Tianle Cao1,2, Yibin Xie1, Debiao Li1, and Anthony G. Christodoulou1
1Cedars Sinai Medical Center, Los Angeles, CA, United States, 2University of California, Los Angeles, Los Angeles, CA, United States
Synopsis
Keywords: Sparse & Low-Rank Models, Sparse & Low-Rank Models
Low-rank tensor modelling is promising for multi-dimensional MR imaging. In this work, we developed a new low-rank tensor reconstruction approach using alternating minimization of spatial and temporal bases from the whole k-t space data instead of from split subsets of data. The approach was evaluated for 2D motion-resolved myocardial T1/T2/T2*/fat-fraction mapping and could potentially be used for imporving reconstruction quality and/or further reducing scan time.
Introduction
Multi-dimensional MRI has drawn increasing attention1-3.
Tensor and array modeling are crucial in enabling highly accelerated
acquisition by leveraging sparsity and/or image correlations4-6.
For example, MR Multitasking framework5
models underlying images as a low-rank tensor, first determining a temporal subspace from auxiliary navigator data, then performing
subspace-constrained reconstruction of spatial coefficient maps from
separate k-t space data.
However, neither step of this workflow is
performed from the complete acquired data, potentially introducing subspace
bias, especially at high undersampling rates. With the goal of improving
reconstruction quality and further reducing scan time, we propose a new image
reconstruction algorithm for MR Multitasking which estimates both the temporal
subspace and spatial coefficients from the complete acquired data. We
evaluated this algorithm for 2D myocardial
T1/T2/T2*/fat-fraction (FF) mapping7,
in a numerical phantom and in-vivo.Methods
Image acquisition
Data were acquired using an interleaved scheme7:
single-echo navigator data were repeatedly acquired at k-space center line to
collect temporally rich information and multi-echo radial imaging data were
acquired with golden angle increments for spatial information.
Image reconstruction
The
underlying image series at the $$$p^{th}$$$ echo is modeled
as a low-rank tensor with Tucker decomposition8, 9:
$$\mathcal{A}^{(p)}=\mathcal{G} \times_1 \mathbf{U}^{(1), p} \times_2 \mathbf{U}^{(2)} \times_3 \mathbf{U}^{(3)} \times_4 \mathbf{U}^{(4)} \times_5 \mathbf{U}^{(5)},\tag{1}$$
Where $$$\mathbf{U}^{(1),p}$$$ is the
spatial basis for the $$$p^{th}$$$
echo, $$$\mathbf{U}^{(2)}$$$ and $$$\mathbf{U}^{(3)}$$$ are the cardiac and respiratory motion bases, and $$$\mathbf{U}^{(4)}$$$ and $$$\mathbf{U}^{(5)}$$$ are the pre-computed basis functions for T1&T2 relaxation. In previous Multitasking implementations7, 1) $$$\mathcal{G}$$$, $$$\mathbf{U}^{(2)}$$$, and $$$\mathbf{U}^{(3)}$$$
were first determined solely from the single-echo
navigator data, via nuclear-norm based low-rank tensor completion and
high-order SVD; 2) spatial factors $$$\mathbf{U}^{(1),p}$$$ were
then estimated solely from the multi-echo imaging data with all temporal
components fixed.
In
this work, we improved both reconstruction steps, as shown in Fig. 1: 1)
$$$\mathcal{G}$$$, $$$\mathbf{U}^{(2)}$$$, and $$$\mathbf{U}^{(3)}$$$ were
initialized (but not fixed) from the navigator data using explicit rank
constraints and alternating minimization, i.e.:
$$\begin{aligned}&\hat{\mathcal{G}}, \widehat{\mathbf{U}}_{\text {nav}}^{(1)}, \widehat{\mathbf{U}}^{(2)}, \widehat{\mathbf{U}}^{(3)} \\&=\arg \min _{\mathcal{G}, \mathbf{U}_{\text {nav }}^{(1)}, \mathbf{U}^{(2)}, \mathbf{U}^{(3)}} \| \mathbf{d}_{\text {nav }} \\&-\Omega_{\text {nav }}\left(\mathcal{G}\times_1 \mathbf{U}_{\text {nav }}^{(1)} \times_2 \mathbf{U}^{(2)} \times_3 \mathbf{U}^{(3)} \times_4 \mathbf{U}^{(4)} \times_5 \mathbf{U}^{(5)}\right) \|_2^2+\lambda \sum_{i=2}^3 R_{\mathrm{t}}\left(\left(\mathcal{G} \times_1 \mathbf{U}_{\text {nav}}^{(1)} \times_2 \mathbf{U}^{(2)} \times_3 \mathbf{U}^{(3)} \times_4 \mathbf{U}^{(4)} \times_5 \mathbf{U}^{(5)}\right){ }_{(i)}\right)\end{aligned}\tag{2}$$
where $$$d_{nav}$$$ is the
acquired navigator data, $$$\Omega_{\text {nav }}(\cdot)$$$ is the
sampling operator, $$$(\cdot)_{(i)}$$$ is the mode-$$$i$$$ unfolding of
a tensor and $$$R_{\mathrm{t}}(\cdot)$$$ imposes total
variation regularization for motion. 2)
temporal factors $$$\mathcal{G}$$$, $$$\mathbf{U}^{(2)}$$$, and $$$\mathbf{U}^{(3)}$$$ and spatial
factors $$$\mathbf{U}^{(1),p}$$$ were updated
and calculated from the entire k-t space data through alternating least-squares minimization:
$$\hat{\mathcal{G}}, \widehat{\mathbf{U}}^{(1), \mathbf{p}}, \widehat{\mathbf{U}}^{(2)}, \widehat{\mathbf{U}}^{(3)}=\arg \min _{\mathcal{G},
\mathbf{U}^{(1),
p}, \mathbf{U}^{(2)},
\mathbf{U}^{(3)}}
\left\|\mathbf{d}^{(\mathbf{p})}-\Omega
\mathbf{F}
\mathbf{S}\left(\mathcal{G}
\times_1 \mathbf{U}^{(1),
\mathbf{p}}
\times_2 \mathbf{U}^{(2)}
\times_3 \mathbf{U}^{(3)}
\times_4 \mathbf{U}^{(4)}
\times_5 \mathbf{U}^{(5)}\right)\right\|_2^2
+\lambda R_{\mathrm{S}}\left(\mathbf{U}^{(1),
\mathbf{p}}\right),\tag{3}
$$
where $$$\mathbf{d}^{(\mathbf{p})}$$$ is the entire
k-space data for $$$p^{th}$$$ echo, $$$\mathbf{F}$$$ is the
Fourier sampling operator, $$$\mathbf{S}$$$ denotes
sensitivity maps, and $$$R_{\mathrm{s}}(\cdot)$$$ is a wavelet
sparsity regularizer.
Numerical simulations
A numerical phantom was created from XCAT
phantom10 with 20 cardiac phases for each of the 6
respiratory phases. The sequence diagram and acquisition scheme shown in
previous work7
was used and typical T1/T2/T2*/FF values at 3T were assigned. The average heart and respiration rates were set as 75 bpm and
15 bpm, respectively, with 10% standard deviation.
In-vivo study
One healthy volunteer was consented and scanned on a 3T scanner (MAGNETOM Vida, Siemens) at the mid-ventricular slice with an 18-channel body coil.
Analysis
All reconstructions were performed at two
different scan times (2.5min and 1.5min) to test their performances at different
acceleration factors. RMSE was used to evaluate the images from numerical
simulations. For in-vivo maps, myocardium was segmented based on the AHA
16-segment model. The intra-segment average and standard deviation (SD) were
compared between different approaches using two-way ANOVA test (N= 12, with 6
segments and 2 scan times). Results
Fig. 2 shows the reconstructed parametric maps and the
difference maps (x2) against gold standard. Lower residuals were seen in
maps generated from the proposed approach than those from the previous
approach. Table 1 shows that the proposed approach substantially reduces T1 and
T2 map error at both scan times, and better maintains performance when reducing
scan time from 2.5-min to 1.5-min. T2* and FF maps also show small improvements
in all but one case.
The in-vivo results are shown in Fig. 3. There is a
visual quality improvement in myocardial homogeneity with the proposed approach,
as pointed out by the arrows. The results of two-way ANOVA are summarized in
Table 2, where a significant reduction is found in intra-segment SD (92 ms vs. 135 ms) of T1 with the proposed approach. No significant difference was found
for other parameters.Discussion and Conclusion
In this work, we developed a new approach for low-rank
tensor reconstruction, which employed explicit rank constraints and provided
flexibility for simultaneous updates of spatial and temporal basis. The
approach can be seen as an extension of matrix recovery algorithm
PowerFactorization11
for tensor recovery. Validation was performed with numerical simulations and a
human subject, and the results indicated the potential for improving image
quality and/or reducing scan time. Although Multitasking data was used, the
approach should generalize to other low-rank tensor MRI methods. Further
in-vivo validations are warranted.Acknowledgements
This
work was supported by the National Institutes of Health (Grant/Award Nos.
R01EB028146 and R01HL156818).References
1. He
J, Liu Q, Christodoulou AG, Ma C, Lam F, Liang Z-P. Accelerated
high-dimensional MR imaging with sparse sampling using low-rank tensors. IEEE Trans Med Imaging. Sep
2016;35(9):2119-2129. doi:10.1109/Tmi.2016.2550204
2. Yaman B, Weingartner S, Kargas N,
Sidiropoulos ND, Akcakaya M. Low-Rank Tensor Models for Improved
Multi-Dimensional MRI: Application to Dynamic Cardiac T1 Mapping. IEEE Trans Comput Imaging.
2019;6:194-207. doi:10.1109/tci.2019.2940916
3. Ma D, Gulani V, Seiberlich N, et al.
Magnetic resonance fingerprinting. Nature.
Mar 14 2013;495(7440):187-192. doi:10.1038/nature11971
4. Feng L, Axel L, Chandarana H, Block
KT, Sodickson DK, Otazo R. XD-GRASP: Golden-Angle Radial MRI with
Reconstruction of Extra Motion-State Dimensions Using Compressed Sensing. Magn Reson Med. Feb 2016;75(2):775-788.
doi:10.1002/mrm.25665
5. Christodoulou AG, Shaw JL, Nguyen C,
et al. Magnetic resonance multitasking for motion-resolved quantitative
cardiovascular imaging. Nat Biomed Eng.
Apr 2018;2(4):215-226. doi:10.1038/s41551-018-0217-y
6. Bustin A, Lima da Cruz G, Jaubert O,
Lopez K, Botnar RM, Prieto C. High-dimensionality undersampled patch-based
reconstruction (HD-PROST) for accelerated multi-contrast MRI. Magn Reson Med. Jun
2019;81(6):3705-3719. doi:10.1002/mrm.27694
7. Cao T, Wang N, Kwan AC, et al.
Free-breathing, non-ECG, simultaneous myocardial T1, T2, T2*, and fat-fraction
mapping with motion-resolved cardiovascular MR multitasking. Magn Reson Med. Oct
2022;88(4):1748-1763. doi:10.1002/mrm.29351
8. Kolda TG, Bader BW. Tensor
Decompositions and Applications. Siam Rev.
Sep 2009;51(3):455-500. doi:10.1137/07070111x
9. Tucker LR. Some mathematical notes
on three-mode factor analysis. Psychometrika.
Sep 1966;31(3):279-311. doi:10.1007/BF02289464
10. Segars WP, Sturgeon G, Mendonca S,
Grimes J, Tsui BM. 4D XCAT phantom for multimodality imaging research. Med Phys. Sep 2010;37(9):4902-15.
doi:10.1118/1.3480985
11. Vidal
R, Tron R, Hartley R. Multiframe motion segmentation with missing data using
PowerFactorization and GPCA. Int J Comput
Vision. Aug 2008;79(1):85-105. doi:10.1007/s11263-007-0099-z