Takayuki Sada1, Hajime Yokota2, Ryuna Kurosawa1, Takafumi Yoda1, Keisuke Nitta1, Koji Matsumoto1, Takashi Namiki3, Masami Yoneyama3, Adam Wu4, Yoshitada Masuda5, and Takashi Uno2
1Department of Radiology, Chiba University Hospital, Chiba, Japan, 2Diagnostic Radiology and Radiation Oncology, Graduate School of Medicine, Chiba University, Chiba, Japan, 3Philips Japan, Tokyo, Japan, 4Philips Healthcare (Shanghai) Ltd., Shanghai, China, 5Chiba University Hospital, Chiba, Japan
Synopsis
Keywords: Nerves, Diffusion Tensor Imaging
The brachial plexus is difficult to obtain
diffusion tensor image (DTI) with good image quality due to the inhomogeneity
of the magnetic field and motion artifacts. Zoom imaging based on 2D RF (iZoom)
could both reduce the distortion dramatically and has better fat suppression
effects. In this study, we compared DTI using conventional Zoom with that using
iZoom for the brachial plexus and diffusion tractography with iZoom created
higher mean length and fiber count than the conventional Zoom sequence,
suggesting its feasibility as a new method for diffusion tensor imaging for the
brachial plexus.
Introduction
Diffusion tensor imaging (DTI) is a sequence
that uses multiple diffusion-weighted images (DWI) with different directions of
motion probing gradient (MPG). Since this sequence can produce anisotropy of
water molecules, it can image nerve fibers indirectly. Although it has been used
mainly in the brain, it has recently begun to be applied for the evaluation of
peripheral nerves such as the sciatic nerve and brachial plexus.1
The brachial plexus is an area that is easily affected by peripheral neuropathy,
so it is often the target of DTI. However, it is difficult to obtain DTI with
good image quality due to the inhomogeneity of the magnetic field and motion
artifacts.
The reduced FOV (Zoom) method has been
reported to be effective in reducing distortion.2 3 However, this
method uses non-coplanar 90 and 180 selection slices with outer volume
suppression, which theoretically suffers from crosstalk artifacts from
neighboring slices. To avoid such artifacts, Zoom is basically applying
interleaved scan packages for odd and even number slices, resulting in doubled
scan time if the TR is kept constant. On
the other hand, iZoom applies a tilted 2D Echo-Planar RF excitation with only
tilting the k-space along the phase-encoding direction. iZoom is not suffering
from crosstalk artifacts thanks to tilted 2D RF excitation, separation of scan
packages is not needed, resulting in shorten the scan time compared to
conventional Zoom. Furthermore, iZoom inherently has better fat suppression
effects, is allows to use weaker (lower flip-angle) fat suppression pre-pulse,
which reduces the magnetization transfer effects, hence leads to improve the
signal-to noise ratio (SNR).456
In this study, we compared DTI using conventional Zoom
with that using iZoom for the brachial plexus and investigated the feasibility
of iZoom in the peripheral nerve.Methods
Nine volunteers: 2
women (age 26 ± 2.5 years) and 7 men (age 28.7 ± 2.8 years) with
electrophysiological studies demonstrating the absence of peripheral neuropathy
were examined on 3T MR unit (Ingenia 3T, Philips Healthcare). Scan parameters are
shown in Figure 1. Diffusion-weighted axial images were acquired with a
16-channel torso array coil using the conventional Zoom and iZoom methods,
respectively.
1. Tractography measurement
Fiber tracking was performed for nerve
roots C6 , C7and spinal cord (using DTI FiberTrack application provided by
Philips). The mean length (the average length in mm for all streamlines
belonging to a fiber tract) and fiber count (the number of DTI streamlines
extracted for a fiber tract) were compared. The settings for fiber tracking
were as follows: maximum angle = 20°, minimum FA = 0.2.
2. Visual assessment
Two independent
blinded readers compared Zoom and iZoom by evaluating four ordinal (1 worst to
5 best) image quality criteria (distortion artifacts, ghosting and motion
artifacts, quality of fat suppression, overall image). (Fig. 2)
Tractography
measurement and visual assessment between Zoom and iZoom were compared using
Wilcoxon signed-rank test.Results
1. Mean lengths were
longer for iZoom than Zoom (P<0.001), and fiber counts were higher for iZoom
than Zoom (P<0.005).
2. Three image quality criteria (Ghosting/Motion artifacts, quality of fat suppression, and overall image scores were rated significantly higher in iZoom than in Zoom
for both readers (P <0.005 for reader1 and P = 0.005 for reader 2) (Fig. 3).
The Distortion Artifact evaluation showed no significant difference between
both sequences for both readers (reader 1, P = 0.052; reader 2, P = 0.681).Discussion
The brachial nerve
is susceptible to failure for fat suppression due to air distortion and motion
artifacts from respiratory motion because it passes near the pulmonary apex. Zoom
is applying interleaved scan packages, therefore prone to image misalignment
due to motion. However, iZoom is not needed the separation of scan packages. iZoom
had better quality of fat suppression than Zoom (Fig4). Therefore, the image
misalignment and failure fat suppression may have influenced tractography
(Fig5). iZoom can be an effective way to
provide DTI with improved image quality, and a potential application for
research and clinical diagnosis in diffusion imaging for the brachial plexus. Conclusion
In the brachial
plexus, iZoom was feasible as a scan method of DTI in the peripheral plexus.
Acknowledgements
No acknowledgement found.References
1.
Ryckie G. , Alexander
W, Irvin T et al. Diffusion tensor imaging of the roots of the brachial
plexus:a systematic review and meta‑analysis of normative values. Clin Transl
Imaging. 2020;8(6):419-431
2.
Wilm BJ, Svensson J, Henning A,
Pruessmann KP, Boesiger P, Kollias SS. Reduced field-of-view MRI using outer
volume suppression for spinal cord diffusion imaging. Magn Reson Med. 2007
Mar;57(3):625-30. doi: 10.1002/mrm.21167.
3.
Jeong, H., et al. High
resolution human diffusion tensor imaging using 2D navigated multi-shot SENSE
EPI at 7 Tesla. Magn Reson Med. 2013, 69 (3): 793-802.
4.
Wu ZG, Zhang J, Fang WX, Huang
F, B1 insensitive zoomed FOV imaging, ISMRM., 2015; 0953.
5.
Banerjee S, Nishimura DG,
Shankaranarayanan A, Saritas EU. Reduced field-of-view DWI with robust fat
suppression and unrestricted slice coverage using tilted 2D RF excitation. Magn
Reson Med. 2016 Dec;76(6):1668-1676. doi: 10.1002/mrm.26405.
6.
Zhigang Wu et al. iZoom with 2
order flow compensated diffusion for Improving cardiac diffusion imaging: a
preliminary study, ISMRM,2022;4137