Verónica Aramendía-Vidaurreta1,2, Sergio M. Solís-Barquero1,2, Marta Vidorreta3, Ana Ezponda1,2, Gorka Bastarrika1,2, and María A. Fernández-Seara1,2
1Radiology, Clínica Universidad de Navarra, Pamplona, Spain, 2IdiSNA, Instituto de Investigación Sanitaria de Navarra, Pamplona, Spain, 3Siemens Healthineers, Madrid, Spain
Synopsis
Keywords: Heart, Perfusion
Myocardial
perfusion can be quantitatively measured noninvasively using arterial spin labeling
(ASL). Motion due to the cardiac cycle is tackled with the use of single- or
double-ECG gating. The goals of this study were to investigate the performance
of double-gating, in synchronized breathing myocardial ASL with presaturation
pulses by comparison with single-gating, and to compare the different
quantification strategies for double-gated data. This study showed that double-gating
is more robust to heart rate variability than single-gating in synchronized
breathing ASL sequences with presaturation pulses, but accurate
saturation-recovery fitting requires the acquisition of several baseline
images.
INTRODUCTION
Myocardial
blood flow (MBF) can be quantitatively measured noninvasively in MRI using
arterial spin labeling (ASL) and most typically a flow-sensitive alternating
inversion recovery (FAIR-ASL) approach has been used. The ASL signal is
computed by subtraction of label from control images acquired with different
time stamps, increasing the likelihood of motion artifacts. Thus, strategies to
mitigate the effects of motion are critical to obtain a reproducible signal. In
cardiac ASL, motion due to the cardiac cycle is tackled with the use of either
single- or double-ECG gating. The former uses one trigger to gate the labeling
pulse, and the latter two independent triggers (one for labeling and another
for readout) to assure they both occur in the same cardiac phase even in the
presence of heart rate variability (HRV) [1]. However,
despite the potential of double-gated myocardial ASL observed under breath-holding
conditions, this technique has not been tested using other breathing
strategies. Moreover, different quantification approaches for double-gated data
have been reported [2], [3]. Therefore,
the goals of this study were to investigate the performance of double-gating, in
synchronized breathing myocardial FAIR-ASL with presaturation pulses by
comparison with single-gating, and to compare the different quantification
strategies for double-gated data.METHODS
Data
acquisition: 4 healthy volunteers (2 female, 31 ± 6 years)
underwent a cardiac MRI scan at 1.5 T (Siemens, AERA). The cardiac MRI protocol
consisted on: localizers to identify a mid-ventricular short axis slice, myocardial
FAIR-ASL sequences with double and single-ECG gating with presaturation and hyperbolic
secant inversion pulses(Figure 1) and T1
mapping using MOLLI. For each FAIR-ASL series, 15 pairs of label/control images
were acquired. For single-gating, inversion time (TI) was fixed at 1s.
Additionally, 4 images at short TI and 6 images with no inversion (M0)
were acquired and used for quantification. In FAIR-ASL sequence, a balanced
steady-state (bSSFP) readout was used with the following parameters: field of
view (FOV) = 300x243mm2, matrix = 128x104, repetition time (TR) = 5s,
pixel size = 2.3x2.3 mm2 and GRAPPA-2. Volunteers were instructed to
synchronize their breathing to the sequence sounds to minimize respiratory
motion. Heart rate was recorded for each sequence.
Data
analysis: All FAIR-ASL images were pairwise
registered [4]. Outliers
were identified based on the final metric obtained after registration and
discarded. Regions of interest covering the left ventricular myocardium were
manually drawn for each sequence.
For
double-gating, a three-parameter (M0, Mz0, T1) saturation
recovery model was used to fit the label
and control image data separately. Fitting was repeated employing different
number of baseline images (6, 1 and 0). MBF was quantified using two different models
(Figure 2). For single-gating, pairwise subtraction between individual control
and label image pairs was performed and MBF was estimated using the equation
derived from Buxton’s general kinetic model (Figure 2). T1 of arterial blood at
1.5T was 1.434s [5].
Temporal
signal-to-noise ratio (tSNR) was computed as the mean
quantitative perfusion signal divided by its temporal standard deviation. For
the double-gating quantification model that used the fitting approach, temporal
data were obtained by subtracting each
control image data minus the fitted label signal obtained at the corresponding
inversion time.
Statistical analysis: MBF and tSNR results were compared between
double and single-gating using Wilcoxon signed-rank test.RESULTS AND DISCUSSION
Figure
3 shows the mean and standard deviation of label and control images separately for
two representative subjects with different levels of heart rate variation. It
can be observed that images obtained with double-gating present a lower degree
of motion.
Figure
4 shows quantitative data obtained per subject together with the median and
interquartile range across subjects. MBF values obtained with double and single-gating
were not different (p=1,p=0.57 for
quantification model 1 and 2, respectively), in agreement with previous work [1]. MBF
values obtained with both quantitative methods show similar results. tSNR was
greater for double-gating than single-gating sequences, although differences
were non-significant (p=0.14 and 0.09 for quantification model 1 and 2,
respectively), likely due to the small sample size. Median TI variabilities of
40.91 (30.24-51.93) ms were observed.
Figure
5 shows three-parameter fitting comparison using different numbers of baseline
images. Although similar MBF values were obtained, T1 estimates were closer to
myocardial T1 (measured with MOLLI) if more than one M0 image was used. CONCLUSION
Double-gating
is more robust to heart rate variability than single-gating in synchronized
breathing ASL sequences with presaturation pulses, but accurate
saturation-recovery fitting requires the acquisition of several baseline
images.Acknowledgements
Spanish
Ministry of Science and Innovation (grant: PI21/00578)References
[1] H. P. Do, A. J. Yoon, M. W. Fong, F.
Saremi, M. L. Barr, and K. S. Nayak, “Double-gated myocardial ASL perfusion
imaging is robust to heart rate variation,” Magn. Reson. Med., vol. 77,
no. 5, pp. 1975–1980, 2016, doi: 10.1002/mrm.26282.
[2] B.
P. Poncelet et al., “Measurement of human myocardial perfusion by
double-gated flow alternating inversion recovery EPI,” Magn. Reson. Med.,
vol. 41, no. 3, pp. 510–519, 1999, doi: 10.1002/(SICI)1522-2594(199903)41:3<510::AID-MRM13>3.0.CO;2-G.
[3] M.
Henningsson, C.-J. Carlhäll, and J. Kihlberg, “Myocardial arterial spin
labeling in systole and diastole using flow-sensitive alternating inversion
recovery with parallel imaging and compressed sensing,” NMR Biomed., vol.
n/a, no. n/a, p. e4436, 2020, doi: 10.1002/nbm.4436.
[4] V.
Aramendía-Vidaurreta et al., “Reduction of motion effects in myocardial
arterial spin labeling,” Magn. Reson. Imaging, pp. 1261–1275, 2022.
[5] X.
Zhang et al., “In vivo blood T 1 measurements at 1.5
T, 3 T, and 7 T,” Magn. Reson. Med., vol. 70, no. 4, pp. 1082–1086,
2013, doi: 10.1002/mrm.24550.
[6] M.
Henningsson, C. J. Carlhäll, T. Ebbers, and J. Kihlberg, “Non-contrast
myocardial perfusion in rest and exercise stress using systolic flow-sensitive
alternating inversion recovery,” Magn. Reson. Mater. Physics, Biol. Med.,
no. 0123456789, 2021, doi: 10.1007/s10334-021-00992-3.
[7] R.
B. Buxton, L. R. Frank, E. C. Wong, B. Siewert, S. Warach, and R. R. Edelman, “A
general kinetic model for quantitative perfusion imaging with arterial spin
labeling.,” Magn. Reson. Med., vol. 40, no. 3, pp. 383–396, 1998, doi:
10.1002/mrm.1910400308.