Afis Ajala1, Seung-Kyun Lee1, Nastaren Abad1, Yihe Hua1, and Thomas Foo1
1GE Global Research, Niskayuna, NY, United States
Synopsis
Keywords: Gradients, Gradients, Concomitant Field Correction
The use of higher-performance
gradient coils results in stronger second-order concomitant magnetic fields,
which can lead to image artifacts such as signal dropout, blurring, and phase
errors that cannot be corrected by pre-emphasis of gradient waveforms and/or radio
frequency modulation alone. We have developed an axially symmetric second-order
field coil that is insertable, and demonstrate its ability to prospectively correct
the additional phase generated by second-order concomitant fields in 2D phase
contrast and spiral-out gradient echo imaging in a 3.0 T high-performance
head-gradient (MAGNUS) system.
Introduction
High-performance gradient systems combine
high gradient amplitudes (Gmax) and high maximum slew rates (SRmax)
to enable rapid and high-resolution magnetic resonance imaging (MRI) for various applications1–3. High-quality diffusion, perfusion
and functional MRI have been demonstrated in the MAGNUS (GE Research,
Niskayuna, NY, USA) asymmetric gradient system that simultaneously delivers a Gmax
and SRmax of 200 mT/m and 500 T/m/s, respectively, using a 620 A/1400
V gradient driver in a clinical whole‐body 3.0 T magnet1,4–7. However, utilization of higher Gmax
at a given main magnetic field strength is accompanied by stronger second-order
concomitant gradient (SOCG) fields that cannot be corrected by pre-emphasis of
gradient waveforms and/or radio frequency (RF) modulation alone8,9. Such fields can lead to signal
dropout and phase errors in phase contrast (PC) MRI, as well as in-plane and through-plane
blurring in spiral imaging. Most published methods for the correction of SOCG rely
on sequence modification and image processing10,11. Here, we report the use of an
axially symmetric coil for prospective correction of SOCG in PC and 2D
spiral-out gradient echo imaging at 3.0 T.Materials and Methods
Second Order Correction Coil Design
and Setup: An axi-symmetric
second order $$$($$$z2$$$-($$$x2$$$+$$$y2$$$)/$$$2$$$)$$$ magnetic field
(B-field) coil was designed by nulling the zeroth and fourth order Taylor’s
expansion terms of the axial B-field created by concentric circular loop coils arranged
on the surface of a 32 cm-diameter cylindrical former in a mirror-symmetric
manner. The diameter of the copper wire used for the coil construction was 1.67
mm, and other design parameters are detailed in Figure 1A. The correction coil
was inserted inside the head transmit/receive coil of the MAGNUS system, and
was driven by an audio amplifier (AE Techron 7220$$$–$$$Ekhart, IN, USA), which
received a unipolar pulsed voltage signal from a signal generator (Tektronix
AFG 3252$$$–$$$Beaverton, OR, USA) triggered by the scanner (Figure 1B and 1C).
Simulation and Phantom Experiments: The B-field inside the designed correction
coil was calculated using the Biot-Savart Law on a 3D grid of 1$$$\times$$$1$$$\times$$$1 mm3 spatial resolution. A pickup coil was used to experimentally
sample the B-field along the axis of the coil while applying an alternating
current of 0.5 A. Imaging experiments were carried out at 3.0 T using the MAGNUS
system with firmware patches for zeroth and first order eddy current and
concomitant gradient fields correction12,13. The correction coil was used to null
phase errors (non-zero zeroth moment) due to SOCG in single-sided PC MRI with bipolar
gradients in the slice selection direction of a coronal and axial acquisition.
For the coronal acquisition (Figure 2A), the correction coil current, Gmax,
SRmax and plateau time of each bipolar gradient lobe were 0.23 A, 59 mT/m, 400 T/m/s and 1 ms respectively. The axial PC experiment was identical to
the coronal acquisition except that the correction current and Gmax of
the lobes were increased to 0.50 A and 150 mT/m respectively (Figure 2B). A
silicone gel phantom was used for both PC experiments, and other acquisition
parameters include slices$$$=$$$3, TE/TR$$$=$$$6.6/105 ms and field of view$$$=$$$20 cm. An
acquisition with zero amplitude bipolar gradients was used to remove background
phase from the PC acquisitions. Additionally, the correction coil was used to reduce
blurring artifacts due to SOCG generated during readout in a 2D multi-slice
gradient echo (GRE) MRI with spiral-out readout. A 10
cm-diameter American College of Radiology phantom was used for the 2D-GRE experiment
and the scan parameters are summarized in Figure 2C. SOCG was calculated as
previously described:10
$$
\label{eq:decoder}
\mathrm{SOCG} \simeq \frac{\mathrm{G_{z}}^{2}}{8\mathrm{B}_{0}}\mathrm{x}^{2}+\frac{\mathrm{G_{z}}^{2}}{8\mathrm{B}_{0}}\mathrm{y}^{2}+\frac{\mathrm{G_{x}^{2}+G_{y}^{2}}}{2\mathrm{B}_{0}}\mathrm{z}^{2}-\frac{\mathrm{G_{x}G_{z}}}{2\mathrm{B}_{0}}\mathrm{xz}-\frac{\mathrm{G_{y}G_{z}}}{2\mathrm{B}_{0}}\mathrm{yz}
\tag{1}
$$
The current amplitude in the correction
coil was empirically tuned to best cancel the SOCG effect in each experiment.Results and Discussions
Figure 3A and 3B show the quadratic dependence of the simulated correction
coil B-field on z location. The measured B-field along the axis of the coil
showed good agreement with simulation (Figure 3C). The correction coil B-field reduced
the SOCG phase accrued in the coronal and axial PC acquisition by 100% and 83% respectively.
This reduction is indicated in the fit to the experimental data by the
coefficients of z2 and x2 for the coronal and axial
acquisitions respectively (Figure 4). The remaining linear and constant
components of the phase can be eliminated by frequency shifting or retrospectively
during image reconstruction. The in-plane blurring in 2D GRE spiral-out imaging
due to SOCG phase accrual was substantially reduced by the correction coil (Figure
5). As expected, the amount of blurring is highest in slices that are farthest
from the isocenter of the gradient (Figure 5A). The width of the line profile
on the images at z$$$\rm{=}$$$40 mm demonstrates 42.7% reduction in blurring and
substantial increase in contrast when the correction coil was turned on (Figure
5C). The z2-dependent B-field generated by the correction coil also
has x2 and y2 dependence (to satisfy Laplace equation)
that can cause other phase errors in 3D imaging, which suggests the need for an
additional (x2$$$-$$$y2) coil and retrospective correction for a more complete
SOCG compensation on all three axes.Conclusion
A
custom-built SOCG correction coil has been used to prospectively correct phase
accruals due to SOCG on single-sided 2D PC and spiral-out GRE MRI in the MAGNUS high
performance gradient system.Acknowledgements
This work
was supported by CDMRP W81XWH-16-2-0054, and this presentation does not necessarily represent the official views of the
funding agency. We also acknowledge the help of Eric
Fiveland for helping with 3D printing of parts needed for coil construction.References
1. Foo,
T. K. F. et al. Highly efficient head-only magnetic field insert
gradient coil for achieving simultaneous high gradient amplitude and slew rate
at 3.0T (MAGNUS) for brain microstructure imaging. Magnetic Resonance in
Medicine 83, 2356–2369 (2020).
2. Setsompop, K. et al. Pushing the
limits of in vivo diffusion MRI for the Human Connectome Project. Neuroimage
80, 220–233 (2013).
3. Weiger, M. et al. A
high-performance gradient insert for rapid and short-T2 imaging at full duty
cycle. Magn Reson Med 79, 3256–3266 (2018).
4. Abad, N. et al. Brain
Microstructure Imaging with Ultrahigh B-Encoding using MAGNUS High Performance
Gradients. in ISMRM 2022 Annual Proceedings (2022).
5. Shih, R. et al. Initial Clinical
Experience with MAGNUS Ultra-High-Performance Gradient Coil for Diffusion
Microstructure Imaging of Intracranial Pathology. in ISMRM 2022 Annual
Proceedings (2022).
6. Zhu, A. et al. Characterizing
Restricted Diffusion in Pre-/Post- treatment Gliomas Using Time-dependent
Diffusion MRI at Ultra-high-gradient Human 3.0T. in ISMRM 2022 Annual
Proceedings (2022).
7. Ajala, A. et al. 3D
Pseudo-Continuous Arterial Spin Labeling Acquisition using a High-Performance
Gradient System: A Scan Time and Image Quality Assessment. in ISMRM 2022
Annual Proceedings (2022).
8. Tao, S. et al. Gradient
Pre-Emphasis to Counteract First-Order Concomitant Fields on Asymmetric MRI
Gradient Systems. Magn Reson Med 77, 2250–2262 (2017).
9. Weavers, P. T. et al. B0
concomitant field compensation for MRI systems employing asymmetric transverse
gradient coils. Magn Reson Med 79, 1538–1544 (2018).
10. Bernstein, M. A. et al.
Concomitant gradient terms in phase contrast MR: analysis and correction. Magn
Reson Med 39, 300–308 (1998).
11. King, K. F., Ganin, A., Zhou, X. J. &
Bernstein, M. A. Concomitant gradient field effects in spiral scans. Magnetic
Resonance in Medicine 41, 103–112 (1999).
12. Tao, S. et al. The Effect of
Concomitant Fields in Fast Spin Echo Acquisition on Asymmetric MRI Gradient
Systems. Magn Reson Med 79, 1354–1364 (2018).
13. TAO, S., TRZASKO, J. D., Shu, Y.,
WEAVERS, P. T. & BERNSTEIN, M. A. Systems and methods for concomitant field
correction in magnetic resonance imaging with asymmetric gradients. (2016).