Adam Berrington1 and Olivier Mougin1
1Sir Peter Mansfield Imaging Centre, School of Physics and Astronomy, University of Nottingham, Nottingham, United Kingdom
Synopsis
Keywords: Signal Modeling, Spectroscopy
Performing simultaneous MRS from two-regions is often
desirable in spectroscopy studies. However, achieving an optimal B0 shim in two
regions is difficult, leading to poor data quality. We investigate a model-based
reconstruction for two-voxel MRS, incorporating B0 information, to reconstruct
data with reduced T2* decay. We test this approach using least squares and
regularization approaches for Hadamard encoding in simulations and phantom. Despite
a noise penalty for B0-correction (2-4)-fold, FIDs from two voxels better
matched un-broadened spectra under conditions of sufficient SNR. Regularization
also produced time-domain reconstructions with less noise than a least-squares approach.
Introduction
High-quality MRS data are often desired from two
regions of interest simultaneously, e.g. as a control region or in functional
MRS studies of metabolite response. Using multi-band excitation,
simultaneous two-voxel MRS can be achieved with the addition of spatial encoding
methods such as Hadamard1, PRIAM2 or recently vGRAPPA3. However, obtaining optimal static shims of the
B0 field is difficult across regions4, which limits data quality. To account for B0-induced
effects, knowledge of B0 inhomogeneity has previously been incorporated to
recover ‘natural linewidth’ spectra in MRSI5. Furthermore, modelling the effect of B0 led to
improved reconstruction of SLIM spectra6. In this work, we propose a model-based approach
which incorporates the measured B0 inhomogeneity in each region to improve reconstruction
of two-voxel Hadamard-encoded MRS data. Theory
To account for the effect of B0 field
inhomogeneity, the two-voxel MRS experiment is
formulated as a model-based reconstruction. The measured signal, $$$s_{n}(t)$$$, following
the $$$n$$$th encode, is a weighted sum of the desired signal, $$$x_{m}(t)$$$, from each voxel, $$$m$$$, such that, $$ s_n(t) = \sum_{m=1}^{2}H_{m,n}x_{m}(t)\int_{R_m}e^{-i[\phi(R) + 2\pi\Delta f(R) t]}dR $$ where $$$H_{m,n}$$$ is the Hadamard
encoding applied to the signal in the $$$m$$$th voxel during the $$$n$$$th
encode. The time-dependent complex exponential factor accounts for B0-field
inhomogeneity through off-resonance, $$$\Delta f(R)$$$, within the spatial
region bound by each MRS voxel, $$$R_{m}$$$. Spatial inhomogeneity
in the RF transmit phase, $$$\phi(R)$$$, is also
included.
The acquired signals are concatenated over the
two encodings, $$$\textbf{s} = [s_1(t), s_2(t)]^T$$$, and the signal
model is reformulated as a matrix equation, such that, $$ \mathbf{s}(t) = \mathbf{G}(t)\mathbf{x}(t) $$ where $$$\mathbf{G}(t)$$$ is the ‘geometry’ matrix containing the
Hadamard and time-dependent B0-terms for each voxel and encoding step. This equation can
be solved by least squares for each time point in the FID. Additionally, we
also investigate a Tikhonov regularization, similar to a previous approach5, such that, $$ \mathbf{x}(t) = \left\| \mathbf{G}(t)\mathbf{x}(t) - \mathbf{s}(t)\right\| + \lambda_t \left\| \mathbf{x}(t)\right\| $$ where $$$\lambda_t$$$ is a regularization parameter and $$$\left\| \cdot \right\|$$$ is the L2 norm.Methods
Simulations
Simulations of two-voxel acquisition with 4 different reconstruction
approaches were performed considering (i) separate acquisition (SEP), (ii) standard Hadamard
reconstruction (H), (iii) a time-dependent
least squares solution (B0-LS) and (iv)
the regularized approach (B0-REG) (Eq.
3). FIDs from two voxels (V1, V2) were simulated containing different amplitudes and chemical shifts (Fig. 1A). Each voxel comprised of 1000 isochromats with random frequency offsets, $$$\Delta f$$$, drawn
from a normal distribution ($$$\sigma_{V1}$$$ = 30
Hz,
$$$\sigma_{V2}$$$ = 15
Hz). Complex noise was added over 8 artificial receive channels by sampling
from a multivariate normal distribution. Monte-Carlo simulations were performed
by incrementing simulated SNR exponentially from 1 to 256 and repeating 10
times. The RMSE of the initial 20% of the reconstructed FIDs were compared to ground-truth
(GT) spectra without B0-inhomogeneity. In simulations, the optimal
regularization parameter was chosen automatically using the L-curve method7.
Phantom
Two-voxel data were also acquired from a homogeneous phantom
using a semi-LASER sequence (20x20x20mm3, TE/TR=28ms/6s, NSA=32) on
a Philips 7 T MR system (32-channel receive coil) with dual-band RF excitation for Hadamard encoding. Raw
data were loaded using NIFTI-MRS8, before being reconstructed using methods
(ii-iv) and coil-combined and eddy current corrected9. B0 shimming was performed over the region
spanning both voxels and shims were fixed during a subsequent 3D B0-mapping scan
(2mm isotropic, delta TE/TR=1/20ms), giving 1000 $$$\Delta f$$$ values per voxel.Results
Simulated
data were successfully reconstructed from both voxels (Fig 2A). The time-domain
reconstruction methods (B0-LS and B0-REG) resulted in visibly reduced B0-induced
T2* decay; returning the natural linewidth of the ground truth spectra. At
certain time-points, the reconstruction was poorly conditioned (Fig 2B, red
arrow), which led to noise amplification using least-squares (B0-LS) but not with
the regularized approach (B0-REG). Over all simulated SNR values (Fig 3A),
B0-LS reconstruction resulted in a large increase in noise 15x (p<10-18),
which was lower for B0-REG 3.8x (p=<10-18). Owing to the absence of
T2* decay in the FID, RMSE of B0-REG reconstruction from ground truth spectra
was lower than for standard Hadamard reconstruction for SNR values >16 (Fig.
3B).
Figure 4
shows the reconstructed phantom data. Despite the fact that each voxel had
narrow frequency distribution (2.4 Hz and 4 Hz) (Fig 4B) the reconstructed spectra using
B0-REG resulted in visibly narrower linewidths with a noise increase of 2.1x
for B0-REG. B0-LS also led to amplification of spurious signal in the middle of the
FID.Discussion & Conclusion
We have demonstrated a model-based approach to account for the effects of B0 inhomogeneity on signals acquired simultaneously using two-voxel MRS. Incorporating knowledge of B0 led to spectra with reduced B0-induced T2* decay, similar to previous work5. The regularized approach resulted in less noise than least-squares reconstruction in simulation (Fig. 2B) and phantom. Despite this, a noise penalty (3-4x in phantom) was still introduced relative to standard reconstruction. Experimental data and simulations (Fig. 3) indicate the method works when SNR is sufficient, however, further work is required to investigate the trade-off between linewidth improvement and noise amplification. B0 inhomogeneity could also be incorporated during spectral fitting and this approach could be applied to other multi-band MRS approaches (SENSE, vGRAPPA) and over multiple regions.Acknowledgements
AB would like to acknowledge the support of the Royal Academy of Engineering.References
[1] M. Dehghani, R. A. E. Edden, and J. Near, “Hadamard-encoded
dual-voxel SPECIAL: Short-TE MRS acquired in two brain regions simultaneously
using Hadamard encoding,” Magn Reson Med, vol. 87, no. 4, pp.
1649–1660, Apr. 2022, doi: 10.1002/MRM.29129.
[2] V. O. Boer, D. W. J. Klomp, J. Laterra, and P. B. Barker,
“Parallel reconstruction in accelerated multivoxel MR spectroscopy,” Magn
Reson Med, vol. 74, no. 3, pp. 599–606, 2015, doi: 10.1002/mrm.25718.
[3] L. T. Riemann et al., “Fourier-based decomposition
for simultaneous 2-voxel MRS acquisition with 2SPECIAL,” Magn Reson Med,
vol. 88, no. 5, pp. 1978–1993, Nov. 2022, doi: 10.1002/MRM.29369.
[4] V. O. Boer, M. Andersen, A. Lind, N. G. Lee, A. Marsman,
and E. T. Petersen, “MR spectroscopy using static higher order shimming with
dynamic linear terms (HOS-DLT) for improved water suppression, interleaved
MRS-fMRI, and navigator-based motion correction at 7T,” Magn Reson Med,
vol. 84, no. 3, pp. 1101–1112, 2020, doi: 10.1002/mrm.28202.
[5] A. Bashir and D. A. Yablonskiy, “Natural linewidth chemical
shift imaging (NL-CSI),” Magn Reson Med, vol. 56, no. 1, pp. 7–18,
2006, doi: 10.1002/mrm.20917.
[6] P. Adany, I. Y. Choi, and P. Lee, “B0-adjusted and
sensitivity-encoded spectral localization by imaging (BASE-SLIM) in the human
brain in vivo,” Neuroimage, vol. 134, pp. 355–364, 2016, doi:
10.1016/j.neuroimage.2016.04.016.
[7] A. Cultrera and L. Callegaro, “A simple algorithm to find
the L-curve corner in the regularisation of ill-posed inverse problems,” IOP
SciNotes, vol. 1, no. 2, p. 025004, Aug. 2020, doi:
10.1088/2633-1357/ABAD0D.
[8] W. T. Clarke et al., “NIfTI-MRS: A standard data
format for magnetic resonance spectroscopy,” Magn Reson Med, Dec. 2022,
doi: 10.1002/MRM.29418.
[9] W. T. Clarke, C. J. Stagg, and S. Jbabdi, “FSL‐MRS: An
end‐to‐end spectroscopy analysis package,” Magn Reson Med, vol. 85, no.
6, pp. 2950–2964, Jun. 2021, doi: 10.1002/mrm.28630.