Lana Bautz1, Oluwabusayo A. Oni2, Hivnu Toraman3, Olav Jansen3, Jan-Bernd Hövener1, Naomi Larsen3, and Mariya S. Pravdivtseva1
1Department of Radiology and Neuroradiology, Section Biomedical Imaging, Molecular Imaging North Competence Center (MOIN CC), University Medical Center Schleswig-Holstein (UKSH), Kiel University, Kiel, Germany, Kiel, Germany, 2University of Florida, Gainesville, FL, United States, 3Department of Radiology and Neuroradiology, University Medical Center Schleswig-Holstein (UKSH), Kiel, Kiel, Germany
Synopsis
Keywords: Data Analysis, Treatment, Evaluation of aneurysm treatment
Flow diverter stents are effective for the treatment of intracranial
aneurysms at a high risk of rupture. Still,
complications may occur after stent implantation, so evaluation of aneurysm
treatment is required. Treatment follow-up can be performed with 4D phase-contrast
MRI. In this work, flow within the aneurysm and adjacent vessels with and
without stent was measured with 4D flow. Strong metal artifacts were observed
at the parent vessel which resulted in MRI signal voids and a 30 % reduction of
measured flow. No artifacts were seen within an aneurysm sac, here flow was
reduced up to 83 %.
Introduction
Intracranial
aneurysms (IAs) can cause life-threatening hemorrhages if they rupture and
therefore require preventive treatment. Aneurysms could be treated by placing flow
diverter stents (FDS) to reduce the blood flow into the aneurysm sac. But
postoperative complications, such as the persistence of flow inside the
aneurysm, can still occur, and even result in a delayed aneurysm
rupture1.
Four-dimensional (4D) phase contrast magnetic resonance imaging (PC MRI) can provide 3D directional time-resolved
comprehensive velocity maps non-invasively. Therefore, it can assess the aneurysm
treatment successes by measuring flow changes . However, 4D PC MRI can be affected by metal artifacts,
making treatment evaluation challenging1.
To assess this matter, this work aims to evaluate flow at the aneurysm and adjacent vessels with and
without FDS. The flow was measured with 4D PC-MRI using a patient-specific aneurysm
model.Methods
Aneurysm
treatment: A
commercially available stent (Derivo) was placed in patient-specific aneurysm
model fabricated in-house2. Aneurysm model with and without (w/o) a
stent was supplied with an average flow of 250 ml/min and underwent MRI
measurements (glycerol-water mixture, PD-1100).
MRI measurements were performed on a 3 T MR system equipped with a 32-ch
head coil. Time-resolved T1-weighted PC MRI with 3D velocity encoding was
acquired (TE/TR: 5/8.4 ms; FOV: 110 × 110 × 50 mm3, iso voxel size
0.75 mm3). The velocity encoding parameter (venc) was set at 80 and
100 cm/s for the aneurysm model w/o and with stent respectively.
Data analysis: Flow and velocity values were evaluated in the aneurysm,
parent vessel before the stent (Fig. 1a, section 1) and within the stent
(section 2) (GTflow, version 3.2.13). For qualitative evaluation, velocity path
lines in the aneurysm were generated.
Then, 2D quantitative evaluation of flow and velocity was performed as follows:
-
Equidistant
slices were created perpendicular to the vessel and within the aneurysm (Fig.
1a);
- The flow was
measured in selected areas in each slice (Fig. 1b)
- The
resulting flow was averaged across the slices and time (Fig. 1c);
- The maximum,
mean, and standard deviation was calculated. The flow reduction (FR)
was calculated as
$$\text{FR} = 1-\frac{Q_{\text{FDS}}}{Q_{\text{w/o}}}.$$
To
further investigate velocity reduction within the aneurysm, a volumetric region
(3D ROI) of the aneurysm was created (Fig. 4c, Meshmixer).
Results
Evaluation of flow: At the parent vessel in section 1,
flow with and w/o a stent was similar, 7.7 and 7.8 ml/s respectively (Fig. 2a
and b). A correlation coefficient of 0.92 was calculated between the flow
values and the linear fit has a slope of 0.98 (Fig. 2c). However, in section 2,
flow with stent was smaller than w/o stent, 7.5 and 5.4 ml/s (Fig. 2d and e). The
correlation coefficient had a value of 0.91, but the slope of linear fit decreases
to 0.65 (Fig. 2f). After implanation, the maximum flow was reduced by 28% in
section 2 but stayed constant in section 1. This reduction is clearly a
non-physical effect and likely attributed to the metal-induced artefacts of the
FDS.
Within the aneurysma, the flow w/o stent was
reduced as a function of distance from the parent vessel (from 7.7 to 3.5 ml/s
at the aneurysm center and external wall, respectively, Fig. 2g). The FDS
reduced the intraaneurysmal flow drastically, from 11.2 to 1.8 ml/s (Fig. 2h),
and the maximum flow by 84 % (Tab. 1). The mean values support the observed
trend in flow reduction within section 2 and aneurysm.
Evaluation of velocities: A strong vortex and flow pulsatiliy
in the aneurysm (Figs. 3a,c) vanished when the stent was placed (Fig. 3b,d).
The mean velocity in the aneurysm was reduced from 26.9 to 10.0 cm/s
(Tab. 1).
Evaluation
of the aneurysm volume as a whole showed a similar reduction in flow, although lower
maximum values (from 20 to 5.9 cm/s, a reduction of 71 %, Fig. 4a and b).Discussion
The
reduction of flow inside the aneurysm of 83% after implantation was similar but
not the same as described by Velvaluri et al.3, where a flow
reduction of 97% was found in with the same stent.
Pereira et al.1
described a velocity reduction, which ranged from 34.6% to 71.1%. This is in
good agreement with the values found here.
Differences
between studies may be caused by different MRI parameters, which have been
shown to affect flow and velocity values. In addition, the choice of the
position of a 2D slice within an aneurysm is crucial for flow calculation.
Velocity calculation within the slice also differs from volumetric velocity
analysis. Standardized MRI protocols and a full description of the quantitative
analysis are needed to allow comparison.
Our study
has several limitations: 1) one aneurysm geometry and one stent model were
studied; 2) venc must be reduced to eliminate velocity noise. Next, we repeat
4D PC MRI with lower venc, add different aneurysm geometries from different
stent manufacturers. Finally, we will determine whether all commercial stents
cause the same in-stent flow reduction due to metal artifacts.Conclusion
4D PC MRI
enables non invasive monitoring the FDS effect on the flow in aneurysms,
despite the fact that metal artefacts occur within the stent. The data
analysis is complex, affects the outcome and must be considered carefully.Acknowledgements
No acknowledgement found.References
1. Pereira VM, Brina
O, Delattre BMA, et al. J NeuroIntervent Surg. 2014; 0:1-7. doi: 10.1136/neurintsurg-2014-011348
2. Pravdivtseva MS, Peschke E, Thomas Lindner et al. Med Phys. 2021; 48:1469-1484. doi: 10.1002/mp.14714
3. Velvaluri P, Pravdivtseva MS, Berg P et
al. AMT 2021; 6:1-13. doi: 10.1002/admt.202100384