John Neri1, Matthew Koff1, Kevin Koch2, and Ek Tsoon Tan1
1Hospital for Special Surgery, New York, NY, United States, 2Medical College of Wisconsin, Milwaukee, WI, United States
Synopsis
Keywords: Quantitative Imaging, Diffusion/other diffusion imaging techniques
Multi-acquisition with variable resonance image combination (MAVRIC) with
DWI can reduce artifacts associated with metal implants to characterize
synovial reactions near total hip replacements. The purpose of this study was to
evaluate the quantitative accuracy of DWI-MAVRIC near actual hip arthroplasty
components and demonstrate the feasibility of spatially mapping ADC biases. DWI-EPI and DWI-MAVRIC images were acquired using a diffusion phantom in the presence
of femoral heads of different material composition. It was found that
cobalt-chromium femoral heads cause more ADC error than oxinium femoral heads. Spatial
mapping of ADC accuracy near total hip implants was found to be feasible.
Introduction
Diffusion weighted imaging (DWI) yields quantitative
apparent diffusion coefficient (ADC) measurements which are useful in assessing
soft tissues1. Multi-acquisition with variable
resonance image combination (MAVRIC) is a novel imaging technique that reduces artifact
due to metal implants2. Two-dimensional MAVRIC3 combined with periodically rotated
overlapping parallel lines with enhanced reconstruction (PROPELLER) FSE for DWI4 has been used to characterize synovial
reactions near total hip arthroplasties5 (THA)(Figure 1). Previous work evaluating
quantitative accuracy of this DWI-MAVRIC acquisition near metal uncovered
biases of ADC values, attributable to the multi-shot FSE acquisition6, which would impede interpretation of
diffusivity values. Therefore, the goal of this work was to evaluate the
quantitative accuracy of DWI-MAVRIC near total hip implant components and demonstrate the feasibility of spatially mapping ADC biases.Methods
We measured known diffusivity values (DC)
of an ADC phantom at different zenith angular (θ:[0o,90o])
and radial (ρ:[ρmin, ρmax])
positions relative to a femoral head (Figure 2) to map the accuracy of the DWI-MAVRIC
sequence. Different femoral heads of varying material compositions were placed
at a known position and orientation relative to B0.This analysis assumed
DC was dependent on susceptibility effects due to B0
and were hence azimuthally-invariant and symmetric about the equator. Extrapolation
of DC to (θ:[0o,360o]) and
interpolation between unsampled positions were performed by sparse
interpolation. Positions that could not be assessed due to signal dropout were excluded.
ADC phantom scanning was performed using a Diffusion Phantom
Model 128 (CaliberMRI, Boulder, CO). To enable scanning at room temperature and
real-time temperature estimation, the phantom was fitted with liquid-crystal
vials with MR signal that displays temperature-sensitive MR signal changes7. For this work, two of the 13 vials of
polyvinylpyrrolidone (PVP) at 10% wt/wt PVP with diffusivity values closest to soft
tissue (1614 µm2/s at 20.5°C) 8 were considered. All images were acquired at
iso-center on a clinical 1.5T MRI (MR450, GE Healthcare) with an 8-channel
cardiac coil. To assess the effect of orthopedic hardware on ADC accuracy, femoral
heads of different sizes and material composition were affixed to the phantom
collinear to the two 10% PVP vials (Figure 2). Next, DWI-EPI (NEX= 3) and
DWI-MAVRIC (b=600, 3 spectral bins) images were acquired for the following
configurations: No metal (NM); 36mm cobalt-chromium (CC) head; 28mm oxidized
zirconium head (oxinium [OX]); and 36mm OX. For the configurations with a
component present, images were acquired with the phantom rotated 22.5°,
45°,
67.5°,
and 90° clockwise
in the coronal plane relative to B0 (Figure 2B-E). Internal phantom water
temperature was monitored using an SPGR sequence to image liquid crystal vials.
Images were processed using in-house software (MATLAB, Mathworks, Natick, MA)
to generate ADC maps with gradient nonlinearity correction (GNC)9. ADC values were acquired by placing two
circular regions-of-interest (44 voxels, 121 mm2) on each vial using
ITK-SNAP10. The actual phantom θ was
manually measured from the images, and ρ was
measured at 29, 41, 62 and 75 mm from the surface of the implant6. ADC error was calculated using theoretical
ADC at the phantom water temperature8 to generate error heat maps of eC=2*(DC
– Dtheoretical)/(DC + Dtheoretical),
showing error eMetal as a function of (ρ,θ)
relative to the implant surface. To
offset for effects due to the EPI or PROPELLER pulse sequences, adjusted error
(eMetal–eNM) heat maps were also generated,
using in-house software (MATLAB, Mathworks, Natick, MA).Results
Figure 3 shows ADC maps in all four scanning configurations,
displaying severe image distortion with the EPI acquisition and CC femoral head.
DWI-MAVRIC shows its largest bias patterns at θ=0o
and with CC. The DWI-EPI eNM bias was small and negative (~ 2%),
while DWI-MAVRIC eNM showed a strong positive bias(~3-6%). Offsetting
by eNM , the greatest errors eCC–eNM
occurred at the angles of 0o and 180o for both DWI-EPI
and DWI-MAVRIC, followed by 90o and 270o for both
DWI-EPI and DWI-MAVRIC (Figure 4). Overall, DWI-MAVRIC acquisitions showed less
error as compared to DWI-EPI. CC implants caused higher eMetal–eNM
spatial variation than OX. The 28 mm OX showed less error than 36 mm OX for
DWI-EPI, but produced a different error pattern in DWI-MAVRIC. Discussion
This study demonstrated the feasibility for mapping diffusivity
bias patterns using the DWI-MAVRIC sequence in the presence of different femoral
head components commonly used in total hip arthroplasty. We observed a positive
DWI-MAVRIC bias similar to previous work6,11 with
bilobular bloom patterns that conform to well-known B0
susceptibility patterns6. These
results suggest that distance from the metal implant, zenith angle relative to the
metal, and material composition of the implant should be considered when evaluating
ADC values in close proximity to implanted metallic devices. In this
preliminary demonstration, we limited the measurements to only five zenith angles
and four distances, which may be increased to improve data interpolation of the
heat maps. ADC data also could not be
acquired near the surface of the implant which prevents conclusions from being
drawn about ADC accuracy immediately adjacent to implanted hardware. Conclusion
DWI-MAVRIC has less ADC error than DWI-EPI when
imaging in the presence of hip implants. Cobalt-chromium femoral heads cause
more ADC error compared to oxinium femoral heads. Spatial mapping of ADC
accuracy near total hip implants was found to be feasible.Acknowledgements
The authors would like to thank the MRI administrative staff and MRI
technologists at Hospital for Special Surgery for their assistance in acquiring
images. Research reported in
this publication was supported by National Institute of Arthritis and
Musculoskeletal and Skin Diseases of the National Institutes of Health under
award number R01AR064840 and National Institute of Biomedical Imaging and
Bioengineering of the National Institutes of Health R21EB023415 and R21EB030123. This content is solely the
responsibility of the authors and does not necessarily represent the official
views of the NIH.References
1.Bihan D le. Apparent
diffusion coefficient and beyond : What diffusion mr imaging can tell us about
tissue structure. Radiology. 2013;268(2):318-322.
doi:10.1148/radiol.13130420
2. Koch KM, Lorbiecki JE, Hinks RS, King KF. A multispectral
three-dimensional acquisition technique for imaging near metal implants. Magn
Reson Med. 2009;61(2):381-390. doi:10.1002/mrm.21856
3. Hargreaves BA, Taviani V, Litwiller D v, Yoon D. 2D
multi-spectral imaging for fast MRI near metal. Magn Reson Med.
2018;79(2):968-973. doi:https://doi.org/10.1002/mrm.26724
4. Koch KM, Bhave S, Gaddipati A, et al. Multispectral
diffusion-weighted imaging near metal implants. Magn Reson Med.
2018;79(2):987-993. doi:10.1002/mrm.26737
5. Gao MA, Neri JP, Lin B, et al. Diffusion MRI of THAs for
the Classification of Synovial Reactions. In: ISMRM. 2021.
6. Neri J, Koff M, Koch K, Tan E. Quantitative Assessment of
Diffusion Weighted Imaging Near Metal Implants. In: ISMRM. 2022.
7. Keenan KE, Stupic KF, Russek SE, Mirowski E. MRI-visible
liquid crystal thermometer. Magn Reson Med. 2020;84(3):1552-1563.
doi:10.1002/mrm.28224
8. Wagner F, Laun FB, Kuder TA, et al. Temperature and
concentration calibration of aqueous polyvinylpyrrolidone (PVP) solutions for
isotropic diffusion MRI phantoms. PLoS One. 2017;12(6):1-13.
doi:10.1371/journal.pone.0179276
9. Tan ET, Marinelli L, Slavens ZW, King KF, Hardy CJ.
Improved correction for gradient nonlinearity effects in diffusion-weighted
imaging. J Magn Reson Imaging. 2013;38(2):448-453.doi:10.1002/jmri.23942
10. Yushkevich PA, Piven J, Hazlett HC, et al. User-guided 3D
active contour segmentation of anatomical structures: significantly
improved efficiency and reliability. Neuroimage. 2006;31(3):1116-1128.
doi:10.1016/j.neuroimage.2006.01.015
11. Neri JP, Koff MF, Koch KM, Tan ET. Validating the accuracy of
multispectral metal artifact suppressed diffusion-weighted imaging. Med Phys.
Published online October 1, 2022. doi:10.1002/mp.15925