Petra Albertova1,2, Maximilian Gram1,2, Verena Schirmer2, Martin Blaimer3, Martin J. Herrmann4, Matthias Gamer5, Peter Nordbeck1,6, and Peter Michael Jakob2
1Department of Internal Medicine I, University Hospital Würzburg, Würzburg, Germany, 2Experimental Physics 5, University of Würzburg, Würzburg, Germany, 3Fraunhofer Institute for Integrated Circuits IIS, Würzburg, Germany, 4Department of Psychiatry, Psychosomatics, and Psychotherapy, University Hospital Würzburg, Würzburg, Germany, 5Department of Psychology, University of Würzburg, Würzburg, Germany, 6Comprehensive Heart Failure Center (CHFC), University Hospital Würzburg, Würzburg, Germany
Synopsis
Keywords: Bioeffects & Magnetic Fields, Quantitative Imaging, spin-lock
Rotary excitation based MRI, which enables spatially
resolved, non-invasive, direct detection of biomagnetic fields, complements the
measurement range accessible via MEG since detection is independent of the
distance to the body surface. We present an extension of the REX method towards
quantitative imaging. A calibration function can be composed from REX
data of two measurements of adjustable fields projected onto the tissue by the scanner’s
gradient system. In subsequent measurements, the field strength can be inferred
from the calibration model. First
in vivo
proof-of-concept measurements show that the calibration successfully eliminates
influences of tissue properties and accurately quantifies oscillatory nT-fields.
Introduction
In
recent years, spin-lock (SL) based MRI approaches have been introduced which
can potentially provide direct and spatially resolved detection of neuronal
activity [1,2] and thus may overcome critical limitations of Electroencephalography
(EEG), Magnetoencephalography (MEG) as well as blood-oxygenation-level-dependent
(BOLD) functional MRI. So far, the highest sensitivity has been demonstrated
for SL methods that rely on Rotary Excitation (REX) rather than Rotary Saturation
[3]. The REX detection of magnetic oscillations consists of the observation of
a sinusoidal signal variation of the MRI signal SREX and the
determination of the corresponding amplitude AREX. However, the
detected values alone do not provide quantification of magnetic oscillations,
since AREX is tissue-dependent and is influenced by relaxation times
as well as proton density and local field properties.
In
this work, we demonstrate that due to the high linearity of the REX effect,
quantification of oscillating nT-fields is feasible via a linear model
accounting for tissue properties. The calibration applies a method which
utilizes ultra-weak gradient waveforms to transmit the REX effect onto the
tissue under investigation (transmission of Rotary Excitation, tREX) [4]. In
the present study, we validate the concept in a healthy volunteer in brain tissue with magnetic oscillations
in the low nT-range.Methods
All experiments were carried out on a clinical
3.0T scanner (MAGNETOM Skyra, Siemens, Erlangen, Germany) using a 20-channel
head coil. The tREX sequence (Fig. 1) was implemented within the open-source
Pulseq framework [5] and employs a balanced-SL preparation for robust field
detection [4,6] with ramped flip angle acquisitions for fast spiral imaging.
For the reproducible generation of magnetic field oscillations (stimuli), the
built-in gradient system was utilized.
As can be seen in Fig. 2, the REX contrast depends
on the relative phase between the SL-preparation and the stimulus. Since the
signal magnitude of a single image is not conclusive, the amplitude AREX,
which is calculated by the standard deviation over images of different phases,
serves as the indicator for field detection. For AREX determination,
20 phase scans were performed (total measurement duration ≈20s). However, AREX is solely scaled in arbitrary units (a.u.) and
does not allow direct reconstruction of the detected stimulus strength. Therefore,
for AREX quantification, two additional calibration experiments were
conducted using tREX at well-known stimulus field strengths of 7.5nT and 60nT. Since
tissue and field properties critically affect AREX, a site-dependent
calibration function was calculated based on a linear model by determining an AREX
slope along with an offset that accounts for noise in the REX signal. After
calibration, further detections were performed with magnetic stimuli in the
10…50nT range to demonstrate spatially resolved field quantification under in vivo conditions.Results
Fig. 3 illustrates the concept of setting up the
linear calibration function by calculating a REX slope and a corresponding
offset map. The slope of the REX amplitude clearly indicates anatomic
structures. Fig. 4 displays the AREX detection for field strengths
in the range 10…50nT. The amplitude increases with stimulus field strength, however,
AREX is not homogeneous and reveals tissue properties. This is also
reflected by the values averaged over an entire brain tissue ROI. The mean
value increases with field strength. The variation over the ROI also increases and
amounts to on average ±10.7%. In Fig. 5, the detected AREX maps were converted to
quantitative field maps using the calibration model. The field maps are
homogeneous within the subject and deviate from the nominal field strength
values only in the low percentage range. The variation over the brain tissue
ROI amounts to ±4.3% and remains approximately constant at ±0.96nT. The
quantitative field maps reveal no significant site dependence and no apparent influence
of tissue properties.Discussion
With
a linear tREX calibration, the highly sensitive REX method for detection of
magnetic field oscillations can be extended, enabling the quantification of nT-fields.
The calibration function can be determined by only two tREX experiments and was
generated in less than 1min in this case. The calibration places low demands on
the experimental setup, since no additional hardware is required and
measurements can be carried out in vivo.
Due to the limitations of our gradient system, only the range above 2.5nT was
accessible for calibration [4]. Accordingly, REX quantification of
sub-nT-fields can currently only be accomplished by extrapolating the
calibration function. This procedure needs to be further validated in future
studies.Conclusion
Calibration
of the Rotary Excitation effect by means of tREX can be applied to quantify
ultra-weak field oscillations with accuracy better than 1nT. The proposed concept
thus paves the way for spatially resolved and quantitative sensing of neuronal
activity via MRI.Acknowledgements
This work was supported by the Federal Ministry for Education and Research of the Federal Republic of Germany (BMBF 01EO1504, MO6).References
[1]
Truong TK, et al. Toward direct MRI of neuro-electro-magnetic oscillations in
the human brain. Magn Reson Med. 2019 Jun;81(6):3462-3475. doi:
10.1002/mrm.27654
[2]
Capiglioni M, et al. Analysis of the robustness and dynamics of spin-locking
preparations for the detection of oscillatory magnetic fields. Sci Rep. 2022
Oct 10;12(1):16965. doi: 10.1038/s41598-022-21232-1
[3] Jiang X, et al. Detection of subnanotesla oscillatory magnetic fields
using MRI. Magn Reson Med. 2016 Feb;75(2):519-26. doi: 10.1002/mrm.25553
[4]
Gram M, et al. Towards robust in vivo quantification of oscillating biomagnetic
fields using Rotary Excitation based MRI. Sci Rep. 2022 Sep 13;12(1):15375.
doi: 10.1038/s41598-022-19275-5
[5] Layton KJ, et al. Pulseq: A rapid and
hardware-independent pulse sequence prototyping framework. Magn Reson Med. 2017
Apr;77(4):1544-1552. doi: 10.1002/mrm.26235
[6]
Gram M, et al. Balanced spin-lock preparation for B1 -insensitive and B0
-insensitive quantification of the rotating frame relaxation time T1ρ. Magn Reson
Med. 2021 May;85(5):2771-2780. doi: 10.1002/mrm.28585