Jian Wu1, Taishan Kang2, Simin Li1, Weikun Chen1, Zhigang Wu3, Congbo Cai1, and Shuhui Cai1
1Department of Electronic Science, Xiamen University, Xiamen, China, 2Department of Radiology, Zhongshan Hospital of Xiamen University, School of Medicine, Xiamen University, Xiamen, China, 3Clinical & Technical Solutions, Philips Healthcare, Shenzhen, China
Synopsis
Keywords: Quantitative Imaging, Kidney
The overlapping-echo detachment (OLED) imaging can capture
a T
2 map within about 150 ms. However, when the region of interest
(ROI) is smaller than the object, it is inefficient to acquire the whole object
because many unnecessary phase-encoding steps should be acquired. The zonal
oblique multislice (ZOOM) is a method to reduce the field of view (FOV) by
using two pulses to excite two slabs that have a specific angle between them. This study combined ZOOM with OLED
to capture the T
2 map with reduced FOV. The ZOOM-OLED can reduce
geometric distortion and improve image resolution.
Introduction
The overlapping-echo detachment (OLED) imaging is a
fast quantitative magnetic resonance imaging (MRI) method which can quantify multiple MR parameres within a single shot.
As the EPI readout is used in data acquisition, OLED is sensitive to the B0
inhomogeneity, and the image resolution is limited. For the EPI-based sequence,
reducing the field of view (FOV) in the phase-encoding direction can reduce the
image distortion when the region of interest (ROI) is much smaller than the
object. Zonal oblique multislice (ZOOM) is a technique that can reduce the FOV
by rotating the excitation slab by a specific angle. Here, we combined OLED and
ZOOM to obtain the T2 map with high image resolution and less
geometric distortion. The ZOOM-OLED can obtain the T2 map in a
single shot, which allows us to capture the real-time T2 values of the
kidney without breath-hold.Methods
Pulse sequence
design:
The OLED sequence can be easily combined with the
ZOOM technique. The ZOOM-OLED sequence is shown in Figure 1c. Two excitation RF pulses are applied to generate
two magnetization
pathways (Figure 1b) that can
be considered
replicas of the k-space signal with different T2 weightings. The
excitation and refocusing slabs have a specific angle (Figure 1a). Before the two excitation pulses, regional saturation (RS) pulses are used to suppress
the signals outside the ROI.
Training
samples generation: The open-source MRI simulation software
(MRiLab) was used to generate the training samples. Details about training
samples generation can be found
in literature [1].
Reconstruction: The flowchart of T2 map reconstruction
is shown in Figure 2. U-Net was used to reconstruct the T2
map from the acquired OLED image. At the training stage, we took the simulated
OLED image as input, and the corresponding T2 map was taken as the
label. At the testing stage, the acquired
DICOM data were fed into the pre-trained
U-Net to obtain the T2 map.
Experiments: This study was approved by the IRB at Zhongshan
Hospital of Xiamen University. One volunteer (male, age = 27) was recruited and
imaged on a 3.0 T MR scanner (Ingenia CX, Philips Healthcare). The experiments were
performed with the ZOOM-OLED sequence (TE/TR, 41 ms/2000 ms; FOV, 250 × 120 mm2;
matrix, 125 × 60). The flip angle of the excitation pulses was α = 55°.Results
The OLED image acquired with the original OLED
technique was compared with that acquired with ZOOM-OLED (Figure 3), and the volunteer was scanned under free
breathing. It shows that the stripe is finer in ZOOM-OLED image, resuting
higher spatial resolution of the T2 map. The ZOOM-OLED scans
were also performed multiple times with TR = 2000 ms. During the scan, the volunteer was asked to keep his breath. The
reconstructed real-time T2 maps are shown in Figure 4. We can see that due to its ultrafast characteristic, ZOOM-OLED
provides T2 maps without motion artifacts. The regions of interest (ROI) are located almost at the same place in
different T2 maps. The boxplot of the T2 value for
the ROI shown in Figure 4 (TR = 1) is given in Figure 5. The T2 variation could probably be attributed to the dynamic
physiological process of the blood flow and
the volunteer’s tiny movement.Conclusion
In this work, we
proposed a fast quantitative MRI method, ZOOM-OLED, to obtain less-distortion and high-spatial-resolution
T2 maps.Acknowledgements
This
work was supported by the National Natural Science Foundation of China under
grant number 11775184.References
[1]
Ma LC, Wu J, et al. Single-shot multi-parametric mapping based on multiple
overlapping-echo detachment (MOLED) imaging. Neuroimage, 2022, 263:119645.