Malte Laustsen1,2, Jakob Slipsager2, Thomas Gaaß2, Robert Frost3,4, André van der Kouwe3,4, Melanie Ganz5,6, and Lars G. Hanson1
1Magnetic Resonance Section, DTU Health Tech, Technical University of Denmark, Kgs. Lyngby, Denmark, 2TracInnovations, Ballerup, Denmark, 3Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 4Department of Radiology, Harvard Medical School, Boston, MA, United States, 5Neurobiology Research Unit, Copenhagen University Hospital, Rigshospitalet, Copenhagen, Denmark, 6Department of Computer Science, University of Copenhagen, Copenhagen, Denmark
Synopsis
Keywords: Motion Correction, Artifacts
Retrospective
motion correction (RMC) can substantially reduce motion artifacts in 3D brain
MRI. However, for extensive motion, RMC performance is limited. We evaluate RMC
with selective reacquisition (RMC+reacq) to expand the range of correctable
motion, while directly comparing to prospective motion correction (PMC) for
volumetric brain MRI, using external motion tracking. Both approaches lead to
significant image quality improvement and the performance of RMC+reacq and PMC was
only found statistically significant in 1 of 9 comparisons. These results
suggest that RMC with selective reacquisition can match the performance of PMC
for 3D-MPRAGE and 3D-FLAIR sequences.
Introduction
Head motion
remains a major challenge for brain MRI. Prospective and retrospective motion
correction (PMC, RMC) aims to correct for motion during and after data
acquisition, respectively. While PMC is most general, its dependence on MR-scanner’s complex real-time
feedback architecture, and the “all-or-nothing” character of prospective
correction can be major roadblocks for adoption. Using RMC, head rotation can
cause Nyquist violations in k-space and spin history effects when confronted
with extensive motion1, but RMC does not require high
computational speed of motion estimation, preserves the original data, and can
be implemented independent of scanner software.
We evaluate
the performance of RMC combined with selective reacquisition. We hypothesize that
the reacquisition of data affected by the most severe pose changes will limit
the remaining motion artifacts to a level manageable by RMC for 3D-MPRAGE and
3D-FLAIR sequences.Methods
MRI and
head motion tracking data were acquired from 22 healthy adults (16 females, 6
males) aged 23.5 ± 4.3 years. For details on recruitment and protocol please
refer to 2,3. Data from 3 subjects were discarded
due to inaccurate synchronization between the scanner and the external motion
tracker needed for RMC. A total of 122 scans from 19 subjects were analyzed.
Rigid-body
head motion was measured with the optical motion tracking system Tracoline
TCL 3.24 (TracInnovations, Ballerup,
Denmark). MRI was performed using a 3T
Siemens Magnetom Prisma scanner with a 64-channel head coil (Siemens Healthcare
GmbH, Erlangen, Germany) and prototype sequences with PMC and selective
reacquisition capabilities (3D-MPRAGE, 3D-FLAIR)5. Motion experiments consisted of a still
reference scan, a scan with nodding motion (“YES”), and a scan with shaking motion (“NO”) (only
for 3D-MPRAGE). Motion scans were repeated with and
without PMC (“PMC-ON”,
“PMC-OFF”).
After each
acquisition, an additional 30 seconds was spent reacquiring the most
motion-corrupted portions of k-space (16 k-space planes) based on a periodic motion
score6. Original and reacquired data were
corrected for motion using the same method (RMC or PMC).
PMC-OFF scans were retrospectively
corrected for motion using pose estimates generated by the motion tracking system.
For PMC, the tracking is smoothed with a 20-point moving median filter with a period of
670 ms. For RMC, tracking is smoothed with a 4-pole non-causal Butterworth low-pass filter with a cut-off of 0.5
Hz. All images were generated using the same off-scanner reconstruction
pipeline.
Three image
quality metrics were calculated for each corrected image using 7: structural similarity index
(SSIM), peak signal-to-noise (PSNR), and Tenengrad (TGRAD) chosen based on superior
correlation with radiologist scoring8. Average edge strength (AES) was
also calculated but did not show any significant difference to uncorrected data
independent of correction method and was therefore not included. Before
calculation of quality metrics, all images were aligned to the reference image
using sinc interpolation (SimpleITK rigid registration). For each subject, images were masked using a common brain mask derived from the uncorrected reference
image (Freesurfer skull strip). For AES and TGRAD, the mask was used to crop
image dimensions tightly around the brain, rather than region nulling, to avoid
introducing false edges that can contribute to image-gradient-based metrics.
Statistical
significance of quality metrics was evaluated with Wilcoxon signed rank test
between no correction and each correction method, between RMC and PMC, and
between RMC+reacq and PMC, compensated for multiple comparisons (false
detection rate: 0.05).Results
Figure 1
shows examples of each correction method for one patient. Figure 2 shows motion
tracking with/without substituting reacquired data. Figure 3 provides quality
metric data for each reconstructed image, and Figure 4 shows ranges and
statistical significance for each quality metric difference.
RMC+reacq, and PMC+reacq show significant
improvement over no correction
for all quality metrics and experiments, whereas RMC and PMC show significant
improvement in all but three comparisons. RMC and PMC showed significant improvement
in five comparisons. Comparison between RMC+reacq and PMC showed no significant
difference except RMC+reacq being favored in one comparison.Discussion
In this
work, we based the comparison of two motion correction methods on volunteers
reproducing prescribed motion patterns. While this pattern is not
representative of all patient motion, it serves as a test case for the used
methods favoring none of the correction approaches. The performance of RMC with
reacquisition depends strongly on the specific characteristics and the duration
of motion. The performance also depends on motion during the 30 second
reacquisition period after the host sequence, which can be mitigated by gating
the reacquisition with the motion tracking. In these motion experiments, the
volunteers were always closer to their original position during the
reacquisition period, which is not guaranteed in practice. The criteria for
reacquisition used here are designed for PMC. A reacquisition criterion that
seeks to fill undersampled regions of k-space might improve RMC+reacq
performance and allow for shorter reacquisition times.Conclusion
Retrospective
motion correction combined with selective reacquisition (RMC+reacq) can extend
the working range of RMC by reducing the maximum motion magnitude affecting
reconstructed data. In these motion experiments, RMC+reacq matched the performance
of PMC for 3D-MPRAGE and 3D-FLAIR. The approach could reduce requirements of real-time
computation and updating, and retain the original data while incurring a low penalty to overall
scan time.Acknowledgements
No acknowledgement found.References
1. Slipsager JM, Glimberg SL, Højgaard L, et al. Comparison of
prospective and retrospective motion correction in 3D-encoded neuroanatomical
MRI. Magn. Reson. Med. 2022;87:629–645 doi: 10.1002/mrm.28991.
2. Eichhorn H, Frost R, Kongsgaard A, et
al. Evaluating the performance of markerless prospective motion correction and
selective reacquisition in a general clinical protocol for brain MRI. PsyArXiv.
Nov 9, 2022. web: psyarxiv.com/vzh4g, doi: 10.31234/osf.io/vzh4g.
3. https://openneuro.org/datasets/ds004332/versions/1.0.1.
4. Olesen OV, Paulsen RR, Hojgaard L,
Roed B, Larsen R. Motion Tracking for Medical Imaging: A Nonvisible Structured
Light Tracking Approach. IEEE Trans. Med. Imaging 2012;31:79–87 doi:
10.1109/TMI.2011.2165157.
5. Frost R, Wighton P, Karahanoğlu FI, et
al. Markerless high-frequency prospective motion correction for neuroanatomical
MRI. Magn. Reson. Med. 2019;82:126–144 doi: 10.1002/mrm.27705.
6. Tisdall MD, Hess AT, Reuter M,
Meintjes EM, Fischl B, van der Kouwe AJW. Volumetric Navigators (vNavs) for
Prospective Motion Correction and Selective Reacquisition in Neuroanatomical
MRI. Magn. Reson. Med. 2012;68:389–399 doi: 10.1002/mrm.23228.
7.
https://github.com/melanieganz/MoCoProject/tree/main/ImageQualityMetrics.
8. Eichhorn H, Chemnitz-Thomsen S, Vouros
E, et al. Evaluating the match of image quality metrics with radiological
assessment in a dataset with and without motion artifacts. In: Joint Annual
Meeting ISMRM-ESMRMB 2022. London, UK.