Can Wu1, Qi Peng2, Ramin Jafari3, Yansong Zhao3, Victoria Yu1, and Ricardo Otazo1,4
1Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 2Department of Radiology, Albert Einstein College of Medicine and Montefiore Medical Center, Bronx, NY, United States, 3MR Clinical Science, Philips Healthcare, Cambridge, MA, United States, 4Department of Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, United States
Synopsis
Keywords: Liver, Relaxometry
An
efficient diamond radial sampling strategy was proposed for free-breathing 3D
T1ρ abdominal imaging at 3T. The phantom experiment shows that diamond radial
sampling provides T1ρ measurement values comparable to that of 3D Cartesian and
radial stack-of-stars sampling. In-vivo volunteer studies illustrate that
diamond radial sampling is superior to Cartesian sampling, where image quality
is significantly compromised by breathing motion artifacts. In addition, scan
time can be drastically reduced using the fast MAPSS method. This work
demonstrates the feasibility of quantitative free-breathing 3D T1ρ imaging with
diamond radial sampling in the abdomen.
INTRODUCTION
T1ρ
imaging has been mostly employed to studying cartilage due to its high
sensitivity to low-frequency interactions between water molecules and their
local macromolecular environment [1]. The application of T1ρ imaging in the
abdomen is more challenging due to respiratory motion, field inhomogeneities within
a large field of view, and unavailability of localized transmit/receive coils. Most
of the previous work on abdominal T1ρ imaging were performed under breath-holds
[2-4], which leads to low spatial resolution and limited volumetric coverage. Radial
sampling is inherently less sensitive to respiratory motion compared to
Cartesian sampling and 3D radial stack-of-stars sampling was recently used to
acquire data for each golden angle after T1ρ preparation in human liver [5].
However, data acquisition is slow (4:41 min for each TSL). The aim of this work
was to develop a fast free-breathing 3D abdominal T1ρ imaging technique using
efficient diamond radial sampling in combination with the state-of-the-art magnetization-prepared
angle-modulated partitioned k-space spoiled gradient echo snapshots (MAPSS)
sequence [6].METHODS
3D
T1ρ Sequence:
The 3D T1ρ sequence consists of four components (Figure 1A): fat suppression
(SPAIR), T1ρ preparation, 3D readout, and T1 recovery. RF pulse flip angle
sweeping and T1ρ phase cycling (with and without an adiabatic inversion pulse)
are used for 3D MAPSS [7]. Three different k-space trajectories are used for 3D
readout (Figure 2) with low-high profile ordering: Cartesian, stack-of-stars
golden-angle radial (SOS-GAR), and diamond pseudo-golden-angle radial (D-pGAR).
The sampling density along kz is constant for SOS-GAR but is decreased from the
center (k0) to the peripheral k-space (kmax,
kmin) for D-pGAR.
Phantom
Study: The
3D T1ρ sequences were first tested on a phantom which consists of three pairs
of tubes with three different agarose concentrations (2%, 3%, and 4%) embedded
in a cylindrical plastic holder (The Phantom Laboratory, Greenwich, NY). The
phantom scan parameters are shown in Figure 1B.
Volunteer
Study: Free-breathing
3D T1ρ abdominal imaging was performed on a healthy volunteer (male, 35 years
old) with Cartesian and D-pGAR sampling. SOS-GAR was not performed due to the
prohibitively long scan time. The volunteer scan parameters were similar to the
phantom experiments, except the following: FOV=360×360×200 mm3,
voxel size = 1.5×1.5×4.0 mm3, TR/TE = 5.6/2.5 ms for Cartesian and 5.6/2.2 ms for D-pGAR, SENSE = 1.4×1.8 (ky×kz) for Cartesian and 1.5 (kz) for
D-pGAR, scan time per TSL = 0:52 min for Cartesian and 2:26 min for D-pGAR. All
experiments were performed on a clinical 3T MRI scanner (Ingenia Elition X,
Philips Healthcare) using anterior and posterior coils.
T1ρ
Map and Analysis:
T1ρ maps were generated using an in-house software in Interactive Data Language
(ExelisVis, Boulder, CO) and the fitting methods for traditional MAPSS (8 TSLs)
and fast MAPSS (4 TSLs or 3 TSLs) were reported previously [8]. Interleaved
phase cycling starting with a positive phase (e.g., +0, -15, +30, -45 ms) was
used for the fast MAPSS methods. Five ROIs (liver, pancreas, spleen, anterior
muscle, and posterior muscle) were manually drawn on a selected slice to
evaluate the T1ρ values of these anatomical structures in the volunteer.RESULTS
Quantitative
T1ρ maps of the phantom are shown in Figure 3. The T1ρ maps generated from
SOS-GAR and D-pGAR present high concordance to the ones using Cartesian
sampling. The fast MAPSS methods (4 TSLs and 3 TSLs) show similar T1ρ maps compared
to the reference MAPSS method with 8 TSLs. Table 1.1 summarizes the
quantitative T1ρ values of the six tubes in the phantom experiments. The
results further confirm that the T1ρ measurements were comparable among the three
sampling methods. Figure 4 illustrates the magnitude image (TSL = 0 ms) and the
corresponding T1ρ maps of the volunteer. The image quality and T1ρ maps of
Cartesian sampling was significantly degraded by breathing motion artifacts,
which was drastically improved using D-pGAR sampling. The T1ρ values of the
five ROIs are shown in Table 1.2.DISCUSSION
Free-breathing
abdominal 3D T1ρ imaging with Cartesian sampling is significantly compromised by
breathing motion artifacts if no triggering or gating method is used. This work
demonstrated that the proposed efficient diamond radial sampling approach
(D-pGAR) significantly improved image quality and T1ρ quantification compared
to Cartesian sampling. The phantom experiment demonstrated that T1ρ
measurements from D-pGAR sampling were comparable to that of the traditional
Cartesian and SOS-GAR sampling methods. This work further supports the use of fast
MAPSS with only 4 TSLs or 3 TSLs, which provides similar T1ρ results compared
to the reference MAPSS with 8 TSLs [8]. Fast MAPSS sequences can significantly
reduce scan time for T1rho mapping (i.e., the total scan time is reduced from 19
minutes to 7 minutes if only 3 TSLs are used). Although the use of D-pGAR
sampling was demonstrated for 3D T1ρ imaging, it can be similarly used for
other quantitative parameter mapping with a different magnetization preparation
module, such as T2 mapping, chemical exchange saturation transfer imaging, and magnetization
transfer imaging.CONCLUSION
Free-breathing
3D T1ρ abdominal imaging with clinically feasible scan times was demonstrated
with the use of an efficient diamond radial sampling strategy. The combination
of diamond radial sampling and fast MAPSS acquisition would enable efficient quantitative
evaluation of tissue properties in abdominal organs.Acknowledgements
The work was supported by NIH Grant R01-AR076328.References
1. Li X, Ma B, Link TM,
Castillo D, Blumenkrantz G, Lozano J, Carballido-Gamio J, Ries M, Majumdar S.
In vivo T1rho and T2 mapping of articular cartilage in osteoarthritis of the
knee using 3 Tesla MRI. Osteoarthritis Cartilage 2007; 15(7): 789-797.
2. Chen W, Chan Q, Wang Y.
Breath-hold black blood quantitative T1rho imaging of liver using single shot
fast spin echo acquisition. Quant Imaging Med Surg 2016; 6(2): 168-177.
3. Wang Y, Deng M, Lin J,
Kwok, Liu E, Chen W. Age- and gender-associated liver physiological T1rho
dynamics demonstrated with a clinically applicable single-breathold
acquisition. SLAS Technol 2018; 23(2): 179-187.
4. Wang Y, Deng M, Lo GG,
Liang D, Yuan J, Chen W. Breath-hold black-blood T1rho mapping improves liver
T1rho quantification in healthy volunteers. Acta Radiol 2018; 59(3):
257-265.
5. Sharafi A, Baboli R,
Zibetti M, Shanbhogue K, Olsen S, Block T, Chandarana H, Regatte R. Volumetric
multicomponent T1ρ relaxation mapping of the human liver under free breathing
at 3T. Magn Reson Med 2020; 83(6):2042-2050.
6. Li X, Han ET, Busse RF,
Majumdar S. In vivo T1ρ mapping in cartilage using 3D magnetization-prepared
angle-modulated partitioned k-space spoiled gradient echo snapshots (3D MAPSS).
Magn Reson Med 2008; 59(2): 298-307.
7. Peng G, Wu C, Peng Q.
Improved phase-cycling preparations in quantitative T1rho mapping. Proc Intl
Soc Mag Reson Med 2022; (30): 4063.
8. Peng Q, Wu C, Kim J, Li X.
Efficient phase-cycling strategy for high-resolution 3D gradient-echo
quantitative parameter mapping. NMR Biomed 2022; 35(7): e4700.