Yuguang Meng1, James J. Lah2, Jason W. Allen1,2, and Deqiang Qiu1
1Department of Radiology and Imaging Sciences, Emory University, Atlanta, GA, United States, 2Department of Neurology, Emory University, Atlanta, GA, United States
Synopsis
Keywords: Data Acquisition, Artifacts
In
gradient-echo imaging, the motion effects on the reconstructed images could be
exacerbated by the temporal B
0 field changes due to respiration and
the subjects’ unintentional position/posture changes during scanning. An
efficient 3D spiral navigator independent of the multi-echo gradient-echo (mGRE)
acquisition train was designed and implemented in a 3D mGRE sequence. The
results showed that although there were unintentional movements during
acquisitions, the temporal B
0 changes could be significant at 3.0 T.
The design provides flexible TE options for mGRE in obtaining simultaneous T
1-weighted
and T
2*-weighted contrasts, quantitative T
2*
and/or quantitative susceptibility mapping.
Introduction
Motion
artifacts are common in MRI scans due to unintentional or noncooperative movements
during MRI data acquisitions 1,2. In gradient-echo imaging, the motion
effects on the reconstructed images could be exacerbated by the temporal B0
field changes due to unavoidable respiration and the subjects’ unintentional position
and/or posture changes during scanning 1. In this
work, an efficient 3D spiral navigator to detect temporal motions and B0
shifts was designed and implemented in a 3D multi-echo gradient-echo (mGRE) sequence.Methods
In vivo 3D human brain scans were performed with
written consent under the approval from the University Institutional Review
Board. The 3D mGRE sequence with navigators for motion and
B0 shift tracking was
implemented on a Siemens 3T MRI scanner (Prisma model, Siemens Healthcare) with
32 head receive coils and IDEA software version VE11C. As seen in Fig. 1, at
the beginning of the sequence, an under-sampled 3D spiral navigator was
inserted with the parameters: one interleaf, matrix size = 48×48×32, isotropic
resolution = 4.5 mm, acceleration factor R = 2 in the slice direction, dual echoes
with TEs = 1.6 and 20 ms. In subsequent excitations, each under-sampled spiral
navigator was interleaved with 26 repetitions of (i.e., 26 TRs) Cartesian GRE acquisitions
with 5 echoes, first TE = 4 ms and echo-spacing = 7 ms, matrix size =
320×320×104, acceleration factor R = 4 in the in-plane phase-encoding
direction, resolution = 0.7 mm × 0.7 mm × 1.4 mm. The same RF flip angle of 15o
and TR of 37 ms were used for both the navigators and the mGRE excitations to
maintain a steady state of the signal. The scanning time was 12 minutes. Following
2D gridding in the in-plane directions, the 3D spiral navigators were
reconstructed using GRAPPA in the slice direction. Dual-echo data for each
receive coil was used to remove the phase offsets in the navigator to obtain
the B0 off-resonance field 3.
The rigid-body translations and rotations relative to the initial position/ posture were measured with the “flirt” script in
FSL toolbox (FMRIB, Oxford), and the B0 shifts were measured from
the phase images of the navigators.Results
As
shown in Fig. 2a-b, gradient-echo image with long TE produced more significant
artifacts than with short TE as indicated by white arrows. The translations and
rotations were measured along with the temporal B0 shifts on a
subject (Fig. 2c-d). While there were unintentional movements during
acquisitions for the subject as shown in the measured minor translations
(<0.2 mm) and rotations (<0.2o), the temporal B0
changes were more significant with a standard deviation around 50 Hz. Ongoing
work is being performed to incorporate these motion and B0 estimates
in the reconstruction of mGRE images to minimize artifacts.Discussion
This work developed a dual-echo 3D
navigators for simultaneous motion and B0 tracking in mGRE sequence.
Since the temporal B0 change is inevitable due to respirations 2,
incorporating dynamic B0 corrections along with motion corrections
during imaging reconstructions is important for MRI scans at 3.0T. It is worth
noting that the applied navigators are independent of the mGRE acquisition
train, thus providing flexible TE options for mGRE in obtaining simultaneous T1-weighted
and T2*-weighted contrasts, quantitative T2*
and/or quantitative
susceptibility mapping (QSM). Given the efficient motion tracking
and B0 measurements, such MR contrasts or quantitative measurements can benefit from various motion/
B0
correction methods to minimize artefacts.Acknowledgements
National Institutes of Health Grants: P30AG066511, R01AG072603 and
R21AG064405.References
1.
Liu Jiaen, Gelderen Peter van, Zwart
Jacco A de, et al. Reducing motion sensitivity in 3D high-resolution T2*-weighted
MRI by navigator-based motion and nonlinear magnetic field correction.
Neuroimage. 2020;206:116332.
2. Gretsch Frédéric, Marques José P, Gallichan
Daniel. Investigating the accuracy of FatNav-derived estimates of temporal B0
changes and their application to retrospective correction of high-resolution 3D
GRE of the human brain at 7T. Magn Reson
Med. 2018;80(2):585-597.
3. Robinson
Simon, Jovicich Jorge. B0 mapping with multi-channel RF coils at
high field. Magn Reson Med. 2011;66(4):976-988.