Fatemeh Rastegar Jooybari1,2, Christopher Huynh2, Sharon Portnoy2, Jonathan Voutsas3, Diana Balmer-Minnes2, Ankavipar Saprungruang4, Shi-Joon Yoo5, Christopher Z Lam4, and Christopher K Macgowan1,2
1Medical Biophysics, University of Toronto, Toronto, ON, Canada, 2Translational Medicine, Hospital for Sick Children, Toronto, ON, Canada, 3Queen's University, Kingston, ON, Canada, 4Department of Diagnostic Imaging, Hospital for Sick Children, Toronto, ON, Canada, 5Departments of Medical Imaging and Paediatrics, University of Toronto, Toronto, ON, Canada
Synopsis
Keywords: Cardiovascular, Quantitative Imaging, Flow, Heart, Cardiovascular, Congenital heart disease
The
application of conventional 4D flow to evaluate the hemodynamics of the heart
and vessels is limited by long scan time. Recent advances aim to reduce scan
time while providing sufficient accuracy. Here we investigate a 4D flow
pipeline based on an undersampled 3D golden angle radial trajectory that offers
reconstruction flexibility. We evaluated this technique in a cross-section of 11
pediatric patients with congenital heart disease and compared flows against conventional
2D phase contrast in target vessels.
Introduction
Cardiovascular magnetic
resonance (CMR) imaging plays a key role in the evaluation and monitoring of anatomy
and hemodynamics in congenital
heart disease (CHD). Two-dimensional (2D) velocity-encoded
phase-contrast (PC) cine MRI is the primary technique used to measure blood
flow velocity and volume. However, this method is limited to velocity analysis from
2D planes through targeted vessels1,2. 4D flow MRI is a CMR imaging technique
that is suitable for the evaluation of complex CHD, as it provides a 3D vector
field of blood flow velocities throughout the cardiac cycle. The clinical
application of 4D flow has been limited historically by long scan times (more
than 15 minutes)3,4. However, the last decade has seen the
development of techniques (e.g. parallel imaging, compressed sensing) that have
reduced scan times to approximately 5-10 minutes5,6,7. Here we investigate
a 5-minute 4D flow technique, based on an undersampled 3D radial acquisition, which
offers robustness to motion and flexibility in the reconstruction, and quantify
its accuracy in measuring net flow in healthy adult volunteers and a
cross-section of pediatric patients with CHD.Method
Subjects: 11 pediatric CHD patients (12.1
± 2.9 years, M:F=8:3) and 4 healthy volunteer (26.7 ± 1.9 years, M:F=3:1) were
recruited (see Table 1 for CHD types).
Acquisition and reconstruction: Patients and healthy controls were scanned on
a 1.5T and 3.0T MRI, respectively (Avanto and Prisma, Siemens Healthineers,
Germany).
4D flow data
were acquired using an internally developed sequence involving a 3D center-out
radial trajectory with golden angle spoke ordering and 4-point velocity encoding. A
5-minute free-breathing scan (5 minutes and 17±1.9 seconds) with 20357±223
spokes per velocity encode was performed (~80,000 spokes total) with the
following relevant parameters: VENC = 150 cm/s, number of averages = 1, TR/TE = 3.83-4.01/1.42 ms, flip angle = 8 degrees. Data were retrospectively sorted
into 8 cardiac frames using the ECG signal, and without respiratory
compensation. Images were reconstructed using compressed sensing with
alternating direction method of multipliers (ADMM) solver8 enforcing
sparsity in the temporal and spatial finite difference domains (λ = 6e-7) and
20 iterations. The pipeline is implemented in Python with the SigPy library9 and run on Research IT-High Performance Computing (RIT-HPC) service at Sickkids using 8 processors and 1 Tesla-V100 GPU node. The reconstructed FOV and spatial resolution were 252x504x504 mm3
and 1.8x1.8x1.8 mm3, respectively.
For
reference, 2D phase contrast was also performed in target vessels: ascending
aorta (AAo), descending aorta (DAo), main/left/right pulmonary arteries
(MPA/LPA/RPA), and superior vena cava (SVC). Relevant scan parameters were: VENC = 150% of
expected peak velocity by vessel, TR/TE = 5.51-7.12/2.9-3.25 ms, number of
averages = 2, reconstructed phases = 25, spatial resolution = 1.09-1.33 mm,
slice thickness = 4.5-5 mm, flip angle = 20-25 degrees.
Flow analysis: The reconstructed 2D and 4D flow
data were analyzed using commercial post-processing software (QFlow, Medis Medical
Imaging Systems, the Netherlands; and cvi42, Circle Imaging, Canada
respectively). A degree-3 polynomial was fit and subtracted to correct for
phase background. Following phase unwrapping for the 4D flow images, regions of
interest were segmented for the analysis, positioned based on the prescription of
the 2D phase contrast MR slices.Results and Discussion
The average
reconstruction time was 48.5±9.7 mintues for each dataset. One CHD case was
excluded due to excessive bulk motion during the 4D flow scan. Qualitative visualizations
of flow from one CHD patient are presented in Figure 1, depicting animated pathlines
for a case of repaired
ventricular septal defect (VSD) and coarctation of the aorta. The animation on
the right panel shows that an increased flow remains at the coarctation site
even after the palliative procedure. The velocity map shows the area with
increased velocity in the aorta (maximum velocity was 2 m/s). Measurements by
2D phase contrast were only performed at the aortic root.
Figure 2 (left
panel) shows a comparison of net flow measured by radial 4D flow versus
conventional 2D phase contrast in healthy volunteers. Linear regression (orange
line) showed good-excellent agreement between the measurements from healthy
subjects (R2 = 0.90, slope = 0.94±0.15 with 95% confidence interval, root mean square error
[RMSE] = 8.29 ml/s). Figure
2 (right panel) shows the Bland-Altman comparison between 4D and 2D
measurements, with a difference of 2.3±8.2 ml/s (p=0.13).
As shown in
Figure 3, good correlation was also obtained between 4D vs. 2D net flows from CHD
patients (R2 = 0.94, slope = 0.87±0.06 with 95% confidence interval,
RMSE = 5.63 ml/s). Net flows
within a given vessel varied widely between CHD patients because of differences
in age and pathology. Bland-Altman analysis shows an underestimation of 4D flow
measurements by 9.4±10.1 ml/s (p = 3.2e-9) in the cohort of CHD patients.
Previous studies
that used motion correction, respiratory gating or longer scan times have shown
similar agreement between 2D and 4D flow measurements in healthy volunteers10,11,12,
and in infants with CHD using time-averaged volumetric flow imaging to keep
scan times short (~3 minutes)12.Conclusion
The proposed
4D flow method can provide retrospective in-vivo 3D quantification and
visualization in 5 minutes in technically challenging clinical population. Future
work will evaluate the effect of respiratory motion correction, contrast agent
and field strength on flow accuracy.Acknowledgements
Canadian
Institutes of Health Research (CIHR) PJT 427837. We thank MR technologists Vivian
Tassos and Joti Gill for their assistance with scanning, and patients and families
for participating in this research.References
1. Whitehead, K. K., Harris, M. A., Glatz, A. C., Gillespie, M. J., DiMaria, M.
V., Harrison, N. E., ... & Fogel, M. A. (2015). Status of systemic to
pulmonary arterial collateral flow after the fontan procedure. The American
journal of cardiology, 115(12), 1739-1745.
2. Geiger,
J., Markl, M., Jung, B., Grohmann, J., Stiller, B., Langer, M., & Arnold,
R. (2011). 4D-MR flow analysis in patients after repair for tetralogy of
Fallot. European radiology, 21(8), 1651-1657.
3. Markl,
M., Frydrychowicz, A., Kozerke, S., Hope, M., & Wieben, O. (2012). 4D flow
MRI. Journal of Magnetic Resonance Imaging, 36(5), 1015-1036.
4. Coppo,
S., Piccini, D., Bonanno, G., Chaptinel, J., Vincenti, G., Feliciano, H., ...
& Stuber, M. (2015). Free‐running 4D whole‐heart self‐navigated golden
angle MRI: initial results. Magnetic resonance in medicine, 74(5), 1306-1316.
5. Pruitt,
A., Rich, A., Liu, Y., Jin, N., Potter, L., Tong, M., ... & Ahmad, R.
(2021). Fully self‐gated whole‐heart 4D flow imaging from a 5‐minute
scan. Magnetic resonance in medicine, 85(3), 1222-1236.
6. Ma, L.
E., Yerly, J., Piccini, D., Di Sopra, L., Roy, C. W., Carr, J. C., ... &
Markl, M. (2020). 5D flow MRI: a fully self-gated, free-running framework for
cardiac and respiratory motion–resolved 3D hemodynamics. Radiology:
Cardiothoracic Imaging, 2(6).
7. Cheng,
J. Y., Hanneman, K., Zhang, T., Alley, M. T., Lai, P., Tamir, J. I., ... &
Vasanawala, S. S. (2016). Comprehensive motion‐compensated highly accelerated
4D flow MRI with ferumoxytol enhancement for pediatric congenital heart
disease. Journal of Magnetic Resonance Imaging, 43(6), 1355-1368.
8. Wahlberg, B., Boyd, S., Annergren, M., & Wang, Y. (2012). An ADMM algorithm
for a class of total variation regularized estimation problems. IFAC
Proceedings Volumes, 45(16), 83-88.
9. https://sigpy.readthedocs.io/en/latest/
10. Blanken, C. P., Gottwald, L. M., Westenberg, J. J., Peper, E. S., Coolen, B.
F., Strijkers, G. J., ... & van Ooij, P. (2022). Whole‐Heart 4D Flow MRI
for Evaluation of Normal and Regurgitant Valvular Flow: A Quantitative
Comparison Between Pseudo‐Spiral Sampling and EPI Readout. Journal of
Magnetic Resonance Imaging, 55(4), 1120-1130.
11. Jacobs, K. G., Chan, F. P., Cheng, J. Y., Vasanawala, S. S., & Maskatia, S.
A. (2020). 4D flow vs. 2D cardiac MRI for the evaluation of pulmonary
regurgitation and ventricular volume in repaired tetralogy of Fallot: a
retrospective case control study. The international journal of
cardiovascular imaging, 36(4), 657-669.
12. Schrauben, E. M., Lim, J. M., Goolaub, D. S., Marini, D., Seed, M., &
Macgowan, C. K. (2020). Motion robust respiratory‐resolved 3D radial flow MRI
and its application in neonatal congenital heart disease. Magnetic
Resonance in Medicine, 83(2), 535-548.