Chloé Najac1, Florian Birk2,3, Tom O’Reilly1, Klaus Scheffler2,3, Andrew Webb1, and Rahel Heule2,3,4
1C.J. Gorter MRI Center, Department of Radiology, Leiden University Medical Center, Leiden, Netherlands, 2Department of High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Tübingen, Germany, 3Department of Biomedical Magnetic Resonance, University of Tübingen, Tübingen, Germany, 4Center for MR Research, University Children's Hospital, Zurich, Switzerland
Synopsis
Keywords: Low-Field MRI, Data Acquisition
Point-of-care imaging with low-field MRI (<0.1T) is a potential game
changer for low-income countries and the intensive care unit. The main
challenge is the low SNR. Balanced steady-state free precession (bSSFP)
sequences are fast and SNR-efficient. However, bSSFP is very sensitive to B
0
inhomogeneities resulting in banding artifacts. We evaluated the
feasibility of using bSSFP on our 46 mT MRI scanner. By acquiring 17 bSSFP
datasets with linearly increasing frequency offsets (from 0 to 1/TR), we could
reconstruct banding-free maximum-intensity images as well as F
0 and
F
-1 SSFP configurations, which we employed for rapid relaxation time
mapping.
Introduction
Development of low-field
MRIs (with B0<0.1T) for point-of-care (POC) applications has
recently become increasingly widespread1. The reduced cost and
improved portability drastically increase their accessibility compared to
conventional systems. However, low-field brain images suffer from low SNR and poor
tissue contrast2. The balanced steady-state free precession (bSSFP) imaging
sequence is fast, offers high SNR, and produces a T2/T1-weighted
image contrast3. However, bSSFP is very sensitive to B0 inhomogeneities
(ΔB0), which lead to banding artifacts3. Banding-free
images can be achieved by acquiring multiple images using RF phase-cycling
increments or frequency offsets3. Although optimized passive and
active shimming schemes have been developed to minimize ΔB04,
compact low-field permanent MRI systems based on cylindrical Halbach arrays
have intrinsically large ΔB0 (~thousands ppm) due to the finite
diameter-to-length ratio5 compared to larger low-field fixed
location systems on which bSSFP has been implemented6. Here, we
wanted to evaluate the potential of bSSFP on our 46mT imaging system. Materials and Methods
Hardware: We used a 46mT Halbach
array-based magnet (outer/inner diameter=60/30.1cm, length=49.2cm, weight=111kg),
with custom-built RF and Bruker gradient amplifiers and a Magritek Kea2
spectrometer7. Imaging was performed using a solenoid and an
elliptical spiral-solenoid head coil2,7.
bSSFP implementation: The 3D bSSFP sequence
consisted of a train of hard pulses with alternating flip angle (α), following
an α/2 ramp preparation (Fig.1A). A zig-zag phase-encode trajectory was
used to smoothly sample k-space8 (Fig.1B). To correct for banding
artifacts we used a range of equidistant frequency offsets between 0
and 1/TR.
Sequence
validation (n=3):
5 tubes were
filled with water-based agarose gel (ranging from 0 to 4%) and 2mM copper-sulfate
to obtain different T1/T2 ratios (Fig.2A). T2
mapping was performed using a 3D Carr-Purcell-Meiboom-Gill (CPMG) imaging
sequence (resolution=3x3x10mm3, TR/echo-spacing=1250/20ms, 10 echoes,
acq. time~15min). T1 mapping was performed using 3D inversion
recovery (IR) scans with TSE readout (identical resolution, TR/TE/TEeff=1250/14/14ms,
ETL=6, inversion times=50/100/150/200/300/400ms, acq. time~15min). A series of
11 bSSFP datasets (identical resolution, TR/TE=16/8ms, acq. time~11x3.4min) were
acquired with a flip angle varying from 10º to 170º. Maximum-intensity bSSFP
images were reconstructed for each flip angle using 17 frequency offset increments.
The signal was averaged over 8 slices and a region-of-interest (5x5 pixels) was
defined to extract proton-density (PD), T2, T1 and bSSFP
values within each sample (Fig.2/3). BSSFP signal evolution as a
function of flip angle was fitted with the theoretical expression3: $$$\frac{M_{SS}}{M_{0}}=\frac{sin(\alpha)}{1+cos(\alpha)+(1-cos(\alpha))*\frac{T_{1}}{T_{2}}}$$$. Optimal flip angle and steady-state
level obtained from the fit were compared to values derived from T1/T2
mapping performed with CPMG/IR sequences using the formula3: $$$FA_{optim}=arcos(\frac{\frac{T_{1}}{T_{2}}-1}{\frac{T_{1}}{T_{2}}+1})$$$ and $$$M_{SS}=\frac{1}{2}*M_{0}*\sqrt{\frac{T_{2}}{T_{2}}}$$$.
Brain-like (BrainLo)
phantom experiment:
BrainLo was built using a 3D printer and filled with different solutions of
agarose/copper-sulphate doped-water/deuterated-water to imitate brain tissue
relaxation properties (Fig.3A). T2/T1 mapping was
performed as described above with (resolution: 2x2x10 mm3, total acq
time~10/8 mins for CPMG/IR sequences). bSSFP data (identical resolution, total
acq. time~2 mins) were acquired with α=90º. First, the sum of squares, complex
sum, magnitude sum, and maximum-intensity images were calculated, neglecting
any additional frequency drift during multi-offset series’ acquisition (Fig.4).
Secondly, F0 and F-1 modes were retrieved via a Fourier
transform of the complex bSSFP frequency response, incorporating the measured
frequency drift. Finally, a T2 map was calculated using the F0
and F-1 configurations in an approach similar to MIRACLE relaxometry9-12.Results and Discussion
Sequence
validation:
As
illustrated in Fig.2A, our samples had T1/T2
ratios ranging from ~1:1 to ~4.7:1. Banding-free maximum-intensity image could
be reconstructed (Fig.2B). Fig.3A shows the evolution of the
signal as a function of the flip angle. The sample with the lowest and highest
T1/T2 ratio shows the highest and lowest signal respectively,
which is in agreement with literature3. The optimal flip angle,
steady-state signal level and T1/T2 ratio were close to
expected values (Fig.3B). Note however, that the fit slightly deviated
from data for low flip angle since in this regime, the maximum might not occur
in the passband but at different frequency offsets. The steady-state signal at
the highest T1/T2 ratio was slightly higher than predicted
from the model.
Brain-like (BrainLo)
phantom:
As shown in Fig.4A, GM/WM/CSF compartments had T2 and T1
values measured using conventional techniques which match in vivo values
at low-field2. Slight residual banding was observed in each
reconstruction, except for the maximum-intensity image (Fig.4B). The
cause is probably small drifts in B0 during acquisition. By
calculating the difference between each dataset and the first dataset (Fig.5A),
we estimated this additional drift, corrected it and produced banding-free F0
and F-1 modes (Fig.5B), which were used to reconstruct the T2
map. The relative T2 values are in the right ratio, but absolute
values in WM/GM are ~40% lower than those obtained using the much slower
conventional techniques. In future experiments we will investigate whether this
is a systematic error, or simply reflects the fact that different acquisition
schemes result in different apparent relaxation times. Conclusion
POC imaging with
low-field MRI (B0<0.1T) is a potential game changer for low-income
countries and patients in the intensive care unit. However, POC systems have
intrinsically low SNR and relatively high ΔB0. We illustrated the feasibility
to reconstruct banding-free bSSFP images at 46mT on a portable POC system and
to rapidly quantify T2 based on the retrieved F0 and F-1
configurations. Acknowledgements
This project has received funding from Horizon 2020 ERC
Advanced PASMAR 101021218, ERC advanced grant SpreadMRI 834940 and the Dutch Science Foundation Open Technology 18981.References
1Saracanie, Front. Phys. (2020) ; 2O’Reilly
et al., MRM (2021) ; 3Scheffler et al., Eur Radiol (2003) ;
4O’Reilly et al., ISMRM #63 (2022); 5Koolstra et al.,
MAGMA (2021); 6Saracanie et al., Sci. Rep. (2015) ; 7O’Reilly
et al., MRM (2020); 8Scheffler et al., ISMRM #294 (2003) ; 9Heule
et al., MRM (2013) ; 10Welsch et al., MRM (2009) ; 11Heule
et al., MRM (2014) ; 12Nguyen et al., MRM (2017)