Ramesh Paudyal1, Akash Deelip Shah2, Amaresha Shridhar Konar1, Victoria Yu1, Dariya I Malyarenko3, Nancy Lee4, Thomas L Chenevert3, and Amita Shukla-Dave1,2
1Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 2Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 3Radiology, University of Michigan, Ann Arbor, MI, United States, 4Radiation Oncology, Memorial Sloan Kettering Cancer Center, New York, NY, United States
Synopsis
Keywords: Quantitative Imaging, Tumor
The present study evaluated the correction efficacy
for ADC values bias due to the gradient nonlinearity of the MRI system on primary
tumors, neck nodal metastases, and masseter muscle in the head and neck region
using a vendor-provided LOw VAriance (LOVA) ADC technique. The LOVA ADC technique
was developed and implemented to compensate for gradient linearity errors that
may allow consistent ADC values in large FOV independent of the offset from the
scanner isocenter. Our results confirmed the potential of the LOVA technique to
improve ADC accuracy independent of location and scanner gradient
characteristics.
Purpose
Diffusion-weighted (DW)-MRI-derived apparent diffusion coefficient (ADC),
a surrogate of tumor cellularity, has shown promise in evaluating treatment
response in metastatic nodes of head and neck cancer.1,2 Previous studies have demonstrated
that the apparent diffusion coefficient (ADC) measurement is biased by
spatially-dependent b-value due to gradient nonlinearity3,4, particularly for anatomy
with increasing offset from the MRI scanner isocenter. It has been shown that there
is an improvement in the accuracy of ADC measures by applying gradient
nonlinearity (GNL) correction, for example, for breast DWI.5,6 Recently, the GNL correction method
has been implemented on clinical scanners via an academic-industry partnership with
the three major MRI vendors. In the present study, we aimed to use the vendor-provided
LOw VAriance (LOVA) ADC technique for GNL
correction to improve the accuracy of the ADC values measured for primary
tumors, neck nodal metastases, and masseter muscles in patients with head and
neck cancer. Methods
MRI data
acquisition:
Phantom: DW-MRI study
was performed on the NIST/QIBA diffusion phantom at a 3T MRI scanner (Elition,
Philips Healthcare, Netherlands) using a 16-channel head coil.7 DWI
images of the phantom were acquired using a single shot spin echo planar
imaging (SS-SE-EPI) sequence with 4 b-values (i.e., b= 0, 500, 900, 2000 s/mm2)
and the following parameters: repetition time (TR)=15000 ms, echo time (TE) =
minimum (99 ms), number of averages (NA)=1, acquisition matrix=128×128, the field
of view (FOV) = 220 mm2, number of slices (NS)=20, slice thickness =
4 mm, all 3 orthogonal directions. The total acquisition time for the multiple
b-value DWI data acquisition was ~2 minutes 30 seconds. For real-time
processing, the LOVA ADC technique was selected during acquisition, which
generated two DW- datasets with and without GNL correction. On both ADC maps, a
186 mm2 region of interest was placed in each vial using Image J
software. 8 The three
vials' ( #Oc, #10o, and #50o ) ADC values were reported
herein.
Patient: DW-MRI
data was acquired from fifteen
HPV positive (+) oropharyngeal squamous cell carcinoma (OPSCC) patients (median
age 54 years, 15 male) in this retrospective study between December 2021 and
September 2022. The patients were treated with chemoradiation therapy (CRT). MRI
protocol consisted of multi-planar T1/T2 weighted imaging
followed by multi-b-value DWI on a 3.0T scanner (Elition, Philips Healthcare,
Netherlands) using a neurovascular phased-array coil at pretreatment (Tx). The
multi b-value DW images were acquired using a single shot spin echo planar imaging
(SS-SE-EPI) sequence with TR/TE=4000/80(minimum) ms, FOV=20-24
cm, matrix=128×128, slices=8-10, slice thickness=5mm, number of excitation
(NEX)=2, and b=0,300,800 s/mm2.
Regions of Interest (ROIs): A neuro-radiologist delineated regions of Interest
(ROIs) on primary tumors, neck nodal metastases where tumors lesion were seen, and
masseter muscle on the DW image (b = 0 s/mm2) using ITK-SNAP. All DW
data analysis was performed using in-house software MRI-QAMPER (Quantitative
Analysis Multi-Parametric Evaluation Routines) written in MATLAB (MathWorks,
Natick, MA). ADC values were calculated for b = 0, 300, and 800 s/mm2, and maps were generated using a mono-exponential
diffusion model. The relative percentage change in mean ADC values (rΔADC (%) ) with and without GNL correction
was calculated as follows: rΔADC(%) = (ADCGNL-ADC)/ADCGNL×100,
where ADCGNL and ADC represent ADC values measured with and without
GNL correction, respectively. Mean ADC values with and without the GNL correction were compared using a
Wilcoxon signed rank test, and P <0.05 was considered significant. Results
Phantom: Mean ADC
values with GNL correction for three representative vials # 0c, # 10o,
and # 50o were 1.68% (1.067±0.016 vs. 1.085±0.016 ×10-3
[mm2/s]),4.39% (0.820±0.007 vs. 0.856±0.007×10-3 [mm2/s]),
and 5.11% (0.137±0.008 vs. 0.144±0.009×10-3 [mm2/s]) which
were lower than those without GNL correction (Figure 1).
Patient: The study evaluated 8 primary lesions, 15 metastatic nodes, and
8 masseter muscles from 15 OPSCC patients. In Figure 2, the boxplot compares
mean ADC values with and without GNL correction for primary tumors, metastatic
nodes, and master muscle from OPSCC patients. GNL-corrected means ADCs values
of primary tumors, metastatic nodes, and master muscles were lower than the
uncorrected ADCs, consistent with the phantom study (Table 1). Figure 3 shows
the ADC parametric maps generated with and without GNL correction. Figure 4 shows the representative histogram plot for ADC values
obtained with and without GNL correction from metastatic nodes. The impact of
GNL correction on narrowing the ADC values can be appreciated.Discussion
Mean ADC value with and
without GNL correction differed by 5.74%, 3.98%, and 6.11% for primary,
metastatic nodes, and master muscle, respectively. The results suggest that
implementing GNL correction improves ADC measurement accuracy and reduces
unnecessary variability due to different offsets from the isocenter. GNL
correction needs to be implemented across all scanner platforms to ensure
uniformity and consistency of diagnostic ADC measures, particularly if
considering the incorporation of ADC as a quantitative imaging biomarker in
standardized multiplatform clinical studies.Conclusion
This study showed that
implementing GNL correction significantly improves the accuracy of ADC measures,
enhancing the robustness of ADC thresholds as a quantitative imaging biomarker
used in clinical trials.Acknowledgements
Support: Funding
support from National Institutes of Health Grants: U01CA166104, U24CA237683, U01
CA211205, R01CA190299, and 75N91021C00036References
1. Paudyal R, Oh
JH, Riaz N, et al. Intravoxel incoherent motion diffusion-weighted MRI during
chemoradiation therapy to characterize and monitor treatment response in human
papillomavirus head and neck squamous cell carcinoma. J Magn Reson Imaging. Apr 2017;45(4):1013-1023.
doi:10.1002/jmri.25523
2. Riaz N, Sherman
E, Pei X, et al. Precision Radiotherapy: Reduction in Radiation for
Oropharyngeal Cancer in the 30 ROC Trial. JNCI:
Journal of the National Cancer Institute. 2021;113(6):742-751.
doi:10.1093/jnci/djaa184
3. Bammer R, Markl
M, Barnett A, et al. Analysis and generalized correction of the effect of
spatial gradient field distortions in diffusion‐weighted imaging. Magnetic Resonance in Medicine: An Official
Journal of the International Society for Magnetic Resonance in Medicine.
2003;50(3):560-569.
4. Malyarenko DI,
Ross BD, Chenevert TL. Analysis and correction of gradient nonlinearity bias in
apparent diffusion coefficient measurements. Magnetic resonance in medicine. 2014;71(3):1312-1323.
5. Malyarenko DI,
Newitt DC, Amouzandeh G, et al. Retrospective correction of ADC for gradient
nonlinearity errors in multicenter breast DWI trials: ACRIN6698 multiplatform
feasibility study. Tomography.
2020;6(2):86-92.
6. Newitt DC, Tan
ET, Wilmes LJ, et al. Gradient nonlinearity correction to improve apparent
diffusion coefficient accuracy and standardization in the american college of
radiology imaging network 6698 breast cancer trial. J Magn Reson Imaging. Oct 2015;42(4):908-19. doi:10.1002/jmri.24883
7. Paudyal R,
Konar AS, Obuchowski NA, et al. Repeatability of Quantitative
Diffusion-Weighted Imaging Metrics in Phantoms, Head-and-Neck and Thyroid
Cancers: Preliminary Findings. Tomography.
Mar 2019;5(1):15-25. doi:10.18383/j.tom.2018.00044
8. Amouzandeh G, Chenevert TL, Swanson
SD, Ross BD, Malyarenko DI. Temperature and concentration dependence of water
diffusion in polyvinylpyrrolidone solutions. Medical Physics. 2022;49(5):3325-3332.