Bart de Vos1, Rob Remis2, and Andrew Webb1,2
1C.J. Gorter MRI Center, Leiden University Medical Center, Leiden, Netherlands, 2Circuits and Systems, Delft University of Technology, Delft, Netherlands
Synopsis
Keywords: Low-Field MRI, Artifacts, Concomitant fields
Low field Halbach systems are
susceptible to concomitant gradient effects. The associated fields are
significantly different from clinical systems and are presented
here for the first time. The corresponding equations and associated effects are
verified using a spin echo sequence on a 46 mT Halbach system where a gradient
strength of 10 mT/m on a phantom with maximum dimension of 200 mm is observed
to create distortions. These are most evident in the transverse plane, where the
phase encoding gradient causes blurring and overlapping the other transverse
gradient lobe results in an additional warping effect.
Introduction
Magnetic field gradients have undesired but unavoidable
vector-components known as concomitant fields. Their effects have been
extensively studied at clinical field strengths1, where distortions
occur for large field of views2 and when using sequences, such as
spiral, EPI and RARE4-5, which are sensitive to phase errors. At low
field (<0.6T), gradient fields for a given field of view become more
significant with respect the background field. Therefore, distortions are to be
expected for smaller field of views and using sequences such as conventional spin
and gradient echo6. For low-field point-of-care (POC) systems, the
effects of concomitant fields have not yet been extensively modelled: notably
the cylindrical gradient coils designed for a transverse B07
create concomitant fields which are geometrically different than in clinical systems.
In this work the concomitant fields associated
with a Halbach-array-based system are derived. The effects were simulated and
verified by acquiring spin-echo images on a 46mT system.Methods
Maxwell’s equations require that in a
source-free region the magnetic flux density must be curl and divergence free. Consequently,
the components of the magnetic flux density can be expressed with the following
matrix vector product1
$$\begin{bmatrix}B_x\\B_y\\B_z-B_0 \end{bmatrix}=\begin{bmatrix}-\alpha G_z&0&G_x\\0&(\alpha -1)G_z &G_y\\G_x&G_y&G_z\end{bmatrix}\begin{bmatrix}x\\y\\z\end{bmatrix},\quad(1)$$
where B0 is in the $$$z$$$-direction. For a Halbach system $$$x$$$ represents
the axial, and $$$y$$$ the vertical,
direction. In (1) the $$$\alpha$$$-term is determined by the gradient coil geometry. Conventional
systems have $$$\alpha=0.5$$$: in contrast, numerical simulations of our gradient
coils reveal that $$$\alpha=0$$$.
The magnitude of the magnetic flux density is required
to determine the phase error and deflection from the main concomitant field component.
This is evaluated using a Taylor expansion:
for Halbach systems this has the following form
$$|\textbf{B}(\textbf{r},\textbf{g})|\approx B_0+\textbf{g} \cdot \textbf{r}+\frac{1}{2B_0}\left[ (G_y^2+ G_x^2)z^2+G_z^2 y^2-2G_yG_z yz \right] + \frac{1}{2B^2_0}[..., \quad (2)$$
where $$$B_0+\textbf{g}\cdot\textbf{r}$$$ are the desired components for imaging, and
the remaining terms describe the concomitant fields. The majority of the concomitant
field effects are represented by the $$$\frac{1}{2B_0}$$$ terms, i.e. three self-squared and a single cross-term. The effects of these terms for
a spin-echo sequence were simulated in Matlab using a phantom of multiple
tubes: experimental results were obtained using a corresponding physical
realization of doped water-filled tubes. Additional experimental results were
obtained from a 1cm thick (transverse)
brain slab phantom with dimensions 180x200mm and compartments containing
agarose doped with CuSO4, such that relaxation times correspond to brain
tissue at 50mT8.
The 46mT system has a maximum gradient strength of 10mT/m,
and so to investigate the effects of stronger gradients an additional bipolar
pulse, which only creates concomitant phase is added (Fig.1). Fig.1A. shows the
pulse sequence used to depict the effect of the self-squared-terms. In Fig.1B.
the dephasing and phase gradient lobes are overlapped to add the effect of the cross-term.
Results
The tube phantom simulations were compared with
measurements, the results are shown in Fig.2. Equation (2) shows that when using $$$z$$$ as the
phase encoding gradient, blurring is expected to occur for large values of $$$|y|$$$: this is confirmed by the significant blurring
visible around the top and bottom tubes. In addition, the cross-term causes an
additional warping effect visible in the right column images. A 1D profile of
the tubes shows close correspondence between the simulation and measurement in
terms of broadening and warping.
Fig.3. shows images of the brain phantom oriented
in three planes using the different encoding gradients. Each plane is compared
to a reference image (left column). The red arrows show where blurring due to concomitant
gradients is expected, the green arrows where no blurring should occur. The
concomitant fields are expected to be constant in the $$$x$$$-direction, this is visible from the absence of distortions in the $$$xy$$$-plane (second row). Fig.4. shows that
overlapping the $$$y$$$- and $$$z$$$-gradient (de)phase-lobes adds a warping effect
with respect to the reference image, confirming the single cross-term in
equation (2). The other gradients can be overlapped without additional effects.
Fig.5. shows the brain phantom for a maximum
phase encoding gradient strength of 5mT/m, and “artificial” strengths of 10 and
15mT/m. The left column shows the effect of the self-squared-terms and the
right column the addition of the cross-term. Blurring and warping become
visible at 10mT/m as indicated by the red arrows. Discussion
This work presents first
analysis and demonstration of the effects of concomitant gradient fields on the
image quality of low-field POC systems. The results obtained at 46mT show that
the effects produce blurring for head-sized objects, particularly in the areas
close to the surface in the phase-encoding directions. This is contrary to
clinical MRI systems where the effects are more pronounced at the extremities
in a direction along the magnet bore. An additional difference is that
gradients for Halbach systems have only one, rather than two, cross-terms.
The concomitant field
effects were evident from a gradient strength of 10mT/m upwards for spin-echo
sequences. We anticipate that for more practical sequences such as RARE, phase
accumulation throughout the echo train will mean that distortions may occur at lower
gradient strengths. Further analysis will also allow us to develop different
correction strategies, such as applying phase correction before reconstruction,
or appropriate variations of quadratic nulling5 to resolve the
effects of self-squared terms.
Acknowledgements
This work was supported Simon Stevin Meester PrizeReferences
1. Bernstein MA, King FK, Xiaohong JZ. Concomitant-Field Correction Gradients, in: Handbook of MRI Pulse Sequences, pp:292-304
2. Zhou XJ, Du YP, Bernstein MA, et al. Concomitant Magnetic-Field induced Artifacts in Axial Echo Planar Imaging. MRM. 1998;39:596-605.
3. Bernstein MA, Zhou XJ, Polzin JA, et al. Concomitant Gradient terms in Phase Contrast MR: Analysis and Correction. MRM. 1998;39:300-308
4. King KF, Ganin A, Zhou XJ, Bernstein MA. Concomitant Gradient Field Effects in Spiral Scans. MRM 1999;41:103-112
5. Zhou XJ, Tan SG, Bernstein MA. Artifacts Induced by Concomitant Magnetic Field in Fast Spin-Echo Imaging. MRM 1998;40:582-591.
6. Volegov PL, Mosher JC, Espy MA, Kraus RH. On concomitant gradients in low-field MRI. JMR 2005;175:103-113.
7. de Vos B, Parsa J, Abdulrazaq Z, et al. Design, Characterisation and Performance of an Improved Portable and Sustainable Low-Field MRI System. Front in Phys 2021
8. O’Reilly T, Webb AG, In vivo T1 and T2 relaxation time maps of brain tissue, skeletal muscle, and lipid measured in healthy volunteers at 50 mT, MRM 2021;87:884-895.