Aiming Lu1, Guruprasad Krishnamoorthy1,2, Jacinta G Browne1, Scott M Thompson1, Daniel A Adamo1, and David A Woodrum1
1Mayo Clinic, Rochester, MN, United States, 2Philips Healthcare, Rochester, MN, United States
Synopsis
Keywords: Pelvis, Artifacts, intervention
MRI-guided cryoablation
is a widely used minimal invasive treatment that uses extreme cold to destroy
localized diseased tissues. The cryoneedles used are expected to introduce little
metal artifacts by themselves in the MRI images. However, severe artifacts along
the cryoneedle shaft are sometimes observed during MRI guided cryoablations in
practice, which can not only complicate the cryoablation
procedures, but also potentially cause complications and negatively affect treatment
outcome. This work demonstrated
that the artifacts are likely due to coupling of the cryoneedles and the RF coils, and can be minimized by proper positioning of the
cryoneedles, RF coils and patients.
Introduction
MRI-guided cryoablation
is a widely used minimal invasive treatment that uses extreme cold to destroy
localized diseased tissues (1-4). MRI guidance is desirable as it allows for target
tissue delineation, cryoneedle placement guidance, real-time treatment progress
monitoring and treatment assessment. The cryoneedles are made of non-magnetic
nickel-chromium-based superalloys known as Inconel and are expected to introduce little
metal artifacts by themselves in the MRI images. However, severe artifacts along
the cryoneedle shaft are sometimes present during MRI guided cryoablations in
practice, which can obstruct the visualization of the cryoneedle placement and the
iceball boundary as these artifacts can also appear as a signal void or a low
signal region. As a result, the artifacts can not only complicate the cryoablation
procedures, but also potentially cause complications and negatively affect treatment
outcome. The purpose of this work is to investigate the cause of these
artifacts and potential strategies to minimize them. Materials and methods
MRI guided cryoablations were performed on a 1.5T Ingenia scanner (Philips
Healthcare, Netherlands) using the Visual-ICE MRI system (Boston Scientific, Marlborough, MA) and cryoneedles (IceSeed/IceRod MRI, 17 Gauge). IRB approval was
obtained for patient studies. Ex-vivo cryoablation studies were performed using porcine tissue samples.
A dStream Flex M loop coil and the posterior
coil integrated in the patient table were used for data acquisition in both patient
and ex-vivo tissue sample studies. Figure 1 shows an example of cryoneedle
artifacts during a cryoablation procedure. The artifacts obscured the visualization
of the cryoneedle and also made it difficult to monitor the iceball evolution due
to their similar appearance in the images. These artifacts likely come from the
coupling between the cryoneedles and the RF coils when they are placed close to
each other since the cryoneedles are metallic. To investigate the dependence of the cryoneedle
artifacts on the relative positioning of the cryoneedle to the coils, tissue
sample experiments were performed using different pulse sequences for imaging. A
cryoneedle was inserted horizontally ~10cm into the tissue placed on the
patient table. Unless otherwise stated, the cryoneedle was disconnected from
the mobile connection panel (MCP). MR imaging was performed with the following
coil/cryoneedle configurations: 1. The tissue sample was placed on the patient
table with a layer of procedure underpad. An IceRod cryoneedle was inserted
into the tissue and ~3cm from the patient table. The loop coil was placed on
top of the tissue sample and ~3cm from the cryoneedle. 2.Same as configuration
1 except the IceRod was replaced with a IceSeed. 3. Same as configuration 2
except the loop coil was removed and only the posterior coil was used for data
acquisition. 4. Same as configuration 2 except with the cryoneedle connected to
the MCP. 5. Same as configure 2 except the separation between the tissue sample
and the patient table was increased by 4 cm using a pad. 6. Same as configuration
5 but with the cryoneedle connected to the MCP. The experiments were repeated
with different cryoneedle insertion locations into the tissue sample. MR
imaging was performed using several typically used sequences, including a T2-weightedTSE sequence (TE/TR: 70ms/971ms, ETL: 22), a proton density-weighted (PD-W) TSE
sequence (TE/TR: 25ms/3800ms, ETL: 22). To
reduce metal artifacts, a 3D UTE with a TE of 0.14 ms was also used.Results and discussion
With
the tissue placed closed to the patient table and consequently
the integrated posterior coil, severe artifacts are seen in all MR images
acquired with different pulse sequences (Figure 2a-c). Changing from an IceRod
cryoneedle to a IceSeed cryoneedle, connecting/disconnecting the cryoneedle from
the MCP showed insignificant impact on the artifacts. Removing the loop coil did
not have much impact on the artifacts either, which means the loop coil was not
the source of the artifacts in this case. When the tissue sample was moved away
from the posterior coil, however, the cryoneedle artifacts were minimized and
the cryoneedle could be clearly visualized as negative contrast (Figure 2d). This
confirmed that the coupling between cryoneedle and the posterior coil was the source
of the artifacts in these experiments.
Although
the artifacts shown in the ex vivo experiments here were from coupling between
the cryoneedle and the posterior coil, cryoneedle artifacts due to coupling
between the cryoneedles and other coils (such as the loop coil) have also occurred
(e.g., Figure 1). The severity of the artifacts depends highly on the relative positions
of the cryoneedles and the RF coils. The artifacts could be minimized by changing
the RF coil position relative to the cryoneedles, changing the patient positioning,
or choosing a different cryoneedle insertion path. Although moving the coils
away from cryoneedle helps with reducing the artifacts, it may also negatively
affect the signal-to-noise ratio in the region of interest. Therefore, patient
positioning, cryoneedle and coil selection and placement should be considered to
achieve optimized procedure outcome. Conclusion
This work demonstrated
that although the cryoneedles are made of non-magnetic superalloys expected to
introduce little metal artifacts in MRI images, severe artifacts do occur sometimes
during MRI guided cryoablations due to coupling of the cryoneedles to the RF coils.
These artifacts can be minimized by careful positioning of the
cryoneedles, RF coils and patients.Acknowledgements
No acknowledgement found.References
1. Morrison,
P.R., Silverman, S.G., Tuncali, K. and Tatli, S. (2008), MRI-guided
cryotherapy. J. Magn. Reson. Imaging, 27: 410-420. https://doi.org/10.1002/jmri.21260
2. Baust, J.G. and Gage, A.A. (2005), The
molecular basis of cryosurgery. BJU International, 95: 1187-1191. https://doi.org/10.1111/j.1464-410X.2005.05502.x
3. Gage
AA, Baust J. Mechanisms of tissue injury in cryosurgery. Cryobiology.
1998;37(3):171-186. doi:10.1006/cryo.1998.2115
4. Woodrum
DA, Kawashima A, Karnes RJ, et al. Magnetic resonance imaging-guided
cryoablation of recurrent prostate cancer after radical prostatectomy: initial
single institution experience. Urology. 2013;82(4):870-875. doi:
10.1016/j.urology.2013.06.011