Jason Daniel van Schoor1,2, Mark Gosselink3, Dennis Klomp3, Giel Mens3, Hans Hoogduin3, Wim Prins4, and Tijl van der Velden3
1High Field MRI, UMC Utrecht, Utrecht, Netherlands, 2Utrecht University, Utrecht, Netherlands, 3UMC Utrecht, Utrecht, Netherlands, 4Philips, Best, Netherlands
Synopsis
Keywords: Motion Correction, Motion Correction, Directional coupler
Motion is a prevalent issue in MRI. Here, we present a fast prospective motion correction (PMC) algorithm using directional couplers as head-motion sensors. The mean reflectances (per volume) of an 8 channel pTx head coil are modelled to head-position recorded by the systems built-in PMC during a calibration scan. Thereafter, the model is used for motion correction at a per RF pulse temporal resolution. Results indicate partial motion correction with error in correcting fast motions due to the necessary presence of filters to remove unidentified low-frequency noise. Mitigating low-frequency noise contributions would allow for less filtering thereby improving dynamic performance.
Introduction
The effects of motion in MRI cause image artifacts and corrupt
information in advanced MRI applications. Prospective motion correction (PMC)
aims to alleviate image artifacts and preserve spin history effects. In this
abstract, we provide an update on our previous work of prospective head-motion
correction using directional couplers (DiCo’s) as motion sensors [1,2]. Now, acquisition
updates can be sent on a per-RF-pulse basis. The reflectance measured during
the RF transmit phase from a multi-channel head coil is modelled to head
position in a calibration phase. The model is used in subsequent scans to
detect and correct for motion. The algorithm to achieve this is termed DiCo
motion correction (DMC). DMC has been successfully implemented in the scanner
environment and runs as a separate client application.Methods
Experiments were performed using an 8-channel Nova Medical head coil on a
Philips 7T MR scanner. The sequence used is a 3D FFE scan (TR/TE=5/1ms, SENSE
4x2, half-scan 75% in the phase encoding directions; 0.8s per volume). DiCo’s
in the RF amplifiers record the reflected and forwards signal during every RF
excitation. The DMC algorithm requires a calibration phase prior to performing
PMC. In this phase the volunteer is asked to move while DiCo data is streamed
from the scanner to the DMC where the reflectances are calculated and stored at
a per RF-pulse resolution. Low-pass filters are used to filter high-frequency,
non-motion associated reflectance components. During this phase, a modified
version of the scanner’s built-in PMC algorithm is used to record the
rigid-body head position at a per volume temporal resolution. The calibration
scan used 500 volumes. After the calibration scan, the mean reflectances over a
single volume are modelled to the corresponding position. The models treat the
real and imaginary reflectances from each channel as independent variables. A
linear model is used to fit the data. The same sequence is used during the
motion correction phase. In this phase, the DMC streams DiCo data and computes
the position at a per RF pulse temporal resolution (i.e. 5ms). If the position
is greater than some threshold, then the DMC sends a volume acquisition update
to the scanner. The threshold for sending updates is a detected motion change of
0.5mm translation or 0.5° rotation. Two scans are performed with this motion
correction active. In the first scan, the volunteer remains still and in the
second scan the volunteer moves in the same manner as was done in the
calibration scan. The low-pass filter defines the DMC sensitivity to motion and
has been set to 0.5Hz.Results & Discussion
Figure 1 conveys the motion trajectory used during the calibration scan with an
experienced volunteer. Figure 2 shows the linear model fitting for each
motion to fit well to the data. During a motion correction scan, acquisition
updates can successfully be sent every RF pulse for TR>=5ms. Figure 3
presents a snippet of the interpreted motions during the no-motion scan:
updates are sent during the pre-pulse phase to reach steady state (dummy pulses) and
during the acquisition. Updates during the acquisition are sent due to
false-motion interpretations caused by an unidentified low-frequency component
in the system. Increasing the cut-off frequency introduces larger false motion
interpretations. Figure 4 presents the reconstructed motion trajectory
during the motion correction phase per RF pulse. The algorithm is tracking
real-motions and correlates well with the trajectory seen in Figure 1. However, with DMC motion correction active, a residual motion is still seen by the scanners
built-in PMC – particularly at points of faster, more extreme motion – see Figures
5. This outcome is a result of the low-pass filter. To prevent false-motion
interpretations, the filter frequency is set low, this penalizes the DMCs
capacity to correctly interpret and correct for fast motions. Additional to
this, there is an assumption that a model calibrated at a per volume temporal
resolution will be the same as the model that performs best at the per RF pulse
temporal resolution. This assumption may not be valid and could be a
contributor to the current performance.Conclusion & Outlook
Fast prospective motion correction has successfully been implemented on
the scanner with promising results. Efforts to investigate and obviate
low-frequency signals in the reflectance could see the method become a
successful technique of PMC. Furthermore, investigating calibration sequences
which use fewer RF excitations could improve model performance at a per pulse temporal
resolution. Acknowledgements
No acknowledgement found.References
[1] Proceedings ISMRM
2020, Abstract 1280.
[2] ISMRM Workshop on
Motion Detection & Correction 2022. Abstract 44.