Constantin Slioussarenko1, Pierre-Yves Baudin1, and Benjamin Marty1
1NMR Laboratory, Neuromuscular Investigation Center, Institute of Myology, Paris, France
Synopsis
Keywords: MR Fingerprinting/Synthetic MR, MR Fingerprinting
We
introduce an optimization framework using the longitudinal
steady-state equilibrium for FLASH MRF sequences, and taking into
account signal noise by using a Monte Carlo method. We use this
framework to optimize echo times, flip angles, timings and recovery time for T1
H2O and FF
mapping while shortening our original MRF T1-FF sequence. The
resulting Fast MRF T1-FF sequence yields comparable estimation
results to the original sequence with an almost twice reduced acquisition time.
Introduction
In the field of neuromuscular disorders (NMD), fat fraction (FF) is an established biomarker of disease severity and water T1 (T1H2O) has been shown to be a potential biomarker of disease activity [1]. The use of MR Fingerprinting (MRF) enables the simultaneous quantification of various parameters, such as FF and T1H2O [2]. However FLASH T1 MRF sequences (such as MRF T1-FF proposed in [3]) generally require long recovery times between repetitions for allowing the longitudinal magnetization to grow back to equilibrium. Shortening the sequence would highly benefit 3D imaging, where multiple acquisitions of the MRF scheme are required to encode along the partition encoding direction. However a shorter sequence is likely to degrade the parameter estimation quality. Hence we need to optimize the acquisition parameters of our shortened sequence for maintaining the quality of FF and T1H2O quantification.
In this work, we introduced an optimization framework that takes into account the longitudinal steady-state equilibrium of the MRF sequence and handles acquisition noise using Monte Carlo simulations [4]. Through this framework, we optimized the echo times (TE), flip angles (FA) and recovery time of the MRF T1-FF sequence and compared the mapping accuracy and precision to the original implementation.Methods
The
optimization framework consisted of 4 blocks : i) a simulation block
for steady-state MRF FLASH sequences with intial inversion pulse and
variable TE,TR and FA, ii) an MRF pattern matching algorithm, iii) a
cost function taking into account noise and iv) an optimizer (see Fig
[1]).
Regarding
the MRF FLASH simulation block, for N impulsions with flip angles
(θi)i=1..N
of phase (φi)i=1..N,
echo times (TEi)i=1..N
and repetitions times (TRi)i=1..N,
the magnetization has the following dynamics:
$$
M_x(t+TE_i)=M_z(t)sin(\theta_i)sin(\phi_i)e^{⁻\frac{TE_i}{T_2}}\\M_y(t+TE_i)=-M_z(t)sin(\theta_i)cos(\phi_i)e^{⁻\frac{TE_i}{T_2}}\\M_z(t+TR_i)=M_0+(M_z(t)cos(\theta_i)-M_0)e^{⁻\frac{TR_i}{T_1}}
$$
$$$m_p=M_z(p*T)$$$ represents the initial magnetization after p repetitions of the MRF scheme, where T is the total duration of the sequence.
We have :
$$m_p=Am_{p-1}+B$$
with $$$A=\prod_{i=1}^N{e^{⁻\frac{TR_i}{T_1}}cos(\theta_i)}$$$ and $$$B=M_0\sum_{i=0}^{N-1}{(1-e^{⁻\frac{TR_{N-i}}{T_1}})\prod_{j=N-i}^N{e^{⁻\frac{TR_j}{T_1}}cos(\theta_j)}}$$$.
$$$m_p$$$ is an arithmetic-geometric sequence with A<<1. Hence it reaches
a steady-state after very few repetitions of the sequence, and the
steady-state value is equal to B/(1-A).
MRF
is mainly impacted by noise due to undersampling that we approximated
with a Gaussian distribution representing an average SNR of 20.
For
pattern matching, we used exhaustive search into a bicomponent dictionary of fingerprints [5].
The
following cost function was calculated to assess parameters mapping
accuracy ([4])
:
$$C((\theta_i)_{i=1..N},(TE_i)_{i=1..N},TR_N)=\lambda_{T1}\frac{1}{N_s}\sum_{s=1}^{N_s}{\frac{|\widehat{T1_{H_2O}^s}-T1_{H_2O}^s|}{T1_{H_2O}^s}}+\lambda_{FF}\frac{1}{N_s}\sum_{s=1}^{N_s}|\widehat{FF^s}-FF^s|+\lambda_{T}\sum_{i=1}^{N}{TR_i}$$
where
$$$T1_{H_2O}^s,FF^s $$$(resp.$$$\widehat{T1_{H_2O}^s},\widehat{FF^s}$$$
) are the ground truth (resp. estimated) parameters for each
simulated signal and Ns is the total number of simulated signals.
We
fixed N=760 spokes, the smallest number of spokes for which
estimation did not worsen drastically.
As
using 760 spokes instead of 1400 spokes already reduced the sequence
duration, we used $$$ \lambda_{T}=0$$$. $$$\lambda_{T1}$$$
and $$$\lambda_{FF}$$$
were set empirically to 1 and 2 for those terms to have
similar order of magnitudes.
To
reduce the dimension of the problem, we assumed that TE and FA had
piecewise constant evolutions with respectively 3 and 5 pieces, and
that
(TRi)i=1..(N-1) were kept at their minimum possible value allowed by the gradient system
and bandwidth,
which left us with a total of 15 parameters to optimize for : Flip
angles values (θi)i=1..N ,
echo time values (TEi)i=1..N, breaks timings and recovery time TRN
at the end of each repetition.
The signals were simulated for the following grid of tissue parameters : T1H2O
=500:100:1900 (ms), FF
= 0.:0.1:0.5, df
= [-30.,0,30] (Hz), B1=[0.7,1.0].
The
algorithm used for minimizing the cost function was differential
evolution [6], which is an efficient global optimization method when the gradient of the cost function is not easily accessible.
The maximum functions evaluations was set at 150000 but the algorithm
stopped earlier.
We
compared the reference (MRF T1-FF) and optimal (Fast MRF
T1-FF) sequences on 3D numerical phantoms
(16 slices – 64 square subregions of size 4x4 per slice with randomly varying
T1H2O,
FF, Df and B1) and 3D in
vivo
data acquired at 3T (PrismaFit, Siemens Healthineers) on the thighs of one healthy control (20
partitions, resolution 1x1x5 mm3,
FOV 10x40x40cm3,10 regions of interest (ROIs) drawn on the left thigh muscles).Results
Each
repetition of Fast MRF T1-FF lasts 6.5s against 10.9s for MRF T1-FF
(Fig [2]). The recovery of 3s at the end of each repetition did not
allow for the longitudinal magnetization to grow back to intial
equilibrium, however the steady-state was reached after a maximum of two repetitions of the Fast MRF T1-FF sequence (Fig [3]).
On numerical phantoms, T1H2O
and FF estimation accuracy and precision were similar to the reference sequence (Figure 4).
On
in
vivo
data, accuracy was comparable to MRF T1-FF. Precision was degraded for T1H2O, with a 40 ms increase in standard deviation on average (Figure 5).
Discussion & Conclusion
Using
a new optimization framework taking into account steady-state
magnetization and noise, we designed a new sequence (Fast T1-FF MRF) for fast quantification of T1H2O
and FF, with reduced time and comparable results to those of the reference sequence. In vivo, the
new sequence was less precise, however this
could be alleviated by MRF denoising techniques ([7]).
The explicit calculation of the steady-state value removed the need to simulate several repetitions of the MRF sequence and allowed to reduce the computation time for sequence optimization.Acknowledgements
This study was funded by ANR-20-CE19-0004.References
[1] Marty, B., Reyngoudt, H., Boisserie, J. M., le Louër, J., Araujo, E. C. A., Fromes, Y., & Carlier, P. G. (2021). Water-fat separation in mr fingerprinting for quantitative monitoring of the skeletal muscle in neuromuscular disorders. Radiology, 300(3), 652–660. https://doi.org/10.1148/radiol.2021204028
[2] Ma, D., Gulani, V., Seiberlich, N., Liu, K., Sunshine, J. L., Duerk, J. L., & Griswold, M. A. (2013). Magnetic resonance fingerprinting. Nature, 495(7440), 187–192. https://doi.org/10.1038/nature11971
[3] Marty, B., Lopez Kolkovsky, A. L., Araujo, E. C. A., & Reyngoudt, H. (2021). Quantitative Skeletal Muscle Imaging Using 3D MR Fingerprinting With Water and Fat Separation. Journal of Magnetic Resonance Imaging, 53(5), 1529–1538. https://doi.org/10.1002/jmri.27381
[4] Karsten S. et al. (2016). Towards judging the encoding capability of Magnetic Resonance Fingerprinting sequences. ISMRM Proc 2016
[5] Slioussarenko, C., Baudin ,P.Y. , Reyngoudt, H. & Marty, B (2022). Bi-component dictionary for efficient quantification of fat fraction and water T1 via MR fingerprinting. ISMRM Proc 2022
[6] Storn, R., & Price, K. (1997). Differential Evolution-A Simple and Efficient Heuristic for Global Optimization over Continuous Spaces. Journal of Global Optimization, 11, 341–359.
[7] Mazor, G., Weizman, L., Tal, A., & Eldar, Y. C. (2018). Low-rank magnetic resonance fingerprinting. https://doi.org/10.1002/mp.13078 p { margin-bottom: 0.25cm; direction: ltr; line-height: 115%; text-align: left; orphans: 2; widows: 2; background: transparent }a:link { color: #0563c1; text-decoration: underline }