Keywords: Flow, Velocity & Flow
Background phase errors in PC-MRI flow measurements are caused by eddy currents and mechanical vibrations that produce unwanted magnetic fields, thereby reducing velocity measurement accuracy. In this work, we demonstrate that a GIRF measurement can be used to predict the PC-MRI background phase, then used to design gradient optimized (GrOpt) velocity encoding waveforms that minimize these errors. The method is tested in static phantoms and in vivo, where 4.8x to 6.4x reductions in background velocity are seen. Phantom results showed a reduction from 0.74±0.22 cm/s to 0.15±0.12 cm/s, and in vivo results showed reductions from 0.93±0.10 cm/s to 0.13±0.08 cm/s.
[1] Chernobelsky, A., Shubayev, O., Comeau, C. R. & Wolff, S. D. Baseline correction of phase contrast images improves quantification of blood flow in the great vessels. Journal of Cardiovascular Magnetic Resonance 9, 681–685 (2007).
[2] Walker, P. G. et al. Semiautomated method for noise reduction and background phase error correction in MR phase velocity data. J Magn Reson Imaging 3, 521–530 (1993).
[3] Dillinger, H., Peper, E., Guenthner, C. & Kozerke, S. Background Phase Error Reduction in Phase-Contrast MRI based on Acoustic Noise Recordings. in ISMRM Annual Meeting, Remote (2021).
[4] Loecher, M., Middione, M. J. & Ennis, D. B. A gradient optimization toolbox for general purpose time-optimal MRI gradient waveform design. Magn Reson Med 84, 3234–3245 (2020).
[5] Rahmer, J., Mazurkewitz, P., Börnert, P. & Nielsen, T. Rapid acquisition of the 3D MRI gradient impulse response function using a simple phantom measurement. Magnetic Resonance in Medicine 82, 2146–2159 (2019).
[6] Layton, K. J. et al. Pulseq: a rapid and hardware-independent pulse sequence prototyping framework. Magnetic resonance in medicine 77, 1544–1552 (2017).
[7] Gatehouse, P. D. et al. Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements. Journal of Cardiovascular Magnetic Resonance 12, 5 (2010).
[8] Loecher, M., Magrath, P., Aliotta, E. & Ennis, D. B. Time-optimized 4D phase contrast MRI with real-time convex optimization of gradient waveforms and fast excitation methods. Magnetic Resonance in Medicine 82, 213–224 (2019).
[9] van Gorkum, R. J. H., Guenthner, C., Koethe, A., Stoeck, C. T. & Kozerke, S. Characterization and correction of diffusion gradient-induced eddy currents in second-order motion-compensated echo-planar and spiral cardiac DTI. Magnetic Resonance in Medicine 88, 2378–2394 (2022).
Figure 1: Demonstration of the GIRF-predicted velocity for a bipolar phase-contrast experiment. a) Slice select gradient waveforms, g(t), for two TRs with opposite bipolar velocity encoding. b) measured GIRF, h(t), for zero-order fields. c) convolution of the gradient waveform with the GIRF predicts the field response. d) and e) cumulative phase effect for each individual TR from the time of excitation. f) predicted background velocity after complex differencing and converting to velocity. The red band shows an example of a window around the echo time to be minimized.
Figure 2: The GrOpt gradient waveform optimization results for the win2 position (a,c,e), and the predicted background velocity for each waveform (b,d,f). Red arrows show bipolars, green arrow show spoilers. The red band shows the minimization window. The top row is for conventional bipolar encoding waveforms, showing the non-zero BGv expected. In the next two rows, the gradient waveforms were optimized to minimize the background velocity within the respective windows. The middle row only optimized the bipolars, and the bottom row optimized both the bipolars and spoilers.
Figure 3: Static phantom velocity maps for different TEs (columns) and a range of PC-MRI gradient waveforms (rows). Only 10 of the 40 TEs acquired are shown. Each row represents a different gradient optimized (GrOpt) waveform. Red boxes highlight velocity maps with echo times that fall within the intended window for which the background velocity approaches zero. Background velocities are generally larger for the other evaluated TEs. When including spoilers in the optimization the response also changes as predicted (Fig 2) and maintains the nulling performance.
Figure 4: Background velocity in a static phantom from the multi-TE imaging experiment. Red lines are the measured 0th order background velocity across the entire phantom for each TE. The blue line shows the GIRF predicted response. The horizontal dashed lines show the ±0.6 cm/s window for acceptable background velocity offsets [7]. Red bands show the minimization window and demonstrate near zero background velocity offsets for both measured and predicted responses.