SD were measured by placing a circular ROI (1.43 cm²) per
phantom vial. Coefficient of variation was calculated as CV=mean/SD.Results
Figure 1
shows the simulation results for SNR values between 20 and 2. For the
investigated range of ADCnom, the fitted ADC values are plotted as mean ADCfit and SD. The simulations predict
increasingly negative bias, i.e., a deviation below ADCnom, and a
loss of precision, i.e., higher SD, towards higher ADCnom and
towards lower SNR.
ADC maps
obtained in setup 1 show a progressive increase of noise in the ADC maps that
is strongest in the center of the phantom (Figure
2) related to lower receiver coil sensitivity and g-factor based noise
enhancement. Regarding bias and precision of the measured ADC values (Figure 3), this translates into
increasingly negative bias and loss of precision from a SENSE acceleration of 4
onwards. The effect is stronger for higher ADCnom and for vials that
are situated in an inner or central location of the phantom.
ADC maps
obtained in setup 2 show low and high noise for SENSE reconstruction at
acceleration factors of 3 and 6, respectively (Figure 4). C-SENSE and C-SENSE AI recover the central ADC drop that
is present for the latter case. Qualitatively, SNR is increased by C-SENSE and
more so by C-SENSE AI, which makes phantom contours, but also aliasing
artefacts more visible. Regarding bias and precision of the measured ADC
values, bias is low for all reconstructions for an acceleration of 3 (Figure 5). At an acceleration of 6,
C-SENSE and C-SENSE AI mostly recover the negative bias present in SENSE reconstruction.
Moreover, these techniques lower the CV, which is brought back to a similar
range as for the lower acceleration factor.Discussion
In
low-SNR EPI-DWI scans, SENSE parallel imaging reconstruction at high
acceleration is prone to bias and loss of precision especially in areas with
lower receiver coil sensitivity and/or high g-factor, i.e., where the coil
elements have reduced overlap. C-SENSE and C-SENSE AI mitigate this problem and
enable faster DWI scanning; however, radiologists may need to adapt visual inspection
to the smoother appearance of ADC maps caused by reduced variation. While only
relative changes between different ADC measurements can be assessed in vivo, phantoms
with areas of homogeneous tissue and nominal ADC values thereof provide an
excellent means of testing accuracy and precision. Conclusion
Compressed
SENSE acceleration with deep-learning constrained reconstruction (C-SENSE AI)
reduces ADC bias in diffusion MRI with comparison to the conventional fast
imaging techniques. Improved accuracy and precision may benefit clinical
examinations and lead to further acceleration in routine practice. References
1 Walker-Samuel, et al. Robust Estimation of the Apparent Diffusion Coefficient (ADC) in Heterogeneous Solid Tumors. Magn Reson Med 2009; 62:420-429.
2 Yoneyama M, et al. SNR boost in whole-body DWIBS utilizing deep learning constrained Compressed SENSE reconstruction. ISMRM 2022, p3778.
3 Yoneyama M, et al. SNR enhancement in rapid high b-value prostate single-shot DW-EPI utilizing deep learning constrained Compressed SENSE reconstruction. ISMRM 2022, p4758.
4 Meister RL, et al. Compressed SENSE in Pediatric Brain Tumor: MR Imaging Assessment of Image Quality, Examination Time and Energy Release. Clin Neuroradiol 2022; 32(3):725-733.
5 Bode M, et al. Liver diffusion‑weighted MR imaging with L1‑regularized iterative sensitivity encoding reconstruction based on single‑shot echo‑planar imaging: initial clinical experience. Sci Reports 2022;12:12468.
6 Pezzotti N, et al. An Adaptive Intelligence Algorithm for Undersampled Knee MRI Reconstruction. IEEE Access. 2020;8:204825-204838.