Atsushi Nakamoto1, Hiromitsu Onishi1, Takahiro Tsuboyama1, Takashi Ota1, Hideyuki Fukui1, Keigo Yano1, Kengo Kiso1, Toru Honda1, Shohei Matsumoto1, Mitsuaki Tatsumi1, Hiroyuki Tarewaki2, Yoshihiro Koyama2, Tetsuya Wakayama3, Xinzeng Wang4, and Noriyuki Tomiyama1
1Osaka University Graduate School of Medicine, Suita, Japan, 2Osaka University Hospital, Suita, Japan, 3GE Healthcare, Hino, Japan, 4GE Healthcare, Houston, TX, United States
Synopsis
Keywords: Prostate, Image Reconstruction, Deep learning reconstruction
The
image quality of high-resolution PROPELLER T2WI combined with deep learning
reconstruction (DLR) in prostate MRI was evaluated by the phantom and clinical
studies. In the phantom study, noise reduction was achieved by DLR, and spatial
resolution was remarkably improved in PROPELLER T2WI compared to Cartesian T2WI.
In the clinical study, DLR showed a significant improvement in signal-to-noise
ratio, and the qualitative analyses showed reduced image noise and improved
spatial resolution, with PROPELLER T2WI DLR images showing the highest overall image
quality. PROPELLER T2WI with DLR would be promising technique to improve the
image quality of T2WI in prostate MRI.
Introduction
T2-weighted
imaging (T2WI) is an essential sequence for the morphological evaluation of the
prostate and a dominant sequence in the diagnosis of transitional zone (TZ)
cancers1. Although Prostate Imaging Reporting and Data System
(PI-RADS) recommends a slice thickness of 3 mm and in-plane resolution of ≤0.7 ×
≤0.4 mm for T2WI1, it might be difficult to maintain a sufficient
signal-to-noise ratio (SNR) depending on the performance of the scanner. Recently,
deep learning reconstruction (DLR) has been introduced into clinical practice,
and its usefulness in improving SNR and reducing acquisition time has been
reported in prostate MRI2, 3. DLR has been newly available for
periodically rotated overlapping parallel lines with enhanced reconstruction
(PROPELLER) sequences. The purpose of this study was to evaluate the image
quality of high-resolution PROPELLER T2WI combined with DLR in prostate MRI.Methods
Phantom study:
A
phantom (Magphan® 170, The Phantom Laboratory, NY, USA) was scanned with a
3-Tesla scanner (SIGNA Architect, GE Healthcare, Waukesha, WI). T2WI was
scanned using two acquisition methods (Cartesian and PROPELLER) with various
matrix settings (512 × 512, 448 × 448, and 384 × 384). For Cartesian T2WI only,
an additional 448 × 280 matrix images were obtained, which are used in our daily
clinical practice. Other parameters are as follows: TR/TE, 5400/81.9–82.5 msec;
Echo train length, 14 (Cartesian)/32 (PROPELLER); Band width, 62.5 Hz; FOV, 20
× 20 cm; Slice thickness/gap, 3/0 mm; NEX, 1 (Cartesian)/1.5 (PROPELLER); Slice
number, 24. Both T2WI were reconstructed with conventional method (original
image) and DLR. A radiologist evaluated images of the high-resolution test plate,
which contains rectangular slots forming a test pattern which ranges from one
to eleven line pairs per cm (5mm to 0.45mm resolution), and assigned scores
regarding the spatial resolution using a 6-point scale (6, could completely
separate 11 lines/cm; 5, 10 lines/cm; 4, 9 lines/cm; 3, 8 lines/cm; and 2, 7
lines/cm; 1, < 7 lines/cm). Circular regions of interest (ROIs) were placed
on homogeneous areas within the phantom, and the standard deviation (SD) of the
signal was measured and employed as the index of image noise.
Clinical Study:
This
study included fourteen patients who underwent prostate MRI including conventional
Cartesian T2WI (C-T2WI) and PROPELLER T2WI (P-T2WI). The written informed
consent was obtained from all patients. The matrix sizes of C-T2WI and P-T2WI
were 448 × 280 and 512 × 512, respectively, and other parameters were the same
as those used in the phantom study. The
acquisition times for C-T2WI and P-T2WI were 4 min 9 sec and 3 min 58 sec,
respectively. A radiologist placed ROIs on the peripheral zone (PZ) and TZ of
the prostate, and the signal and SD of the signal were measured. The SNR was
calculated by dividing the signal by SD, and compared among 4 images (C-T2WI original,
C-T2WI DLR, P-T2WI original, and P-T2WI DLR) using Friedman’s test followed by
post hoc Wilcoxon signed rank test with Bonferroni correction (n = 6). As a
qualitative analysis, a radiologist who was blinded to the acquisition and
reconstruction methods reviewed images and assigned scores regarding the noise,
sharpness, spatial resolution, and overall image quality using a 5-point scale. Results
Phantom study:
For
both acquisition methods, the larger the matrix size, the more image noise was
present, and noise was reduced by using DLR (Figure 1, 2). The spatial
resolution of P-T2WI was obviously improved with DLR, and 11 lines/cm could be
separated in the 512 and 448 matrices, while C-T2WI showed mild improvement in
spatial resolution with DLR.
Clinical Study:
The
results of the quantitative analysis are summarized in Figure 3. DLR images
showed improved SNR compared to original images, with significant differences
in PZ and TZ for C-T2WI and TZ for P-T2WI (P < 0.01).
The
results of the qualitative analysis are summarized in Figure 4. Regarding
noise, scores for both C-T2WI and P-T2WI were significantly increased with DLR
(i.e., image noise was reduced) (P < 0.01). Regarding sharpness and
spatial resolution, scores for P-T2WI were higher than for C-T2SWI, with the
highest scores for P-T2WI DLR. P-T2WI DLR had the highest score for overall
image quality, followed by C-T2WI DLR (Figure 5).Discussion
Our
results showed that DLR reduced image noise and improved SNR in P-T2WI as well
as in C-T2WI. Furthermore, in the phantom study, DLR showed a more prominent
improvement in spatial resolution in PROPELLER than in Cartesian, and in the
clinical study, P-T2WI DLR showed the highest spatial resolution and image
quality. Therefore, the combination of PROPELLER and DLR would be a promising
technique for simultaneously improving the spatial resolution and image quality
of prostate MRI, thereby improving diagnostic performance. This is a preliminary
study using a relatively small number of cases, and further evaluation of the
clinical usefulness of this technique, including its detectability of prostate
cancers, will be needed using a larger number of cases.Conclusion
PROPELLER
T2WI combined with DLR resulted in image noise reduction and improved spatial
resolution. This technique enables high-resolution T2WI imaging with high image
quality and is expected to contribute to the improvement of the quality of
prostate MRI.Acknowledgements
No acknowledgement found.References
1. Turkbey
B, Rosenkrantz AB, Haider MA, et al. Prostate Imaging Reporting and Data System
Version 2.1: 2019 Update of Prostate Imaging Reporting and Data System Version
2. Eur Urol. 2019;76(3):340-351.
2. Wang X,
Ma J, Bhosale P, et al. Novel deep learning-based noise reduction technique for
prostate magnetic resonance imaging. Abdom Radiol (NY). 2021;46(7):3378-3386.
3. Park
JC, Park KJ, Park MY, Kim MH, Kim JK. Fast T2-Weighted Imaging With Deep
Learning-Based Reconstruction: Evaluation of Image Quality and Diagnostic Performance
in Patients Undergoing Radical Prostatectomy. J Magn Reson Imaging. 2022;55(6):1735-1744.