Fuchang Jiang1, Kaylee Henry1, Bhumi Bhusal2, Pia Sanpitak2, Gregory Webster3, Andrada Popescu4, Giorgio Bonmassar5, Christina Laternser6, Daniel Kim2, and Laleh Golestanirad1,2
1Department of Biomedical Engineering, Northwestern University, Evanston, IL, United States, 2Department of Radiology, Northwestern University, Chicago, IL, United States, 3Division of Cardiology, Ann and Robert H. Lurie Children's Hospital of Chicago, Chicago, IL, United States, 4Division of Medical Imaging, Ann and Robert H. Lurie Children's Hospital of Chicago, Chicago, IL, United States, 5A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Boston, MA, United States, 6Center for Cardiovascular Innovation, Ann and Robert H. Lurie Children's Hospital of Chicago, Chicago, IL, United States
Synopsis
Keywords: Heart, Safety, Implants
Children with congenital heart defects often have life-sustaining
indications for an epicardial cardiac implantable electronic device (CIED). However
few data exist on endocardial systems in children. Additionally, the FDA has
never approved an epicardial system as MRI-Conditional due to limited data on
potential heating risk. To provide evidence-based knowledge on RF-induced
heating of CIEDs in children and adults with epicardial and endocardial leads
of different lengths, we recorded the temperature increase of 120 clinically
relevant trajectories positioned into adult and pediatric phantoms. Our results
highlight the need for age and device-specific assessments of MRI safety in
children with CIEDs.
Introduction
At least 75% of patients who receive a cardiac implantable electronic
device (CIED) will require MRI during their lifetime1. MR-conditional endocardial CIEDs have been
approved for adults, where the implantable pulse generator (IPG) is placed in
the pectoral region and leads are passed through the subclavian vein to reach the
interior of the heart. Most children, however, receive an epicardial CIED due
to their smaller anatomy or complex vasculature, where the IPG is placed in the
abdomen and leads are directly sewn to the myocardium (Figure 1). While the
2021 PACES guideline made a class IIb recommendation that MRI of pediatric
patients with epicardial or abandoned leads be considered individually based
upon their risk/benefit ratio, no guidance is available on how to quantify
those risks2. However, due to
limited data availability, children’s hospitals default to either refusing MRI
service to children with CIEDs or scanning patients based on results from adult
studies that have limited applicability in pediatrics. We argue that
improvements can be made to the current guidelines and that the decision
whether to scan children with CIEDs should be made on an individual basis.Method
We
custom-made human-shaped phantoms that mimic the silhouette of an average-sized
adult and a twenty-nine-month-old child. Phantoms were filled with
polyacrylamide (PAA) gel (22 L for the adult and 9 L for the pediatric phantom)
with a conductivity of σ = 0.47 S/m and relative permittivity of εr =
88 representing dielectric values of average tissue (Figure 2A). We reviewed
chest X-rays and computed tomography (CT) images of 200 adult and pediatric
patients with epicardial and endocardial CIEDs to create representative device
trajectories. Our Institutional Review Board approved the retrospective use of
patient imaging data for the purpose of modeling and safety assessment. To
enhance reproducibility, we designed and 3D printed trajectory guides which
helped route the leads along patient-derived trajectories and keep them
securely in place during the experiments, similar to our previous works3,4 (Figure 2C). 120 epicardial and endocardial lead
trajectories of varying lead lengths (15, 25, 35 cm epicardial leads and 35,
45, and 58 cm endocardial leads) were positioned into the adult and pediatric
phantoms following clinically relevant trajectories based on patient images
(Figure 2B & Figure 3).
The
temperature increase was recorded using MR-compatible fiber optic probes
secured at the tip of the lead. RF exposure was performed in a 1.5 T Siemens
Aera scanner (Siemens Healthineers, Erlangen, Germany) with the phantoms
positioned such that the chest was at the isocenter and using a high-SAR
T1-weighted turbo spin echo sequence (TE = 7.3 ms, TR = 897 ms, B1+ = 5 µT, AT
= 280 s).Results
There was significantly higher RF heating of epicardial leads compared
to endocardial leads in the pediatric phantom (p<0.001); however, there was
no significant difference between epicardial and endocardial leads in the adult
phantom (p=0.24). The maximum RF heating was observed to be ~12 °C which
occurred for the 25-cm epicardial lead. For a 10-minute scan, this will be
equivalent to the cumulative thermal dose of CEM43°C = 1280 minutes. This is
high enough to cause necrosis in pig muscles5. This is concerning,
as younger children — with thin hearts which are less tolerant to thermal
injury — are more likely to receive shorter leads (i.e., 25 cm) as opposed to
longer leads (e.g., 35 cm). Therefore, RF safety studies of epicardial leads in
older children and young adults should not be offhandedly generalized to all
children. In contrast, endocardial leads in the pediatric phantom generated
significantly less RF heating than endocardial leads in the adult phantom (0.6
± 0.4 °C vs. 2.0 ± 1.8 °C, p<0.001). Discussion and Conclusion
Tissue heating
from radiofrequency (RF) excitation fields remains a major issue. Two factors
determine the severity of RF heating of implanted CIEDs: the patient’s body
size and the lead’s length/trajectory6-8. Because the position, orientation, trajectory, and length
of the lead within the human body determine the degree to which the MRI
electric field couples with the lead, RF safety in children must be assessed on
an age-specific basis. Due to their smaller anatomy, endocardial leads in
children often have more loops around the IPG compared to the same devices
implanted in adults. Reducing the straight portion of an elongated lead has
been shown to reduce RF heating in neuromodulator devices9, which explains why endocardial leads generated
significantly lower RF heating in children. Also, younger children are more
likely to receive shorter epicardial leads (e.g., 25 cm). The RF field’s half-wavelength
in the tissue is ~26 cm at 64 MHz, which explains the excessive heating of 25 cm
leads whose length approaches the resonance length. We demonstrate that
endocardial leads may be lower risk for hospitals to start imaging children
with CIEDs, moving the field away from a blanket ban on children with CIEDs. Additionally,
our epicardial data highlights the need for age and device-specific assessment
of MRI safety in children with CIEDs. Specifically, safety studies in adults
should not be offhandedly interpolated to children who receive different CIED
lead lengths and follow different trajectories. Acknowledgements
This study was supported by
NIH grant R01EB030324.References
1. Kalin
R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and
defibrillator patients. Pacing Clin
Electrophysiol. 2005;28(4):326-328.
2. Writing
Committee M, Shah MJ, Silka MJ, et al. 2021 PACES Expert Consensus Statement on
the Indications and Management of Cardiovascular Implantable Electronic Devices
in Pediatric Patients: Developed in collaboration with and endorsed by the
Heart Rhythm Society (HRS), the American College of Cardiology (ACC), the
American Heart Association (AHA), and the Association for European Paediatric
and Congenital Cardiology (AEPC) Endorsed by the Asia Pacific Heart Rhythm
Society (APHRS), the Indian Heart Rhythm Society (IHRS), and the Latin American
Heart Rhythm Society (LAHRS). JACC Clin
Electrophysiol. 2021;7(11):1437-1472.
3. Golestanirad
L, Kazemivalipour E, Keil B, et al. Reconfigurable MRI coil technology can
substantially reduce RF heating of deep brain stimulation implants: First
in-vitro study of RF heating reduction in bilateral DBS leads at 1.5 T. PLoS One. 2019;14(8):e0220043.
4. Bhusal B,
Stockmann J, Guerin B, et al. Safety and image quality at 7T MRI for deep brain
stimulation systems: Ex vivo study with lead-only and full-systems. PLoS One. 2021;16(9):e0257077.
5. van Rhoon GC,
Samaras T, Yarmolenko PS, Dewhirst MW, Neufeld E, Kuster N. CEM43°C thermal
dose thresholds: a potential guide for magnetic resonance radiofrequency
exposure levels? European radiology. 2013;23(8):2215-2227.
6. Mattei E,
Triventi M, Calcagnini G, et al. Complexity of MRI induced heating on metallic
leads: Experimental measurements of 374 configurations. BioMedical Engineering OnLine. 2008;7(1):11.
7. Nordbeck P,
Weiss I, Ehses P, et al. Measuring RF-induced currents inside implants: Impact
of device configuration on MRI safety of cardiac pacemaker leads. Magn Reson Med. 2009;61(3):570-578.
8. Van Den
Bergen B vdBC, Kroeze H, Bartels L, Lagendijk J. The effect of body size and
shape on RF safety and B1 field homogeneity at 3T. 2006.
9. Golestanirad L, Kirsch J, Bonmassar G,
et al. RF-induced heating in tissue near bilateral DBS implants during MRI at
1.5 T and 3T: The role of surgical lead management. Neuroimage. 2019;184:566-576.