Xiaobing Fan1, Aritrick Chatterjee1, Zhen Ren1, Aytekin Oto1, and Gregory S. Karczmar1
1Radiology, The University of Chicago, Chicago, IL, United States
Synopsis
Keywords: Quantitative Imaging, Cancer
The
extended Tofts model (ETM) requires calculation of contrast agent concentration
in tissue as function of time (C(t)) using a non-linear model that results in
error propagation. Here, we introduce a signal intensity (S(t)) form of ETM (SI-ETM)
without calculating C(t). QIBA DCE-MRI data was used to validate the SI-ETM,
and then human prostate DCE-MRI data were analyzed to compare physiological
parameters calculated from the ETM and the SI-ETM. The parameters calculated
from S(t) were strongly correlated with the values calculated from C(t). Bland–Altman
analysis showed good agreement between the parameters calculated from the ETM and
the SI-ETM.
INTRODUCTION
Quantitative analysis of
dynamic contrast enhanced (DCE) MRI provides valuable information for detection
and diagnosis of cancers. The standard Tofts model (TM) and extended Tofts
model (ETM) are the most common pharmacokinetic models used to extract
physiological parameters (Ktrans, ve and vp)[1-3].
However, use of pharmacokinetic models requires calculation of contrast agent
concentration in tissue as a function of time (C(t)) based on T1-weighted
signal intensity (S(t)). The C(t) can be calculated by using the gradient echo
signal equation (non-linear model) with pre-contrast tissue T1 values[4];
or by using the ‘reference tissue’ model under a simple linear approximation[5].
The C(t) calculated from the non-linear model is more accurate than the
‘reference tissue’ model, but its precision is strongly influenced by the
native T1 values. Measurements of pre-contrast tissue T1 values also contribute
error to calculations of C(t) using non-linear model.
To address this problem, Fan
et al. developed a signal intensity form of standard Tofts model[6].
In this study, the signal intensity form of extended Tofts model (SI-ETM) was developed and validated using simulated
DCE-MRI data from the Quantitative Imaging Biomarkers Alliance (QIBA)[7].
Furthermore, the physiological parameters calculated from the ETM using S(t)
were compared with results obtained using C(t) for human prostate DCE-MRI data.THEORY and METHODS
Changes of C(t) in tissue
following bolus contrast agent injection are described by the standard TM of
DCE-MRI. Based on a ‘reference tissue’ model for calculating C(t), Fan et al.
developed signal intensity form of standard TM as follows[6]:
$$S_r(t)=\lambda\frac{S_b(0)}{S(0)(1-Hct)}K^{trans}\int_{0}^{t}S_p(\tau)exp[-\lambda(t-\tau)K^{trans}/v_e]d\tau,-----(1)$$
where Ktrans
is the volume transfer constant between blood plasma and extravascular
extracellular space (EES), ve is the volume of EES per unit volume
of tissue, 𝜆=T1blood/T1tissue is ratio of T1
in blood to T1 in tissue, Hct is the hematocrit, $$$S_r(t)=\frac{S(t)-S(0)}{S(0)}$$$ and $$$S_p(t)=\frac{S_b(t)-S_b(0)}{S_b(0)}$$$, Sb(t) is signal intensity in
blood, S(0) and Sb(0) are signal intensities at baseline for tissue
and blood, respectively. Here we introduce the SI-ETM as follows:
$$S_r(t)=\lambda\frac{S_b(0)}{S(0)(1-Hct)}\left[K^{trans}\int_{0}^{t}S_p(\tau)exp[-\lambda(t-\tau)K^{trans}/v_e]d\tau+v_pS_p(\tau)\right],-----(2)$$
where vp is the volume fraction
of the plasma space.
The SI-ETM model was
validated with DCE-MRI data provided by the QIBA. The parameters calculated
from the ETM were compared with the values calculated from the SI-ETM.
As a clinical application
of the SI-ETM, eighteen patients with biopsy-confirmed prostate cancer were
included in this IRB-approved study. MRI data were acquired on a Philips
Achieva 3T-TX scanner. After the required clinical MRI scans, DCE 3D-T1-FFE
data were acquired pre- and post-contrast media injection (0.1 mmol/kg DOTAREM;
TR/TE=3.5/1.0 ms, FOV=180×180 mm , matrix size =160×160, flip angle=10°, slice
thickness=3 mm, typical number of slices=24, SENSE factor=3.5, half scan
factor=0.625) for 150 dynamic scans with typical temporal resolution of 2.2
s/image (1.0-4.3 s).
Regions-of-interest
(ROIs) for prostate cancer, normal tissue in different prostate zones were
drawn on T2W images and transferred to DCE images. ROI’s for blood vessels were
manually traced on the iliac artery on a slice with cancer. For each ROI, the
average S(t) was calculated, and then C(t) was calculated from the non-linear
model.
Normalized
root-mean-square-errors (NRMSE) were calculated between data and corresponding
fits. Pearson’s correlation coefficient was calculated between parameters
obtained from C(t) and S(t). Bland-Altman analysis was performed to evaluate
the agreement of the ETM and SI-ETM in calculating parameters. Receiver
operating characteristic (ROC) analysis was used to evaluate performance in
differentiating cancer from normal tissue.RESULTS
For the QIBA DCE-MRI data, Figure
1 shows comparisons of fits C(t) and Sr(t) between the ETM and SI-ETM.
There were small errors in fits from the SI-ETM model. Figure 2 shows comparisons
of Ktrans, ve, and vp maps obtained between the
ETM and SI-ETM. The scatter plots (Fig. 3(a-c)) shows there were strong
correlations (r>0.85, p<0.0001) for parameters obtained between C(t) vs. Sr(t).
Bland-Altman analysis (Fig. 3(c-d)) shows good agreement for parameters obtained
between C(t) vs. Sr(t).
For human prostate DCE-MRI data, Figure
4 shows plots of C(t) and Sr(t) as well as corresponding fits from the ETM and
SI-ETM for a 66-year-old patient over ROIs of cancer and normal prostate
tissue. There was no difference in average NRMSE (0.046 vs. 0.047) between fits
of the ETM and SI-ETM. Finally for all 18 patients, Figure 5 (a-c) shows scatter
plots of Ktrans, ve, and vp obtained between
the ETM and SI-ETM. There was strong correlation (r>0.85, p< 0.0001)
between parameters calculated between these two methods. The corresponding
Bland–Altman plots (d-f) shows good agreement between parameters calculated
between the ETM and SI-ETM. ROC analysis yielded area under the curve of 0.760
and 0.805 for Ktrans obtained between the ETM and SI-ETM,
respectively.DISCUSSION
The SI-ETM was developed
and validated by the QIBA DCE-MRI data. The physiological parameters calculated
from the ETM and SI-ETM were very similar and provided equal power in differentiation
between cancer and normal tissue. The main advantage of using the SI-ETM model
with Sr(t) is that it avoids error propagation associated with
calculation of C(t). Implementation of the SI-ETM in clinical practice may
facilitate quick estimation of physiological parameters.CONCLUSION
The SI-ETM of DCE-MRI could
be used as alternative of the ETM in calculating physiological parameters when
there was difficulty to calculate tissue contrast agent concentrations.Acknowledgements
This
research is supported by National Institutes of Health (R01 CA172801-01, R01
CA218700-01, and 5U01 CA142565-09).References
[1] Tofts PS. J Magn Reson Imaging. Modeling tracer kinetics
in dynamic Gd-DTPA MR imaging. 1997;
7(1):91-101.
[2] Tofts PS, Brix G,
Buckley DL, Evelhoch JL, Henderson E, Knopp MV, Larsson HB, Lee TY, Mayr NA,
Parker GJ, Port RE, Taylor J, Weisskoff RM. Estimating kinetic
parameters from dynamic contrast-enhanced T(1)-weighted MRI of a diffusable
tracer: standardized quantities and symbols. J Magn
Reson Imaging. 1999; 10(3):223-32.
[3] Sourbron SP, Buckley DL. On
the scope and interpretation of the Tofts models for DCE-MRI. Magn Reson Med.
2011;66(3):735-45.
[4] Dale BM, Jesberger
JA, Lewin JS, Hillenbrand CM, Duerk JL. J Magn Reson Imaging. Determining and
optimizing the precision of quantitative measurements of perfusion from dynamic
contrast enhanced MRI. 2003;
18(5):575-84.
[5] Medved M, Karczmar G,
Yang C, Dignam J, Gajewski TF, Kindler H, Vokes E, MacEneany P, Mitchell MT,
Stadler WM. Semiquantitative analysis
of dynamic contrast enhanced MRI in cancer patients: Variability and changes in
tumor tissue over time. J Magn
Reson Imaging. 2004; 20(1):122-8.
[6] Fan X,
Chatterjee A, Medved M, Oto A, Karczmar GS. Signal intensity
form of the Tofts model for quantitative analysis of prostate dynamic contrast
enhanced MRI data. Phys Med Biol. 2021 Jan
22;66(2):025002.
[7]
Quantitative Imaging Biomarkers Alliance (QIBA) - ‘QIBA_v4 Extended Tofts’. https://sites.duke.edu/dblab/qibacontent/.