0668

Accurate B1 Mapping with Actual Flip Angle Imaging (AFI) in the Presence of Fat
Alexey Samsonov1, Julia Velikina2, and Vasily Yarnykh3
1Radiology, University of Wisconsin, Middleton, WI, United States, 2Radiology, University of Wisconsin, Madison, WI, United States, 3University of Washington, Seattle, WA, United States

Synopsis

Keywords: Quantitative Imaging, Artifacts, B1 mapping; quantitative; parameter mapping

Actual Flip angle Imaging (AFI) is an efficient B1 mapping method requiring the proper spoiling of transverse magnetization. Optimal spoiling can be achieved using large spoiling gradients enabling water diffusion-based signal decay. However, spoiling the non-aqueous signal like from fat is typically ignored in AFI optimizations. We demonstrate that infinitesimal diffusion in the fat signal makes fat spoiling in AFI unachievable in the reasonable scan time and that incomplete fat spoiling is a major source of previously unexplained AFI errors. We propose method to minimize them using a superposition model of the spoiling artifacts and chemical shift encoded fat/water separation.

Introduction

Actual Flip angle Imaging (AFI) is an efficient $$$B1$$$ mapping method, which requires proper spoiling of transverse magnetization [1]. Optimal spoiling can be achieved in the presence of large spoiling gradients enabling water diffusion-based signal decay [2]. This approach provides high $$$B1$$$ mapping accuracy in tissues of diagnostic interest due to sufficiently high diffusion coefficients in them [2]. Simultaneously, spoiling the non-aqueous signal, i.e. from fat, is typically ignored in AFI optimizations, due to fat not being a primary imaging target in many standard quantitative MRI approaches, especially in neuroimaging. Here, we argue that infinitesimal diffusion in the fat makes fat spoiling in AFI unachievable in a reasonable scan time. We demonstrate that incomplete fat spoiling is a major source of previously unexplained AFI errors affecting not only fatty but also other tissues. Finally, we propose a method to minimize these errors using a superposition model of the spoiling artifacts and chemical shift encoded (CSE) fat/water separation.

Theory

AFI sequence collects the signals $$$S_1$$$ and $$$S_2$$$ from two alternating spoiled gradient echo (SPGR) intervals with repetition times $$$\rm{TR}_1$$$ and $$$\rm{TR}_2$$$ ($$$\rm{TR}_2/\rm{TR}_1=n,n\approx4-5$$$ [2]), with the areas of spoilers $$$G_1,G_2$$$ at the end of each $$$TR$$$ related as $$$A_{G_2}=n\cdot{A_{G_1}}$$$. The effect of diffusion on the spoiling can be estimated based on the published data using diffusion dampening approach [3]. The dampening coefficient $$$d=D\,b$$$ depends linearly on the diffusion coefficient $$$D$$$ and quadratically on area of the gradient spoilers: $$b=\left(\gamma\int_0^{\rm TR_1}{\bf G}(\tau)d\tau\right)^2\cdot\rm{TR}_1.$$ Following results of [3], the tissue-dependence of optimal RF spoiling can be ignored if $$$d\ge1$$$. For example, for a typical range of tissue water diffusion coefficients, assuming gradient strength 28 mT/m, this condition leads to attainable 16ms $$$G_1$$$ duration [3]. However, the diffusion of fatty acids and triglycerides is about two orders of magnitude smaller than that of water [4], requiring the duration to increase by an order of magnitude, thus making it infeasible to implement without violating the scan time limits and AFI assumption ($$$\rm{TR}_1,\rm{TR}_2\ll T1$$$). Therefore, the operation within the feasible spoiling regimen would result in a situation where the fat signal is partially spoiled, and the associated breakdown of steady-state condition may lead not only to localized but also to distributed errors, typically appearing as ghost signals [5]. Following the linearity of the signal formation, we note that the fat ghosts retain the distinct chemical shift signature of fat. Therefore, we hypothesize that they can be removed from source images prior to AFI processing by separating AFI images into fat and water channels using CSE F/W separation methods provided that multiple echoes are collected in AFI.

Methods

Experiments were performed on a 3.0T GE Discovery MR750 (Waukesha, WI). AFI was implemented with four-echo readout [6]. AFI spoiling was implemented using RF phase increments and gradient spoilers identified, according to [2], as standard, strong, and complete spoiling regimes for water-containing tissues. The phantom experiments were performed in objects comprising a plastic sphere either empty or filled with Gd-dopped water and two peanut oil containers. VFA $$$R1$$$ mapping was performed in MS subjects to study the effect of different $$$B1$$$ mapping approaches on $$$R1$$$ quantification ($$$\rm{FA}=4/24^o,\rm{TR}=20\,\rm{ms}$$$). To exclude contributions from other potential mechanisms, modified imaging experiments (results not shown here) were also performed to control for other possible sources of image artifacts including vibrational motion (by decoupling the phantom from the table/coil through external support) and Eddy currents (by derating the spoiling gradients). Multi-echo AFI was used either in standard (the first echo processing) or in the proposed CSE regime (separating AFI signals $$$S_1/S_2$$$ into F/W followed by AFI on each channel). The fat fraction map from the separation was used to guide the backward combination of F/W AFI maps into a single map for demonstration purposes (FF>0.9 was considered to contain fat-only tissue, otherwise water). Additionally, we explored standard approach for improving AFI maps quality by spatial smoothing.

Results

Figure 1 illustrates the proposed processing pipeline and correction of fat spoiling artifact in phantom. Results in Figure 2 further confirm that fat is the source of the $$$B1$$$ mapping artifacts observed in Fig. 1. Figure 3 compares the method’s performance in thighs and demonstrate removal of $$$B1$$$ mapping errors with the proposed CSE-AFI. Figure 4 demonstrates superior performance of CSE-AFI in brain in a variety of spoiling regimes. Figure 5 shows results of evaluation of different AFI methods for correction of excitation FA in VFA $$$R1$$$ mapping demonstrating that $$$B1$$$ with proposed CSE-AFI provides more complete correction of $$$R1$$$ value from flip angle variations.

Discussion and Conclusions

This work identifies incomplete spoiling of fat signal as a major source of AFI $$$B1$$$ mapping errors. The standard approach to their correction necessitates infeasible requirements to gradient strength and/or scan time. Instead, we propose to use chemical-shift signature of the fat in conjunction with multi-echo AFI acquisition to eliminate the artifacts, thereby enabling $$$B1$$$ mapping with minimized errors even in standard spoiling regimen. While this method was demonstrated to improve significantly $$$B1$$$ mapping in locations with relatively limited amount of fat (e.g., in intracranial imaging), we anticipate the method to be especially useful in body imaging applications, where fat is more abundant.

Acknowledgements

The work was supported by NIH (R01EB027087) and GE Healthcare.

References

1. Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med 2007;57(1):192-200.

2. Yarnykh VL. Optimal radiofrequency and gradient spoiling for improved accuracy of T1 and B1 measurements using fast steady-state techniques. Magn Reson Med 2010;63(6):1610-1626.

3. Nehrke K. On the steady-state properties of actual flip angle imaging (AFI). Magn Reson Med 2009;61(1):84-92.

4. Steidle G, Eibofner F, Schick F. Quantitative diffusion imaging of adipose tissue in the human lower leg at 1.5 T. Magn Reson Med 2011;65(4):1118-1124.

5. Leupold J, Hennig J, Scheffler K. Moment and direction of the spoiler gradient for effective artifact suppression in RF-spoiled gradient echo imaging. Magn Reson Med 2008;60(1):119-127.

6. Yu H, Shimakawa A, McKenzie CA, Brodsky E, Brittain JH, Reeder SB. Multiecho water-fat separation and simultaneous R2* estimation with multifrequency fat spectrum modeling. Magn Reson Med 2008;60(5):1122-1134.

Figures

Figure 1. Correction of fat spoiling artifact in phantom. Standard AFI B1 map from 1st echo shows visible ghosting in spherical water phantom, consistent with the pattern of signal modulation in poorly spoiled signal in fat (small vial on top and large bottle on the bottom). Preprocessing AFI signals with F/W separation removes the ghosting in corresponding channels. CSE-AFI map after combination of F/W AFI maps shows removal of the error from the water object.

Figure 2. Further investigation on the source of ghosting. The sphere in the previous experiment (Fig. 1) was emptied, and the experiment was repeated in the same setup. Scaling the image intensity to emphasize background (10x) reveals that there are ghosts from the fat, which are considerably lower for the first TR. This is consistent with expected improved spoiling of fat signal in the short TR due to more efficient signal decay in the preceding long TR. The difference allows further appreciating the artifacts (with reduced Gibbs ringing).

Figure 3. Correction of B1 mapping errors in thighs. (a) The abundance of fat causes significant artifacts in standard AFI B1 map from 1st echo (arrows). They are which are eliminated in water-only map and (b) in CSE-based AFI B1 maps obtained by fat-fraction based combination of AFI maps from fat and water channels (b).

Figure 4. Performance of the methods in brain. Subcutaneous fat causes ghosting in standard AFI B1 maps obtained with different spoiling regimes (Sp1 – weak, Sp2 – strong, Sp3 – complete, defined in [2]). The fat ghosting is removed with CSE-AFI for all spoiling regimes. The results allow appreciating other errors due to incomplete spoiling of other tissues in Sp1/Sp2 regimes. B1 map acquired in the complete water spoiling regime (Sp3) and processed with the proposed CSE-based procedure demonstrates the best regularity.

Figure 5. Effect of several AFI B1 mapping methods on R1 mapping. (a) Propagation of B1 errors from standard AFI results in artifacts (arrows), all diminished in the proposed approach. The effect can be appreciated on the histograms (b), which are blurred for the standard AFI and sharpened for the proposed method, indicating improved correction of B1-induced inhomogeneity. Postprocessing AFI maps by spatial smoothing improves histograms for standard AFI, which still is not sufficient to reach fidelity of the similarly processed CSE-AFI (c).

Proc. Intl. Soc. Mag. Reson. Med. 31 (2023)
0668
DOI: https://doi.org/10.58530/2023/0668