Suen Chen1, Haoyu Sun1, Hao Chen1, Suma Anand2, Michael Lustig2, and Zhiyong Zhang1
1School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China, 2Department of Electrical Engineering and Computer Sciences, University of California, Berkeley, Berkeley, CA, United States
Synopsis
Keywords: Motion Correction, Low-Field MRI, Pilot Tone
To improve the image quality in low field, the conventional method
is to repeat scanning many times, and finally increase the SNR by averaging
multiple imaging data. Unfortunately, long scan time can make images highly
susceptible to motion artifacts. A recent contact-free motion detection
technology Beat Pilot Tone (BPT) improves the sensitivity compared with Pilot
Tone (PT) and is not limited by Larmor frequency. We introduce BPT in a 0.25T low-field
MRI system, and successfully reduce the motion artifacts while improving SNR by
binning the continuously acquired data into different motion states in image
domain and k-space via BPT signal.
Introduction
Motion sensing
technologies are important in MRI to reduce the motion
artifacts caused by unavoidable physiological movements
such as cardiac and respiratory motion. At low-field MRI, motion detection is
significant because it may suffer more motion artifacts besides physiological
signals for repeated and longer scanning[1]. This is a compromise to improve its low SNR.
Recently a contact-free motion sensor Beat Pilot Tone (BPT) is proposed[2]. It overcomes Larmor frequency limitation and further
improves motion detection sensitivity compared with PT[3-5]. At present, BPT has
only been tested in the 3T system, and its performance and application prospects in
low-field have not been exploited.
In this work, we applied the BPT technique to a 0.25T
low-field MRI system in order to reduce motion artifacts while improving the
SNR of low-field images. The essence was to bin continuously acquired data into
different motion states by the acquired BPT signal which carries motion
information. Two strategies were adopted. The first work was done in the image
domain based on fast EPI acquisition. The motion state of each image was detected
by the acquired BPT signal, and images in similar states were averaged. The second work
was carried out in k-space. After being classified by the BPT signal, the phase
encoding lines with similar motion states were rearranged and re-filled into one
k-space, so as to reconstruct images in this state. Methods
Hardware
Figure
1 shows the hardware system. We used ADF4350 and HMC833 to generate two signals
separated by the frequency of fBPT.
The two signals were amplified respectively, combined, high-pass filtered, and
transmitted to an omnidirectional antenna.
At
the receiving end, due to the nonlinear characteristics of the preamplifier,
the two signals modulated by the motion information will produce a second-order
intermodulation effect, thus the BPT signal within the MR receiver bandwidth is obtained. The signal carries high-precision human motion information and is
then further processed. In the experiment,
is set at 10.87MHz for the head coil and 10.93MHz
for the abdominal coil based on the center frequency which is changed by loading.
Adjust the location of the antenna carefully so that the patient's motion
information can be modulated when the transmitted signals propagate since receiving
coils in the low field have fewer channels than in 3T MRI.
Imaging experiments
EPI
(slice=5, TR for each slice=420ms, repeat 300 times) and GRE (single slice, TR=30ms,
repeat 48 times) were performed on our prototype 0.25T system. Two scans were
acquired on three volunteers with informed consent: a head scan using a
two-channel head coil and an abdominal scan using a two-channel abdominal coil.
In the first experiment, the volunteers stayed still for ~30s, then rotating head
continuously (left/right). In the second experiment, the volunteers hold their
breath for ~20s and then breathe normally.
Data analysis
Due
to the high acquisition efficiency of EPI, assuming that single slice scan would
not be affected by motion, the motion states of different slices and repetition
times were different. The images of the same slice with similar BPT signals were
averaged directly in the image domain. As for GRE acquisition, there are still
non-negligible motions in
single-slice phase encodings. BPT signal is used to rearrange the phase-encoding-line
data into the corresponding k-space by their motion states and refill them in a
motion-solved k space.
Results and Discussion
Figure 2a and b show the BPT signal imposed on the image and
frequency domain. The received BPT signal was sufficiently spaced from the
image signal in the frequency domain and could be easily extracted. Figure 2c lists
the extracted BPT signal in the abdominal scan at the 0.25T system, the motion
information of the breath-hold state and free-breathing periods can be clearly
observed. This demonstrates the feasibility and high sensitivity of BPT in low-field
MRI.
Figure 3 shows the results of EPI acquisition. The blurring caused
by motion severely degraded using the simply averaging image. With BPT-guided averaging,
the SNR in the averaged image is significantly improved. This result
demonstrates the feasibility of BPT for signal averaging under motion.
Figure 4 shows the results of GRE reconstructed with the motion
state extracted from BPT. Motion artifacts are not visible in the rearranged
images.Conclusions
A contact-free and plug-and-play BPT technique at low-field
MRI was proposed. The results demonstrate that BPT keeps its high sensitivity to motion detection
in the 0.25T system. Based on EPI and GRE, we improve SNR by averaging BPT-guided
images or rearranging k-space signal. Acknowledgements
This work is supported
by the National Natural Science Foundation of China National Science Foundation
of China (No. 62001290), Shanghai Science and Technology Development Funds
(21DZ1100300), and Shanghai Sailing Program (20YF1420900).References
[1] MARQUES
J P, SIMONIS F F, WEBB A G. Low‐field MRI: An MR physics perspective. Journal of
magnetic resonance imaging, 2019, 49(6): 1528-42.
[2] ANAND S, LUSTIG M. Beat Pilot Tone:
Exploiting Preamplifier Intermodulation of UHF/SHF RF for Improved Motion
Sensitivity over Pilot Tone Navigators; proceedings of the ISMRM Meeting and
Exhibition Abstract, 2021 .
[3] GRAESSLIN I, MENS G, GUILLAUME A, et al.
Advancements in contact‐free respiration monitoring using RF pick‐up coils; proceedings
of the Proceedings of the 18th Annual meeting of ISMRM, Stockholm, Sweden, 2010 .
[4] LENK M C. Respiratory Motion Tracking in
Magnetic Resonance Imaging with Pilot Tone Technology ; The Ohio State
University, 2018.
[5] SOLOMON E, RIGIE D S, VAHLE T, et al.
Free‐breathing radial
imaging using a pilot‐tone radiofrequency transmitter for detection of
respiratory motion. Magnetic resonance in medicine, 2021, 85(5): 2672-85.