Sri Kirthi Kandala1, Kwanjoon Song2, and Sung-Min Sohn1
1Arizona State University, Tempe, AZ, United States, 2Yonsei University, Gangwon-do, Korea, Republic of
Synopsis
Keywords: Non-Array RF Coils, Antennas & Waveguides, New Devices, Liquid metal RF coil
In this work, we present a novel inflatable radio-frequency
receive coil for Magnetic Resonance Imaging at 7T. A small sample was imaged
with and without inflation, and the overall signal-to-noise ratio of the sample
was improved by 12.7% in dB with inflation. This coil design will offer many opportunities
in the field of endorectal imaging, imaging for irregular sample shapes, and
reducing motion artifacts. Inflation also changes the tuning and matching conditions,
which compensate for loading effects by precisely adjusting the air volume inside the cavity.
Introduction
Liquid metal antennas have been actively developed for
stretchable and flexible sensors with discrete or continuous tuning control.1 Flexible and stretchable liquid metal (LM) based radio frequency (RF) coils are
shown to improve the signal-to-noise ratio (SNR) of the image at high field
strengths.2-4 In this work, we introduce an inflatable RF coil with the
ability to physically reconfigure and move closer to sample by pushing air into
a small cavity. This allows the user to place the RF coil close to the sample
structure which improves peripheral SNR and reduces the motion artifact by
restricting movement. The induced magnetic field (B1)
in a loop coil is defined as
B1=(µ0 IR2)/〖2(R2+d2 )〗(3⁄2) by the Biot-Savart Law, where, R is the radius of the coil, I is the
magnitude of current flow and d is the distance between a coil and a subject. As
the B1 field intensity is proportional to
and
, to the distance and
radius of the coil respectively, an inflatable coil can have superior
performance in MR signal induction and reception in comparison with all other
coil types including flexible and conformal RF coils as the coil is moves much
closer to the region-of-interest by inflation.Method
Inflatable RF coil was designed in two steps i) Elastic
silicone polymer (Ecoflex Series, Smooth-On, USA) is used to facilitate a
flexible and stretchable material for inflation and ii) LM loop coil element was
designed to bend and
stretch along with the inflating base structure. A 3D printed mold (30 mm x 30
mm x 2 mm) was designed to make elastic substrate layers. Two layers are
stacked to create an air cavity with an intake tube and a third layer
containing the LM coil is placed on top of the air cavity to move the RF coil
as shown in Fig.1a. An L-matching network with two variable trimmer capacitors is
connected to the coil element for a 50 Ω match. This stack is placed in a 3D print
coil frame to provide one-directional inflation i.e., towards the sample as
shown in Fig.1b. An air intake tube is located between the bottom two layers
which is connected to a syringe. This syringe is connected to a wirelessly
controlled linear actuator to create a push-pull motion to inflate and deflate
the cavity. The coil is placed 3 mm away from the sample and at full inflation
can raise up to 4.5 mm from the base. Elastomers such as Ecoflex provide high
elasticity, therefore enabling 100% shape retention of the LM coil after
inflation. Results and Discussion
A two-port VNA (FieldFox N9923A, Keysight Technologies,
USA) was used to gather the S-parameters of the inflatable coil with a saline
phantom for various inflation ranges. Fig.2a shows the bench test setup along
with front and top views of the inflatable LM coil frame. Fig.2b shows the
impact of inflation on the coil resonance based on the volume of air inside the
coil cavity. The impedance mismatch caused by inflating the coil and increasing
its dimension was compensated with two variable capacitors located on the
L-matching network. All the MR imaging studies were conducted at Barrow
Neurological Institute - Arizona State University (BNI-ASU), Center for
Preclinical Imaging, using a 7T small-animal, 30-cm horizontal-bore magnet and
BioSpec Avance III spectrometer (Bruker, Billerica, MA) with a 116-mm
high-power gradient set (600 mT/m). Fast-Low-Angle-Shot (FLASH) Sequence with a
repetition time (TR) of 350 ms, echo time (TE) of 5.4 ms, and a flip angle (α)
of 20 degrees was used for MR imaging experiments. A field of view (FOV) of 40
x 40 mm and 256 x 256 matrix leading to an in-plane resolution of 156 x 156 μm,
in addition 5 slices were acquired along the sample with a slice thickness of 1
mm. A Bruker Linear Birdcage coil was used as the transmit coil. The peak power
used for the FLASH sequence was 700 W. A rat bed of 72 mm diameter was used to
hold the receive coil housing. Cherry tomato was then placed inside the receive
coil housing and fed into the scanner. Fig.3a
and 3b show the MR images before and after inflation respectively. All the
parameters of the imaging sequence were kept same including the reciever gain
to validate the performace of the inflateble coil. Same slices (5/5) were of
the MR images were selected to calculate the SNR and mean signal intensity as
shown in Fig.3c. 12.7% (from 22.8 dB to 25.9 dB) in SNR and 43.4% (from xxx to
xxx) in signal intensity are observed with inflation. This novel design will
offer opportunities in many imaging fields and further studies are being
carried out to carefully analyze the benefits and drawback of the coil in
multiple scenarios.Conclusion
In this work, we have designed, fabricated, and tested a
novel inflatable RF coil with liquid metal and elastic silicone polymer. By
inflating, the coil closer to the sample, we have demonstrated the novel design
method for physically reconfigurable RF coils at a preclinical 7T.Acknowledgements
This work was supported by the National Institute of
Biomedical Imaging And Bioengineering of the National Institutes of Health
under Award Number R00EB020058.References
1. Motovilova,
Elizaveta, and Shao Ying Huang. 2020. "A Review on Reconfigurable Liquid
Dielectric Antennas" Materials 13, no. 8: 1863. https://doi.org/10.3390/ma13081863
2. Motovilova, E.,
Tan, E.T., Taracila, V. et al. Stretchable self-tuning MRI receive coils based
on liquid metal technology (LiquiTune). Sci Rep 11, 16228 (2021). https://doi.org/10.1038/s41598-021-95335-6
3. Martin JF, Hajek
P, Baker L, Gylys-Morin V, Fitzmorris-Glass R, Mattrey RR. Inflatable surface
coil for MR imaging of the prostate. Radiology. 1988 Apr;167(1):268-70. https://doi.org/10.1148/radiology.167.1.3347731 PMID: 3347731
4. Noworolski, S.M.,
Crane, J.C., Vigneron, D.B. and Kurhanewicz, J. (2008), A clinical comparison
of rigid and inflatable endorectal-coil probes for MRI and 3D MR spectroscopic
imaging (MRSI) of the prostate. J. Magn. Reson. Imaging, 27: 1077-1082. https://doi.org/10.1002/jmri.21331