Clement Debacker1,2, Geoffroy Pouliquen1,2, Sylvain Charron2, Anna Fayolle1,2, Valentin H. Prevost3, Wolter de Graaf4, Alexandre Roux2,5, Johan Pallud2,5, and Catherine Oppenheim1,2
1Radiology department, GHU Paris Psychiatrie et Neurosciences, Site Sainte-Anne, Paris, France, 2IMA-BRAIN, Université Paris Cité, Institute of Psychiatry and Neuroscience of Paris (IPNP), INSERM U1266, Paris, France, 3Canon Medical Systems Corporation, Tochigi, Japan, 4Canon Medical Systems Europe, Zoetermeer, Netherlands, 5Neuro-surgery department, GHU Paris Psychiatrie et Neurosciences, Site Sainte-Anne, Paris, France
Synopsis
Keywords: Tumors, Tumor
Our work is the clinical validation of a
deep-learning algorithm (DLR) used to denoise MR images on quantitative MR biomarkers.
Since it has been trained on including T1- and T2-weighted conventional images,
in healthy volunteers, its effects on multiparametric quantitative MRI in
patients, are uncertain. It could potentially improve brain tumors
characterization by providing quantitative biomarkers under clinical
time-constraints.
INTRODUCTION
Brain neoplasm remains a diagnostic
challenge (1). Quantitative information can be extracted from relaxometry and
Diffusion Tensor Imaging (DTI). Despite their clinical potential, their long
duration prevents their wide integration in clinical settings. Denoising using
deep learning, deep learning-based reconstruction (DLR), has been used to improve image quality or to compensate for the
degradation induced by acquisition time reduction (2). Since it has been trained on T1 and T2-weighted conventional
images, in healthy volunteers, its effects on multiparametric quantitative MRI
are uncertain.METHODS
Subjects
22 patients (11 women; mean age, 56.0 ±
14.2 years) patients with a known or suspected supratentorial brain tumor were
prospectively enrolled before surgery from April to June 2021 and scanned on a
3T whole-body MRI scanner (Vantage Galan 3T / XGO; Canon Medical Systems
Corporation, Tochigi, Japan) with a 32-channel head coil.
Imaging protocol
A 3D T1 MPRAGE was performed for intra-subject registration. DTI was
performed using a 2D single-shot spin-echo EPI imaging sequence with whole brain coverage: in-plane resolution=2
mm2; STH=2mm; 30 directions; b-values=0-1000 s/mm²; TR/TE=5.2 s/85
ms; NEX=1; TA=2min58s. A 3D T1 MP2RAGE
sequence was performed to generate T1-map: in-plane resolution=0.7x0.5
mm²; STH=2.5 mm; TR/TE=7.5 ms/3.3 ms; TI=664/3300 ms; inter-shot TR=7 s; NEX=1;
TA=3min09s. A 2D T2 FSE sequence was performed at 4 echo times values to generate T2-maps: in-plane
resolution=0.7x0.5 mm²; STH=2 mm; TR=6 s; TE=20/60/100/140 ms; NEX=1; TA=2min44s.
To evaluate the denoising effect of DLR, quantitative sequences were acquired
with one number of acquisitions (NAQ1) and 3 number of acquisitions (NAQ3) as
ground truth data. The vendor-supplied DLR algorithm was applied to NAQ1
(DLR-NAQ1).
Data processing
T1 and T2 maps have been generated using
Olea Sphere software. Data were pre-processed using FSL (3). For DTI data the processing steps were: susceptibility-induced
off-resonance field estimated using top-up (4), then correction for eddy current and motion-distortions using eddy
(5) and finally fractional anisotropy (FA) and mean diffusivity (MD)
computed using the dtifit command. The parametric maps were co-registered onto
the 3D-anatomical scan of each patient using a 3D rigid-registration via FLIRT (6). Deep Learning reconstructions have been performed with the
application of a deep Convolutional Neuronal Network combined with a
low-pass filtered component, in order to maintain original contrasts (2). For healthy tissues, six region-of-interest (ROIs) were placed in three
normal brain areas (cortex, white matter, deep gray matter). For lesions, four
ROIs were placed in the lesion center and in the lesion periphery.RESULTS
Figure
1 shows representative images for FA, MD, T1 and
T2-maps calculated from NAQ1, DLR-NAQ1 and NAQ3 images. For all studied
parameters, except for FA, the main effect of ROIs
location was always significant with expected variations depending on ROIs
location, thus all ROIs were pooled for further analysis. Note
that the putamen ROI has been removed in the analysis of the FA-parameter
considering the significant interaction effect between the ROI location and
acquisition-reconstruction methods.
As shown in Figure 2, there was no significant difference in mean FA
values between DLR-NAQ1 and NAQ3. Interestingly, mean FA significantly differed
(P < 0.001) between NAQ1 and the other two acquisition-reconstructions, with
values being higher in NAQ1. Similar findings were observed for SDs. In
particular, SDs were smaller (P < 0.001) in DLR-NAQ1 and NAQ3 than NAQ1. As
shown in Figure 3, there was no significant difference in mean MD
values between the three groups. SDs significantly differed between the three
groups. As shown in Figure 3, there was no significant difference in mean T1
values between the three groups. SDs of T1 values significantly differed
between NAQ1 and the other two acquisition-reconstructions, with values being
lower in DLR-NAQ1 (P < 0.01). The results for the T2 values are similar to
those of T1 values (data not shown).DISCUSSION
Overall, in
a series of 22 consecutive patients with brain tumors, we found that DLR
applied to fast and noisy MR sequences improved the reliability of quantitative
parameters (FA, MD, T1- T2-relaxation times). Indeed, values were similar to
those obtained from longer sequences, with less variability compared to fast ones
without DLR. This was observed irrespective of the brain region.
Interestingly regarding FA values, a significant
difference was identified in NAQ1 compared to that of DLR-NAQ1 and NAQ3, with
higher values in NAQ1. In contrast, there were no significant differences
regarding mean MD values between the three groups. It has been reported that an
upward bias of FA and no significant bias in MD are observed as signal-to-noise
ratio decreased (7). By reducing image noise using DLR, the FA values
of DLR-NAQ1 became much closer to that of NAQ3 to the point that there was not
any significant difference between DLR-NAQ1 and NAQ3.CONCLUSION
To our
knowledge, this is the first study that assessed the effect of DLR on
quantitative MR biomarkers, including T1 and T2-relaxation time maps and
DTI-derived metrics, in patients with a supratentorial brain tumor. Showing
that reliable quantitative biomarkers can be derived from short sequences (≈3
minutes each), will promote their use in a clinical setting. The parametric
maps could help characterize brain tumor subtypes or to differentiate a unique
metastasis from a high-grade glioma owing to quantitative biomarkers and
improve diagnosis.Acknowledgements
No acknowledgement found.References
1. Kern M, Auer TA, Picht T, Misch M, Wiener E.
T2 mapping of molecular subtypes of WHO grade II/III gliomas. BMC Neurol
2020;20:1–9 doi: 10.1186/s12883-019-1590-1.
2. Kidoh M, Shinoda
K, Kitajima M, et al. Deep Learning Based Noise Reduction for Brain MR Imaging:
Tests on Phantoms and Healthy Volunteers. Magn Reson Med Sci 2019;advpub doi:
10.2463/mrms.mp.2019-0018.
3. Jenkinson M,
Beckmann CF, Behrens TEJ, Woolrich MW, Smith SM. FSL. NeuroImage
2012;62:782–790 doi: 10.1016/j.neuroimage.2011.09.015.
4. Andersson JLR,
Skare S, Ashburner J. How to correct susceptibility distortions in spin-echo
echo-planar images: application to diffusion tensor imaging. NeuroImage
2003;20:870–888 doi: 10.1016/S1053-8119(03)00336-7.
5. Andersson JLR,
Sotiropoulos SN. An integrated approach to correction for off-resonance effects
and subject movement in diffusion MR imaging. NeuroImage 2016;125:1063–1078
doi: 10.1016/j.neuroimage.2015.10.019.
6. Jenkinson M,
Smith S. A global optimisation method for robust affine registration of brain
images. Med Image Anal 2001;5:143–156 doi: 10.1016/S1361-8415(01)00036-6.
7. Landman BA,
Farrell JAD, Jones CK, Smith SA, Prince JL, Mori S. Effects of diffusion
weighting schemes on the reproducibility of DTI-derived fractional anisotropy,
mean diffusivity, and principal eigenvector measurements at 1.5T. NeuroImage
2007;36:1123–1138 doi: 10.1016/j.neuroimage.2007.02.056.