Kelly Payette1,2, Carla Avena Zampieri1, Jordina Aviles Verdera1,2, Lisa Story3, Raphael Tomi-Tricot4, Joseph V Hajnal1,2, Mary Rutherford1,2, and Jana Hutter1,2
1Centre for the Developing Brain, School of Biomedical Engineering and Imaging Sciences KCL, London, United Kingdom, 2Biomedical Engineering Department, School of Biomedical Engineering and Imaging Sciences, KCL, London, United Kingdom, 3Women’s Health, KCL, London, United Kingdom, 4MR Research Collaborations, Siemens Healthcare Limited, Frimley, United Kingdom
Synopsis
Keywords: Fetal, Low-Field MRI
Low
field MRI for fetal imaging has the potential to be a powerful tool for
prenatal diagnosis, with a favourable
combination of scope for a larger bore (adding comfort and reducing claustrophobia),
and lower SAR, susceptibility effects and receiver coil properties. We performed
functional scans of the fetal lung to demonstrate the high image quality that
can be acquired. The quantitative values of the fetal lungs (IVIM parameters,
T1, T2*) calculated from the low field images match the expected values from
higher field strengths reported in the literature.
Introduction
Low field
magnetic resonance imaging (MRI) is emerging as an exciting field, especially
in the perinatal realm1. The lower field strength results in less
magnetic susceptibility artifacts, negating the need for shimming and advanced imaging
pre-processing techniques while increasing the dynamic range for T2*
relaxometry. The shorter T1 times improve image contrast and raise efficiency2. Finally, the wide bore present in low field
systems makes for a more comfortable imaging experience for pregnant women. These
benefits make a low field imaging system ideal for fetal imaging.
As
fetal lung development is critical to survival after birth, it is important to understand
the functional and structural changes occurring throughout gestation. Lung
volumes from MRI have been shown to be a useful prognosticator for disorders
such as fetal growth restriction and congenital diaphragmatic hernia, as well
as for preterm premature membrane rupture3–5. However, functional, quantitative techniques
such as T1, T2* and diffusion MRI have not yet been studied in depth for the
fetal lungs. The most commonly studied, fractional perfusion, using diffusion
MRI and the IVIM model, has been shown to be correlated with gestational age6,7, and early low field studies have shown that
diffusion measurements can be an indicator of fetal lung maturation8. Therefore,
we propose to show that functional quantitative data acquired with a low field
scanner is equivalent to that acquired by standard clinical scanners, and that
functional sequences can be used to explore fetal lung development.Methods
Fetal
MRI was acquired as part of an ethically approved study (MEERKAT, REC
19/LO/0852, Dulwich Ethics Committee, 08/12/2021) performed between May-October 2022 at St Thomas’ Hospital, in London, UK. 38 women were scanned on a
clinical 0.55T scanner (MAGNETOM Free.Max, Siemens Healthcare,
Germany) using a
6-element blanket coil and a 9-element spine coil built into the table
scanner. Three different types of functional sequences were acquired:
diffusion MRI, T2* relaxometry and T1 relaxometry, all sharing a FOV of 400x400
mm2. Diffusion scans were acquired with a single-shot
diffusion-weighted spin echo sequence (resolution=3x3x3mm3, TR= 7200ms,
TE=129ms, 8 b-values between 0-1000s/mm2). Inversion-recovery data
was acquired with a single-shot gradient echo sequence (resolution=4x4x4mm3,
TR=3820ms, TE=80ms) and multi-echo T2* data was acquired with a single-shot
gradient-echo sequence (TR=15480ms, , 5 TEs=[49–345]ms GRAPPA=2). Fetal T2* lung
images were acquired at 3T (resolution: 3x3x3mm2, 4 TEs=[38-240],
TR=7200ms).
The
images were reviewed, and cases with large amounts of motion were excluded. Example
b0 images from the diffusion scan can be seen in Figure
1 overlaid with a manually segmented mask of the
fetal lung.
The IVIM parameter maps (fractional perfusion
(f), diffusion coefficient (D), pseudo-diffusion coefficient (D*), were
calculated using dmipy’s9 implementation of the bi-exponential mode . After
reviewing the fit of each model, two further cases were excluded where the fit
failed. The T1 and T2* maps were created by using mono-exponential fitting in
an in-house tool in python.
The mean of each of the parameters was
calculated for each fetal lung ROI, and the relationship of all parameters with
gestational age (GA) was explored. The calculated values were compared with
existing literature at higher field strengths, where available. The Pearson
Correlation Coefficient (PCC) was calculated to determine the relationship of
each variable with GA. Results
After quality checks, and importantly reviewing
images for the inclusion of lungs, 30 cases with good quality fetal lung diffusion
scans were selected, 14 with T1 scans, and 16 with T2* scans. Figure
2 shows an example diffusion scan and the
resulting IVIM maps. Figure
3 shows the relationship between each of the diffusion
parameters and GA.
An example T1 image and corresponding T1 map
can be seen in Figure 4. The lungs can be clearly seen in both the
original image and in the T1 map. The average T1 values of the fetal lung show
an increase with GA. Figure
5 shows
the T2* image and corresponding T2* map of the same subject. Again, the lungs
can clearly be visualized, but here when the average T2* values of all cases
are compared to GA, there is not a clear trend. Discussion and Conclusion
The presented functional contrasts on the fetal lungs illustrate that low field MRI
scanners are a promising avenue for fetal MRI. The quantitative results are in-line
with the literature where expected (the IVIM parameters), and
deviate as expected (higher T2* values when compared to 3T). At higher
field strengths, extensive pre-processing of diffusion data is required in
order to correct for eddy currents and b0 distortion, and these computationally-intensive
steps were not required for this low field analysis.
The
correlation between each of the IVIM parameters and GA is weak, whereas the
literature has reported a correlation between f and GA. However, the values of
each of the parameters are in the same range as reported on by the existing
fetal lung IVIM literature acquired at 1.5T and 3T at the higher GA6,7. The lack of motion correction may explain
this deviation especially at the lower GA, which is where most of the excluded
motion and discrepancies with the literature lie. Future analysis should
incorporate motion correction in order to improve the number of cases included
in such a study, especially at lower GAs.Acknowledgements
The
authors thank all pregnant women and their families for taking part in this
study. The authors thank the research midwives and radiographers for their
crucial involvement in the acquisition of these datasets. This work was
supported by a Wellcome Trust Collaboration in Science grant [WT201526/Z/16/Z],
a UKRI FL fellowship and by core funding from the Wellcome/EPSRC Centre for
Medical Engineering [WT203148/Z/16/Z]. The views presented in this study
represent these of the authors and not of Guy's and St Thomas' NHS Foundation
Trust.References
1. Aviles, J. et al. A Fast Anatomical
and Quantitative MRI Fetal Exam at Low Field. in Perinatal,
Preterm and Paediatric Image Analysis (eds. Licandro, R., Melbourne, A.,
Abaci Turk, E., Macgowan, C. & Hutter, J.) 13–24 (Springer Nature
Switzerland, 2022). doi:10.1007/978-3-031-17117-8_2.
2. Hori, M., Hagiwara, A., Goto, M., Wada,
A. & Aoki, S. Low-Field Magnetic Resonance Imaging. Invest. Radiol. 56,
669–679 (2021).
3. Corroenne, R. et al.
Cost-effective fetal lung volumetry for assessment of congenital diaphragmatic
hernia. Eur. J. Obstet. Gynecol. Reprod. Biol. 260, 22–28 (2021).
4. Moradi, B. et al. Comparison of
fetal lung maturation in fetuses with intrauterine growth restriction with
control group, using lung volume, lung/liver and lung/muscle signal intensity
and apparent diffusion coefficient ratios on different magnetic resonance
imaging sequences. J. Matern.-Fetal Neonatal Med. Off. J. Eur. Assoc.
Perinat. Med. Fed. Asia Ocean. Perinat. Soc. Int. Soc. Perinat. Obstet. 1–9
(2021) doi:10.1080/14767058.2021.2008349.
5. Messerschmidt, A. et al. Fetal
MRI for prediction of neonatal mortality following preterm premature rupture of
the fetal membranes. Pediatr. Radiol. 41, 1416–1420 (2011).
6. Ercolani, G. et al. IntraVoxel
Incoherent Motion (IVIM) MRI of fetal lung and kidney: Can the perfusion
fraction be a marker of normal pulmonary and renal maturation? Eur. J.
Radiol. 139, 109726 (2021).
7. Jakab, A. et al. Microvascular
perfusion of the placenta, developing fetal liver, and lungs assessed with
intravoxel incoherent motion imaging. J. Magn. Reson. Imaging JMRI 48,
214–225 (2018).
8. Moore, R. J., Strachan, B., Tyler, D.
J., Baker, P. N. & Gowland, P. A. In vivo diffusion measurements as an
indication of fetal lung maturation using echo planar imaging at 0.5T. Magn.
Reson. Med. 45, 247–253 (2001).
9. Fick, R. H. J., Wassermann, D. &
Deriche, R. The Dmipy Toolbox: Diffusion MRI Multi-Compartment Modeling and
Microstructure Recovery Made Easy. Front. Neuroinformatics 13,
(2019).