Abhijit J Chaudhari1, Robert D Boutin2, Yongwan Lim3, Sophia Cui4, Robert M Szabo5, and Krishna S Nayak3
1Radiology, University of California Davis, Sacramento, CA, United States, 2Radiology, Stanford University, Palo Alto, CA, United States, 3University of Southern California, Los Angeles, CA, United States, 4Siemens Medical Solutions, Malvern, PA, United States, 5Orthopaedic Surgery, University of California Davis, Sacramento, CA, United States
Synopsis
Keywords: Joints, Skeletal, Real-time imaging; Wrist Kinematics; Dynamic imaging.
We
acquired real-time MRI of the actively moving wrist in humans utilizing a
high-performance 0.55T system. Resulting images and associated static wrist
scans were assessed by two expert readers blinded to acquisition parameters. Our
results show that images acquired at a high temporal resolution of 12.6 ms per
frame demonstrate minimal image degradation compared to images acquired at
temporal resolution of 100 ms or higher, and enable improved characterization
of wrist motion. These benefits support further studies to assess high-performance
0.55T systems for the evaluation of dynamic dysfunction of the wrist.
Introduction
MRI is commonly employed in the
evaluation of wrist pathology, with assessment relying primarily on static
scans [1, 2]. In previous work we demonstrated the
utility of 3T MRI for imaging the actively moving wrist to overcome
shortcomings of the static scans for assessing wrist kinematics [3, 4]. However, image artifacts arising from
the relatively slow acquisition rate and magnetic field susceptibility frequently
obscured critical wrist structures [3, 5]. To address these limitations, we
conducted a human study utilizing a high-performance 0.55T system for obtaining
real-time (RT) scans of the actively moving wrist and static scans of the wrist
in the neutral position. The resulting images were evaluated by two expert readers
independently, providing qualitative and quantitative assessments of image
quality and specific imaging features of interest for assessing wrist
dysfunction.Methods
Scan
were performed using a whole body 0.55T system (prototype MAGNETOM Aera,
Siemens Healthineers, Erlangen, Germany) equipped with high-performance
shielded gradients (45 mT/m amplitude, 200 T/m/s slew rate). The study cohort
consisted of 5 participants (3 men and 2 women, median age (range) = 46 (31-59)
years). The dominant wrist for each participant was scanned. Participants lay
in the “superman position” with one arm placed on spine RF coil (below) and
6-channel surface coil (above), in a configuration that did not restrict wrist
range of motion. Using the localizer acquisition, a coronal slice showing the
scapholunate (SL) interval was prescribed. Each participant
performed two wrist maneuvers (radial/ulnar deviation and the clenched fist maneuver)
during RT-MRI acquisition utilizing their full, active range-of-motion at a
natural speed. These maneuvers are of significance in the evaluation of dynamic
wrist dysfunction.
A 2D spiral balanced steady-state
free-precession (bSSFP) sequence was implemented within the RTHawk realātime
imaging platform (HeartVista, Inc., Los Altos, CA, USA). Imaging parameters
were: in-plane resolution = 1.38x1.38 mm2, slice thickness = 8 mm,
field-of-view = 18x18 cm2, TR / TE= 12.70/0.78 ms, readout duration
= 8.38 ms, sampling interval = 2 µs, flip angle = 70º. Thirteen spiral-out sampling
scheme and temporal bit-reversed orders of spiral interleaves was used.
Localization, imaging, and reconstructions were performed using the real-time
console while MRI data were analyzed from off-line reconstructed images. The temporal
finite-difference constrained reconstruction implemented using the Berkeley
Advanced Reconstruction Toolbox was used. For static
imaging of the immobilized wrist in the neutral position, the acquisition
involved a T1-weighted 3D pulse sequence with voxel size: 0.61×0.61×0.60
mm3, and 2D proton-density weighted (coronal) and T1-weighted
(coronal and axial) turbo spin echo acquisitions with in-plane resolutions of
0.31×0.31 and 0.25×0.25 mm2, and slice thicknesses of 2.50 and 3.00
mm, respectively.
The RT images were assessed by two experienced
readers, a fellowship-trained musculoskeletal radiologist with 25 years of
post-training experience, and a PhD scientist with expertise in musculoskeletal
kinematics and wrist imaging for 15 years, independently. First, the readers scored the images based on
a 5-point Likert scale (1: ‘extremely poor’; 2: ‘poor’; 3: ‘average’; 4:
‘good’; and 5: ‘excellent’) [6]. Second, two imaging
features, the SL and the capitotriquetral (CT) interval, were evaluated from
the images. The scoring was based on a 3-point scale: (1: not visualized, 2:
visualized but not well-defined, and 3: well-defined [7]). The static scans
were assessed based on the same 5-point Likert and 3-point scales. Further,
visualization of joint distention, sharpness of joint borders, and the ability
to assess triangular fibrocartilage complex (TFCC) tears was evaluated. Each of
these factors was graded poor, fair, good or very good based
on criteria described by Andreisek et al [8].Results
From
static images, the wrist bones were visualized clearly in the T1-weighted
acquisition (Figure 1). The proton density-weighted acquisition demonstrated
TFCC anatomy and the SL ligament (Figure 1). All static scans were rated good
to very good and diagnostic. For the radial-ulnar deviation maneuver, reader
scores of RT-MRI scans with temporal resolution of less than 100 ms were higher
than those with temporal resolutions greater than 100 ms (p<0.05), Table 1.
For the clenched fist maneuver, however, there were no differences in reader
scores with changes in temporal resolution. With finer temporal resolution, images
showed reduced motion blurring and a sharper boundary between the scaphoid and
lunate or capitate and triquetrum, indicating an improved ability to quantify
changes to the SL and CT interval over the range of motion (Figures 2, 3).
Furthermore, visualization of the changes in the SL and CT interval with
temporal resolution down to 12.6 ms was possible with minimal degradation in
image quality during the wrist maneuvers.Discussion and Conclusions
Our
results demonstrate the feasibility of a high-performance 0.55T MRI system in
assessing tissues of the moving wrist during the performance of a wrist
maneuvers relevant to wrist dysfunction. Images down to a temporal resolution
of 12.6 ms showed minimal degradation in image quality compared to images with
temporal resolutions of greater than 100 ms and allowed improved visualization
of joint motion. These benefits combined with other advantages of low-field
imaging make it a promising technology to evaluate further for assessing wrist
biomechanics and dynamic dysfunction. Future studies will include more subjects
with and without wrist pathology and evaluation of 0.55T-specific acquisition
strategies for the moving wrist.Acknowledgements
We
acknowledge grant support from the National Science Foundation (#1828736) and
research support from Siemens Healthineers.References
[1] R.
L. Leake, M. K. Mills, H. Allen, A. M. Crawford, P. H. Kobes, and M.
Soltanolkotabi, "MRI of the Wrist Ligaments," Topics in Magnetic Resonance Imaging, vol. 29, no. 5, pp. 209-220,
2020.
[2] C. A.
Tiegs-Heiden and B. M. Howe, "Imaging of the Hand and Wrist," Clinics in Sports Medicine, vol. 39, no.
2, pp. 223-245, 2020.
[3] R. D.
Boutin et al., "Real-time
magnetic resonance imaging during active wrist motion--initial
observations," PLoS One, vol. 8,
no. 12, p. e84004, 2013.
[4] S. S.
Henrichon et al., "Dynamic MRI
of the wrist in less than 20 seconds: normal midcarpal motion and reader
reliability," Skeletal Radiol, vol.
49, no. 2, pp. 241-248, Feb 2020.
[5] C. B.
Shaw et al., "Real-time three-dimensional
MRI for the assessment of dynamic carpal instability," PLoS One, vol. 14, no. 9, p. e0222704,
2019.
[6] S.
Partovi et al., "Image quality
assessment of automatic three-segment MR attenuation correction vs. CT
attenuation correction," (in eng), Am
J Nucl Med Mol Imaging, vol. 3, no. 3, pp. 291-9, 2013.
[7] S. D.
Sharma, C. L. Fong, B. S. Tzung, M. Law, and K. S. Nayak, "Clinical image
quality assessment of accelerated magnetic resonance neuroimaging using
compressed sensing," Investigative Radiology,
vol. 48, no. 9, pp. 638-645, 2013.
[8] G.
Andreisek, S. R. Duc, J. M. Froehlich, J. Hodler, and D. Weishaupt, "MR
arthrography of the shoulder, hip, and wrist: evaluation of contrast dynamics
and image quality with increasing injection-to-imaging time," AJR Am J Roentgenol, vol. 188, no. 4,
pp. 1081-8, Apr 2007.