David Leitão1, Daniel West1, Raphael Tomi-Tricot2,3, Jo Hajnal1,3, Tobias C Wood4, and Shaihan Malik1,3
1Biomedical Engineering & Imaging Sciences, King's College London, London, United Kingdom, 2MR Research Collaborations, Siemens Healthcare, Frimley, United Kingdom, 3Centre for the Developing Brain, King's College London, London, United Kingdom, 4Department of Neuroimaging, King's College London, London, United Kingdom
Synopsis
Keywords: Magnetization transfer, Magnetization transfer, Low-Field MRI
We used multiband pulses to obtain Magnetization Transfer (MT) weighted
images at 0.55T in a time-efficient manner. White Matter (WM)/Grey Matter (GM)
contrast is compared across several flip-angles for both GRE and bSSFP
sequences. GRE showed larger MTR WM/GM difference but after accounting for
signal-to-noise ratio bSSFP had 50% better contrast-to-noise ratio.
Introduction
Low-field
MRI brings new opportunities in interventional and diagnostic imaging1. Despite lower signal-to-noise ratio
(SNR), longitudinal magnetization recovers faster (shorter T1)
while longer T2 values means
signal can be measured for longer2.
This is particularly beneficial for bSSFP whose signal amplitude is T2/T1-weighted3 but also for GRE. Another
advantage of low field-strength is
lower specific absorption rate (SAR) which allows higher root mean square $$$\mathrm{B_1^+}$$$ amplitude. This is beneficial for
Magnetization Transfer (MT) imaging which uses high power pulses to saturate
semisolid magnetization and thus generate MT contrast. Here we exploit these advantages to
demonstrate high Contrast-to-Noise MTR imaging with GRE and bSSFP at 0.55T.Methods
Magnetization Transfer Ratio (MTR) maps were acquired for the
brain over two scanning sessions on the same healthy volunteer at a clinical 0.55T system (MAGNETOM
Free.Max, Siemens Healthineers, Erlangen, Germany) using a 6-channel head coil.
3D GRE and bSSFP sequences were acquired in combination with non-selective
multiband (MB) pulses to simultaneously excite the free-water magnetization and
saturate the semisolid magnetization4 (Figure 1) – i.e. the pulses were not used for spatial encoding, only for
contrast generation. These pulses offer an efficient means for generating MT
contrast within rapid gradient echo sequences5.
Both GRE and bSSFP had identical settings (cartesian encoding, sagittal
orientation, $$$\mathrm{2mm}$$$ isotropic resolution, 3 averages, $$$\mathrm{TE=3.6ms}$$$, $$$\mathrm{BW_{pixel}=429Hz}$$$) with exception of the
TR ($$$ \mathrm{TR_{GRE}=7.96ms/TR_{bSSFP}=7.19ms}$$$) but resulting in very
similar acquisition times per volume ($$$ \mathrm{T_{acq,GRE}=4m59s/T_{acq,bSSFP}=4m53s}$$$). To investigate the
contrast dependence on the flip angle, GRE and bSSFP were acquired using
several flip angles: $$$\mathrm{\alpha_{GRE}=\{5^\circ,10^\circ,15^\circ,20^\circ,25^\circ\}}$$$ and $$$\mathrm{\alpha_{bSSFP}=\{10^\circ,20^\circ,30^\circ,40^\circ,50^\circ,60^\circ\}}$$$.
MTRs were measured by acquiring images with
either a single band (SB) pulse used only for excitation, or a MB pulse
containing both the on-resonance excitation and off-resonance saturation bands
(Figure 1). The MB pulse was $$$\mathrm{3ms}$$$ long with its side bands at $$$\mathrm{\pm3kHz}$$$, used a Gaussian envelope ($$$\mathrm{TBP=2.26}$$$) and was designed to achieve $$$\mathrm{B_1^{rms}=3.5\mu{}T}$$$. Since the on-resonance band also contributes
to $$$\mathrm{B_1^{rms}}$$$ the pulses with lower flip angle employed
larger amplitude off-resonance bands than the higher flip angle pulses. The
non-MT-weighted volume used a SB pulse with the same duration and TBP. The MTR was calculated using the expression:
$$\mathrm{MTR(\%)=100\times{}\frac{S_{SB}-S_{MB}}{S_{SB}}}\;\;[1]$$
where $$$\mathrm{S_*}$$$ are the signals acquired with the SB and MB pulses.
To compare the contrast across
sequences and flip angles, two ROIs were drawn
manually (Figure 2A) around white matter (WM) and gray matter (GM) to estimate
the MTR. The average $$$\mu{}$$$ and standard deviation $$$\sigma$$$ of MTR were calculated over the ROIs, and the
difference $$$\mathrm{\Delta{}MTR}$$$ between WM-GM MTR was calculated:
$$\mathrm{\Delta{}MTR=\mu_{WM}-\mu_{GM}}\;\;[2]$$
To account for differences in SNR, the contrast-to-noise ratio
(CNR) was also calculated:
$$\mathrm{CNR=\frac{\Delta{}MTR}{\sqrt{\sigma_{WM}^2+\sigma_{GM}^2}}}\;\;[3]$$Results
MTR maps acquired with bSSFP and GRE are shown in Figures 3 and 4
respectively. Overall MTR values decreased with increasing flip angles (Figure
5A,B). However, as the flip angle increases the difference in MTR between WM
and GM increases substantially for GRE (Figure 5C). When accounting for the SNR
the bSSFP has superior WM-GM discernment for an optimal flip angle of $$$30^\circ$$$ ($$$\mathrm{CNR=4.9}$$$), whereas the CNR for GRE is $$$50\%$$$ smaller ($$$\mathrm{CNR=3.2}$$$) for an optimal flip angle of $$$15^\circ$$$.Discussion and Conclusion
In this work MT contrast was achieved by off-resonance
saturation with non-selective MB pulses which are a time efficient way to
simultaneously excite free-water magnetization and saturate semisolid
magnetization. The MB pulses applied a constant $$$\mathrm{B_1^{rms}}$$$ which was maximized to the
system limits. Thus, these pulses created a variable amount of off-resonance
semisolid saturation across different flip angles as the power is moved from
off-resonance to on-resonance with increasing flip angle (Figure 1). Assuming a
constant absorption lineshape, the net semisolid saturation induced by the
MT-weighted scan is the same at all flip angles, but for the higher flip angles
the ‘reference’ image is itself more strongly MT weighted, thus reducing the
measured MT ratio.
Both GRE and bSSFP showed very
uniform MTR maps with virtually no $$$\mathrm{B_1^+}$$$ inhomogeneity and a strong MTR
dependence on the flip angle, with smaller MTR for larger flip angles (Figure
5A,B) but difference between WM and GM difference improving for GRE (Figure
5C). However, in terms of CNR bSSFP proved superior to GRE, which was expected
given its intrinsic higher SNR. This agrees with previous work on MTR imaging
at 0.55T6
where MT contrast was obtained via
on-resonance saturation by scaling the pulses amplitude/duration.
Independently of the saturation scheme, MT contrast increases with the applied
RF power and in this work we found that the maximum achievable $$$\mathrm{B_1^{rms}}$$$ was limited by the hardware and
not by SAR constraints. Therefore, there is potential to safely increase MT
contrast.
Future work will explore more
efficient spatial encoding like spiral7, and other
imaging modalities that could benefit from high $$$\mathrm{B_1^+}$$$ fields like inhomogeneous MT8.Acknowledgements
The research was funded/supported by core funding from the Wellcome/EPSRC Centre for Medical Engineering [WT203148/Z/16/Z] and by the National Institute for Health Research (NIHR) Biomedical Research Centre based at Guy's and St Thomas' NHS Foundation Trust and King's College London and/or the NIHR Clinical Research Facility. The views expressed are those of the author(s) and not necessarily those of the NHS, the NIHR or the Department of Health and Social Care.References
1. Campbell-Washburn, A. E., Ramasawmy,
R. & Restivo, M. C. Opportunities in Interventional and Diagnostic Imaging
by. Radiology 2–11 (2019).
2. Marques, J. P., Simonis, F. F. J.
& Webb, A. G. Low-field MRI: An MR physics perspective. Journal of
Magnetic Resonance Imaging 49, 1528–1542 (2019).
3. Bieri, O. & Scheffler, K.
Fundamentals of balanced steady state free precession MRI. Journal of
Magnetic Resonance Imaging 38, 2–11 (2013).
4. Teixeira, R. P. A. G., Malik, S. J.
& Hajnal, J. v. Fast quantitative MRI using controlled saturation
magnetization transfer. Magn Reson Med 81, 907–920 (2019).
5. Malik, S. J., Teixeira, R. P. A. G.,
West, D. J., Wood, T. C. & Hajnal, J. v. Steady-state imaging with
inhomogeneous magnetization transfer contrast using multiband radiofrequency
pulses. Magn Reson Med 83, 935–949 (2020).
6. Afshari, R., Bauman, G. & Bieri,
O. On the Prospects of Magnetization Transfer Imaging at 0.55T. in Proc.
Intl. Soc. Mag. Reson. Med. 30 2710 (2022).
7. Restivo, M. C., Ramasawmy, R.,
Bandettini, W. P., Herzka, D. A. & Campbell-Washburn, A. E. Efficient
spiral in-out and EPI balanced steady-state free precession cine imaging using
a high-performance 0.55T MRI. Magn Reson Med 84, 2364–2375
(2020).
8. Girard, O. M. et al.
Magnetization transfer from inhomogeneously broadened lines (ihMT):
Experimental optimization of saturation parameters for human brain imaging at
1.5 Tesla. Magn Reson Med 73, 2111–2121 (2015).