Bogdan Dzyubak1, Kay M Pepin1, Jun Chen1, Yuan Le1, Kyle Kalutkiewicz2, Roger Grimm1, Jeremy A Heilman1, Scott Kruse1, Jennifer Kugel1, Meng Yin1, Kevin J Glaser1, and Ehman L. Richard1
1Radiology, Mayo Clinic, Rochester, MN, United States, 2Engineering, Resoundant, Rochester, MN, United States
Synopsis
Keywords: Liver, Elastography
Magnetic Resonance Elastography is a
technique used to effectively stage liver fibrosis by measuring tissue
stiffness. This work optimized several acquisition and processing parameters, and
performed a comprehensive repeatability/reproducibility validation across the
three types of MRE (GRE, EPI, 3DEPI), 1.5T and 3T field strengths, and the
three major manufacturers (GE, Siemens, Philips).
When using the optimized methods and
automated analysis, the repeatability coefficients were 11.8% for GRE, 10.8%
for EPI, and 7.7% for 3DEPI, all superior to QIBA’s 19% benchmark. Agreement
between manufacturers was within 2%.
Introduction
Regulatory-approved
implementations of MR Elastography (MRE) are now widely deployed at more than
2000 locations globally. The main
clinical application is for diagnosis and staging of liver fibrosis and MRE is
now regarded as the most accurate non-invasive technique for this important
clinical task 1. MRE is
increasingly being used longitudinally in patients to follow the progression of
disease and to monitor treatment. The repeatability coefficient
(RC) 2 is used to indicate the level at which a biomarker change
is 95% likely to be caused by biology rather than measurement variation. A test
with a small RC can be used to detect a smaller biological change.
The
Consensus Profile for MRE of the liver, developed by the Quantitative Imaging Biomarker Alliance (QIBA) is an
international standard protocol, and the resulting repeatability performance
that can be achieved 3. Based on a meta-analysis of multiple studies, the QIBA
profile indicates that if properly performed, the RC of 2D liver MRE is
19%. This is a very useful level of
precision because fibrosis causes large changes in liver stiffness (liver
stiffness can more than double with severe fibrosis).
Given recent technical advances in
sequences and protocols for 2D GRE and SE-EPI MRE, wider availability of 3D vector
MRE, and the development and extensive testing of an automated analysis tools
for MRE, the goal of this study was to re-evaluate the test-retest
repeatability of liver MRE on multiple platforms. A second aim was to provide further evidence
that if implemented in a consistent fashion, MRE-based liver stiffness
measurements obtained at different field strengths and on MRI systems from
different manufacturers will be comparable.
In this study, Then, using the optimal parameter
set, we used a comprehensive approach to investigate the repeatability across
major manufacturers (GE, Siemens, and Philips), field strengths (1.5 and 3T),
and acquisition strategies (2D GRE and EPI MRE, and 3D EPI MRE). All stiffness
measurements were performed using an automated method which allowed us to
eliminate the effects of reader variability and bias and narrow the study to
technology-related effects. Methods
As a first step, we investigated parameter permutations for
acquisition and processing methods to improve repeatability of MRE in phantoms
and healthy volunteers. Parameters that were evaluated included resolution,
slice thickness, fractional encoding, confidence threshold, and analysis volume
for 3D MRE. When comparing sequence types, GRE MRE with QIBA profile settings was
used as the baseline.
The optimized acquisition parameters (Table 1) were then used
for the cross-platform repeatability study. Five volunteers were scanned on each
combination of the following: manufacturer (GE, Siemens, Philips), field
strength (1.5T, 3T), and sequence type (GRE, EPI, 3D EPI). Each volunteer was
scanned twice without repositioning on two separate days, for a total of 360
MRE exams. Liver stiffness values were calculated using an automated method
without manual modification 4. The repeatability coefficient was calculated for each type
of MRE sequence across all subjects and platforms 2. A percentage stiffness difference was calculated between
manufacturers. Results
The results are summarized in Table
2. Automated analysis was successful in
all exams. The repeatability coefficient
was 11.8% for 2D GRE MRE, 10.8% for 2D SE-EPI MRE, and 7.7% for 3D vector SE-EPI
MRE. The inter-platform agreement was excellent, systematic with bias values
ranging between 0.2 % and 1.2%
Calculating stiffness from ROIs in 4 slices spaced
out to sample a 2D-like 4-cm volume yielded stiffness within 1% of 12-slice and
20-slice measurements. Confidence thresholding the 96x96 2D EPI data at 0.98,
and 3D MRE at 0.92 was deemed by an expert reader to produce masks equivalent
to the GRE threshold (Figure 4). Discussion
Ongoing incremental advances in clinical liver MRE protocols
and analysis methods can be expected to steadily improve test-retest
repeatability, which is a key performance factor for following the disease
process and assessing the effect of treatment.
The RC of 3D vector MRE demonstrated a further substantial improvement
over 2D MRE, which likely reflects the superiority of fully processing the 3D
wavefield.
The repeatability results obtained in this study
were obtained under ideal conditions - in healthy subjects and in short exams
that that only included the MRE acquisitions.
While it likely that the advances will boost the repeatability of liver
MRE beyond the current QIBA profile claim, results obtained in patient exams and
long protocols are not expected to achieve the same level of performance. The results obtained in this study should be
regarded as a performance target and provide motivation for further studies to
update the test-retest performance in clinical settings.Conclusions
Parameters
for GRE, EPI, and 3D EPI MRE sequences were optimized and validated for
repeatability and reproducibility on 3 platforms and 2 field strengths. When
using the optimized parameters together with automated analysis, this study demonstrated
progress in improving the repeatability of liver MRE and the opportunity to
boost the repeatability claim of the current QIBA standard for liver MRE. The
repeatability of 3D vector MRE was substantially higher than that of 2D MRE. The
results also demonstrated excellent agreement in liver stiffness across field
strengths and manufacturer platforms. Acknowledgements
NIH R37 EB001981. The authors gratefully
acknowledge the assistance of Bradley Bolster of Siemens Heathineers, Sandeep
Ganji of Philips Healthcare, and Daniel Rettmann of GE Healthcare in conducting
this research.References
1.Pepin, K.M., Welle, C.L.,
Guglielmo, F.F., et al.: ‘Magnetic resonance elastography of the liver:
everything you need to know to get started’, Abdom Radiol, 2022, 47, (1), pp.
94-114
2.Raunig, D.L., McShane,
L.M., and Pennello, G.: ‘Quantitative imaging biomarkers: A review of
statistical methods for technical performance assessment’, Statistical Methods
in Medical Research, 2015, 24, (1), pp. 27-67
3.QIBA: ‘QIBA Profile:
Magnetic Resonance Elastography of the liver’, 2018
4.Dzyubak, B.,
Venkatesh, S., Manduca, A., et al.: ‘Automated liver elasticity calculation for
MR elastography’, J Magn Reson Imag, 2016, 43, (5), pp. 1055-1063
5.Trout, A.T., Serai,
S., Mahley, A., et al.: ‘Liver stiffness Measurements with MR Elastography:
Agreement and Repeatability across Imaging Systems, Field Strengths, and Pulse
Sequences’, Radiology, 2016, 281, (3), pp. 793-804
6.Shire, N.J., Yin, M.,
Chen, J., et al.: ‘Test-retest repeatability of MR elastography for noninvasive
liver fibrosis assessment in Hepatitis C’, Journal of Magnetic Resonance Imaging,
2011, 34, pp. 947-955
7.Serai, S., Yin, M.,
Wang, H., et al.: ‘Cross-vendor validation of liver magnetic resonance
elastography’, Abdom Imaging, 2015, 40, (4), pp. 789-794
8.Kim, J.H., Kim, B., Yu, H.J., et al.:
‘Reproducibility of hepatic MR elastography across field strengths, pulse
sequences, scan intervals, and readers’, Abdom Radiol, 2020, 45, (1), pp.
107-115