Georgiy Alekseevich Solomakha1, Dario Bosch 1,2, Klaus Scheffler1,2, and Nikolai Ivanovich Avdievich 1
1High-field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany, 2Department for Biomedical Magnetic Resonance, University of Tübingen, Tuebingen, Germany
Synopsis
Keywords: RF Arrays & Systems, RF Arrays & Systems
Arrays of dipole
antennas were recently introduced as transceiver RF coils for human head
imaging at UHF as a simple and robust alternative to loop arrays. Due to the
head size, dipoles should be significantly shorter than λ/2 at working
frequency. Short dipoles suffer from high SAR and insufficient brain coverage.
In addition, since head arrays are usually placed on rigid holders, the resonance
frequency of dipoles change drastically with head size variation. In this work,
we developed a coaxial dipole array for human head imaging at 9.4T. The
developed coil provides whole-brain coverage, low SAR, and low frequency
variation.
Purpose
To evaluate the coaxial dipole antennas as transceiver elements of a human head array at 9.4T. Introduction
Arrays of dipole
antennas were recently introduced as transmit or transceiver RF coils for human
head imaging at ultra-high field (UHF) (i.e., larger than 7T) (1,2) as a simple
and robust alternative to loop arrays. Due to the sample size, dipole antennas
for head imaging should be significantly shorter than λ/2 at working frequency.
Such short dipoles (around 17 cm) suffer from high SAR and insufficient brain
coverage due to the “triangular” current distribution with a relatively sharp
peak at the center (3). In addition, since head arrays are usually placed on
rigid holders, the resonance frequency of dipoles could change drastically with
the head size variation. To overcome these issues, previously we developed a
novel antenna design by folding ends of the common straight dipole and moving
them away from the subject, i.e. a so-called folded-end dipole (3,4,5). This
design, however, has a disadvantage because folded ends require additional
space inside the coil housing. Another method to flatten and extend the current
distribution longitudinally is using coaxial dipoles in combination with lumped
elements (6). In this work, we developed a 9.4T coaxial human head transceiver
dipole array. To our best knowledge, this is a first example of using coaxial
dipoles for human head MRI at UHF. Methods
The developed array
consisted of eight coaxial dipoles uniformly surrounding the head. Following
work (6), the shield of each coaxial dipole had two gaps at small distances
from both ends and inductors connecting the core and the shield (Fig.1A). We
also modified the original geometry of coaxial dipoles (6) by shorting the
inner and outer conductors at the feeding point (Fig .A). Distances 2 cm and 3
cm from the end to the gap were considered in this work. Smaller distances
required very large inductances, while larger distances did not cause
significant extension of the current distribution. The total length of the
coaxial dipole was 19 cm. For comparison, we also evaluated a folded-end
dipole, straight dipole, and straight dipole with lumped inductors placed near
the ends (Figs.1A and 1B). For all dipoles, we assess the frequency variations
using two setups consisting of single dipole elements and cylindrical phantoms
(Fig.1B) (ε=58.6, σ=0.64 S/m, 17.6 and 14.2 cm in diameter) mimicking small and
large heads. Frequency shifts measured for the coaxial and folded-end dipoles
were similar and about two times smaller than that obtained for the straight
dipole (Fig.1C). We also evaluated and compared current distributions along the
dipole’s conductors and SAR-efficiencies (<B1+>/√pSAR, where pSAR is a local peak SAR). <B1+>
was averaged over 13-cm transversal slab, which includes the majority of the
brain. Based on these results, we chose
the optimal coaxial dipole design with the gap position of 2 cm and inductance
of 40 nH. Figs 1D-1F show the simulation model and photos of the final
8-element coaxial dipole array design. The array performance was compared
to that of the previously developed eight-element folded-end dipole array of
the similar size. B1+ and
pSAR for both arrays driven in the CP-mode (Fig.1D) were simulated numerically
using a head and shoulder (HS) phantom (Fig.1E) mimicking the human tissue
property and the Duke voxel model. All numerical simulations were done using
CST Studio 2021 (Dassault Systèmes, Vélizy-Villacoublay, France). All data were acquired using Siemens Magnetom 9.4T full
body scanner. In the experiment, B1+ was measured using a
satTFL sequence (7) (TR=2.5ms, TE=0.73ms, GRAPPA 2x2, FA=2˚/70˚, matrix:
64x64x64, resolution 3.5mm isotropic) for the HS phantom and healthy volunteer.
In-vivo T1-weighted and T2*-weighted images of a healthy volunteer were
acquired using MPRAGE (TI=1340ms, TR=3360ms, GRAPPA 2, matrix 264x264x224mm,
resolution 0.8mm isotropic) and GRE (TR=11ms, TE=7ms, GRAPPA 2, FA=5˚, matrix
264x264x224mm, resolution 0.8mm isotropic).Results and Discussion
Figs.2 and 3 show
SAR-efficiencies and corresponding current distributions for the single
coaxial, folded-end, and straight dipoles all loaded by the cylindrical phantom
(Fig.1B). While the current distribution
of the straight dipole has an almost triangular shape, using coaxial dipole or
straight dipole with lumped inductors allow flattening the current distribution
and reducing SAR. As seen in Fig.2, straight dipoles with inductances at the
ends also provide an extended current distribution, but the required inductance
is much higher (290 vs 40nH). Smaller inductances allow for a decrease in coil
losses. 8-element coaxial dipole array loaded by the Duke voxel model provides
<B1+>
of 0.395 µT/√W. In contrast, folded-end dipoles provide 0.427 µT/√W. However,
the coaxial dipole design delivers 30% lower local pSAR value. As a result,
SAR-efficiency for the coaxial dipole array is 10% higher than that for the
folded dipole array. From the in-vivo measurements
(Fig 5A) coaxial dipole array can provide whole-brain coverage. Thus, our
numerical and experimental results show that the coaxial dipole array imaging
performance is very similar to the folded dipole array. Conclusion
We developed, constructed,
and evaluated the eight-element coaxial dipole array for human whole-brain imaging
at 9.4T. The developed array provides a more compact and convenient alternative
to the folded-end dipole design with an additional benefit of 10% higher SAR
efficiencyAcknowledgements
Financial support of the ERC Advanced Grant
“SpreadMRI”, No 834940 is gratefully acknowledged.References
1. Chen, Gang, et al. “A 7T 8 channel transmit-receive dipole
array for head imaging: dipole element and coil evaluation.” Proceedings
of the 22nd Annual Meeting of ISMRM, Milan, Italy. Vol. 621. 2014.
2. Clément, J., Gruetter, R., & Ipek, Ö. (2019). A
combined 32‐channel receive‐loops/8‐channel transmit‐dipoles coil array
for whole‐brain MR imaging at 7T. Magnetic resonance in medicine, 82(3),
1229-1241.
3. Avdievich,
Nikolai I., et al. “Unshielded bent folded‐end
dipole 9.4 T human head transceiver array decoupled using modified passive
dipoles.” Magnetic resonance in medicine 86.1 (2021):
581-597.
4.Avdievich,
Nikolai I., et al. “Evaluation of short folded
dipole antennas as receive elements of ultra‐high‐field human head
array.” Magnetic resonance in medicine 82.2 (2019):
811-824.
5.Avdievich,
Nikolai I., et al. “Folded‐end dipole transceiver
array for human whole‐brain imaging at 7 T.” NMR in Biomedicine 34.8
(2021): e4541.
van Leeuwen, Carel C., et al. “The Coax Dipole: A fully
flexible coaxial cable dipole antenna with flattened current distribution
for body imaging at 7 Tesla.” Magnetic Resonance in Medicine 87.1
(2022): 528-540.
Paper about the B1 mapping method