Jinye Li1, Lixin Sun2, Na Hu2, Mengxiao Liu3, Linsheng Wang2, and Chuanting Li2
1shangdong provincial ENT hospital, Jinan, China, 2Shandong Provincial ENT Hospital, Jinan, China, 3Siemens healthineers, Shanghai, China
Synopsis
Keywords: Head & Neck/ENT, Head & Neck/ENT
In order to visualize endolymphatic
hydrops in Ménière’s
disease better, we compared the conventional 3D real IR and ZOOMit 3D
real IR sequence. Results suggest that visualization
of the endolymphatic space might
be higher by zs-3D real IR compared with t-3D real IR, especially in the
cochlea.
BACKGROUND AND PURPOSE
As
a chronic disease, the incidence of Ménière’s disease (MD) reaches 200-500
cases per 1000,000 people1 and seriously affect the quality of
patients’ life. What’s more, endolymphatic hydrops (EH) is the main pathological manifestation of
MD 2. Magnetic resonance contrast provides a
completely new examination method for MD diagnosis, however the
detection rate of premortem MRI EH is lower than that of postmortem autopsy EH,
indicating that current MR technique may underestimate the Ménière’s disease. So, we
prospectively investigated whether a novel high-resolution MR imaging
technique, the 3D-ZOOMit real inversion recovery, improves endolymphatic
hydrops detection compared with conventional 3D-TSE inversion recovery.BACKGROUND AND PURPOSE
Fifty patients (22 male, 28 female, average age ± SD: 53.06 ±
13.28 years, affected side:21right, 29 left) with definite unilateral Ménière’s
disease were enrolled from December 2020 to October 2021. All patients were
scanned on a 3T MR system (MAGNETOM Prisma, Siemens Healthcare, Erlangen,
Germany) with 64 channel head coil. Conventional 3D real IR and 3D-ZOOMit real IR
were performed after 6 h gadolinium injection with a double dose(0.4 0.4 ml/kg
body weight; ProHance). The parameters of conventional
3D real IR is as below: TR 5300ms, TE 191ms, TI 1850, FOV 220mm×220mm, voxel size 0.6mm×0.6mm and scan time 16min47s; The parameters
of ZOOMit 3D real IR is as below: TR
8000ms, TE 491ms, TI 2250, FOV 160mm×80mm, voxel size 0.6mm×0.6mm and scan time 15min12s. The endo- and perilymph spaces
were scored separately by 2 radiologists independently. Scores for the separate
visualization of the cochlear and vestibular endolymph were as follows: 1, impossible
to recognize; 2, some can be recognized; 3, most can be recognized; and 4, all
can be recognized 26.
The contrast-to-noise ratio, SNR, and
signal intensity ratio of the two sequences were respectively calculated using
the following equations:
CNR = (SIperi − SIendo)/SDnois [Eq. 1]
SNR
= SIperi/SDnoise
[Eq. 2]
SIR
= SIperi/SIlmcp
[Eq. 3]
Where, SIperi, SIendo and
SIlmcp present the signal intensity of the cochlear basal
turn’s center, the endolymph and the left middle cerebellar peduncle of the
images. SDnoise is the standard deviation in the artifact-free air
area. And the presence of endolymphatic
hydrops was evaluated in patients.
For quantitative and qualitative analysis, we used the paired
samples t-test or Wilcoxon signed ranked test. Statistical significance was set
at P values < 0.05(SPSS25.0, IBM, Chicago, IL, USA).RESULTS
The
visualization of 3D ZOOMit real IR
in endolymphatic space of cochlea
and vestibule is better than conventional
3D-TSE real
IR (P < .001). For
the visualization of endolymphatic space in the cochlea, all images from ZOOMit
3D real IR performed well (score ≥ 3) and 37 patients scored 4. For 3D TSE real
IR, there are 5 patients’ score ≤2, and only 17 patients scored 4. Meanwhile, for
the recognition of the endolymphatic space in the vestibule, most images can
perform well in both sequences (49 patients’ score ≥ 3), but the number of
score4 in ZOOMit 3D real IR is 40 and the number is 27 in 3D TSE real IR. For
asymptomatic side, the performance of ZOOMit 3D real IR is also better than 3D
TSE real IR (table 1).
The CNR, SNR, and SIR values
of 3D-ZOOMit real IR images were statistically higher than those of conventional 3D-TSE inversion recovery
images (all P < .001), the details
are shown in table 2. DISCUSSION
In our study, CNR, SNR, and SIR values,
endo- and perilymph image quality scores by ZOOMit 3D real IR were higher than
those by 3D TSE real IR. For ZOOMit is a technique with an independent parallel
transmission (pTX) system, thus can provide less image blur and artifacts,
higher spatial resolution, and faster screening time. CONCLUSIONS
3D-ZOOMit real IR sequence is superior
to conventional 3D-TSE IR sequence in visualizing the
endolymphatic space, detecting endolymphatic hydrops, and discovering contrast
permeability.Acknowledgements
NoneReferences
1. Havia
M, Kentala E, Pyykko I. Prevalence of
Meniere's disease in general population of Southern Finland. Otolaryngol Head Neck Surg
2005;133:762-768.https://doi.org/10.1016/j.otohns.2005.06.015
2. Lopez-Escamez
JA, Carey J, Chung WH, et al. Diagnostic
criteria for Meniere's disease. J
Vestib Res 2015;25:1-7.https://doi.org/10.3233/VES-150549