Iman Khodarahmi1, Mahesh B Keerthivasan2, and Jan Fritz1
1NYU Langone School of Medicine, New York, NY, United States, 2Siemens Medical Solutions USA Inc., Malvern, PA, United States
Synopsis
Keywords: Bone, Bone
$$$B_1^+$$$ field inhomogeneity is
a source of metal artifacts in patients with orthopedic hardware. Local $$$B_1^+$$$ shimming can
potentially decrease these artifacts and improve visualization of the
bone-metal interface. Our proposed turbo-spin echo-based $$$B_1^+$$$ mapping technique
enables accurate estimation of the $$$B_1^+$$$ field near the metal hardware. After optimization for in-vivo applications, the
technique was successfully employed on a clinical 3.0 T parallel-transmit
system aiming at $$$B_1^+$$$ shimming near the orthopedic hardware. Our results demonstrate
significant improvement in visualization of the bone-metal interface compared
to standard 1.5 and 3.0 T acquisitions.
Introduction
Susceptibility
artifacts are a well-known source of image artifacts in MRI of patients with
metallic orthopedic hardware. The other much less investigated source of
artifacts is the perturbation in the transmit $$$B_1^+$$$ field, which is caused by metal-induced electromagnetic
fields 1-2. Such $$$B_1^+$$$ inhomogeneities may cause signal alterations at the bone-metal interface and be
misinterpreted as an abnormality and obscure the underlying pathology.
$$$B_1^+$$$ shimming localized to
the area surrounding the metal hardware using a multi-channel transmit system can
potentially provide a more homogeneous $$$B_1^+$$$ field and improve visualization of the bone-metal interface. A successful $$$B_1^+$$$ shim requires (1)
obtaining reliable $$$B_1^+$$$ field maps in the presence of metal and (2) a subject-specific combination of
the individual $$$B_1^+$$$ field maps from each transmit channel.
This work aims to
develop a turbo-spin echo (TSE)-based $$$B_1^+$$$ mapping technique for in-vivo applications and analyze the effect of local $$$B_1^+$$$ shimming on the image
quality of patients with metallic orthopedic hardware using a clinical dual-transmit
system.Methods
$$$\it B_1^+$$$
mapping: We have recently developed and validated a
TSE-based method of $$$B_1^+$$$ mapping which relies on the acquisition of multiple images with various
excitation-refocusing (EX-Ref) flip angles (FA) 3. Briefly, with a $$$B_1^+$$$ amplitude scale factor
of b1 (b1
= actual $$$B_1^+$$$ / nominal $$$B_1^+$$$),
the signal intensity of a TSE sequence can be expressed as: $$$S(b_{1}) = f(b_{1}.\theta,b_{1}.\phi,\psi)$$$, where θ and φ represent excitation and
refocusing flip angles, respectively, ψ other imaging parameters, and f(.) is the signal Bloch model. For
different sets of Ex-Ref FA, b1 can be obtained by solving the following optimization problem: $$\widehat{b_{1}} = \min_{b_1}\sum_i^n\parallel f(b_{1}.\theta_{i},b_{1}.\phi_{i},\psi)-\widehat{S}(\theta_{i},\phi_{i},\psi) \parallel_{2}$$ with $$$\widehat{S}(\theta_{i},\phi_{i},\psi)$$$ being the pixel signal obtained
by the ith set of Ex-Ref FA.
$$$\it B_1^+$$$ mapping optimization for in-vivo
imaging: Monte-Carlo simulations were performed to study the sensitivity
of $$$B_1^+$$$ relative error to the number and
choice of Ex-Ref FA pairs. The TSE signal was simulated for a wide range of
excitation (30-120°) and refocusing (60-180°) FA sets using Bloch equations. Gaussian noise was added to the
simulated signal for each set and the mean $$$B_1^+$$$ map was estimated from one thousand
realizations.
$$$\it B_1^+$$$ shimming: The composite $$$B_1^+$$$ field can be obtained
by complex linear combination of the $$$B_1^+$$$ field maps of the two transmit channels using the $$$B_1^+$$$ amplitudes of the previous step and the
relative $$$B_1^+$$$ phase distribution of the TSE images. For a given region-of-interest (ROI) near
the metal, the shimming was performed by adjusting the relative amplitude, and
phases of the two transmit elements, aiming at minimizing the difference
between the $$$B_1^+$$$ in the ROI and that of the background.
MRI
Experiments: After
obtaining institutional review board approval and informed consent, a volunteer
with a femoral intramedullary nail was imaged on a parallel-transmit 3T
clinical system with the following parameters: TR/TE = 2000/29 ms, voxel size =
0.6 x 0.6 x 3.0 mm3, and turbo factor = 13. Standard and $$$B_1^+$$$ shimmed intermediate-weighted axial images were acquired, focusing on
visualization of the bone-metal interface. For comparison purposes, $$$B_1^+$$$ maps were also obtained using a TurboFLASH (TFL) sequence equipped with a
preceding RF pulse for magnetization preparation 4.Results and Discussion
Monte-Carlo
simulation results on the optimal number and choice of the Ex-Ref FA pairs are summarized in
Figure 1. The lowest mean relative error in $$$B_1^+$$$ estimations were 27%, 19%, and 18% in 2,3,
and 4 Ex-Ref FA sets, respectively. Considering
the increased imaging time and specific absorption rate (SAR) values, these
results suggest that acquiring more than three Ex-Ref FA sets will not result
in improved $$$B_1^+$$$ estimation. In addition, the error in $$$B_1^+$$$ estimation is more
pronounced at low $$$B_1^+$$$ values, as the overall low signal at low $$$B_1^+$$$ values reaches the noise floor.
Using the optimal
three Ex-Ref FA sets, the $$$B_1^+$$$ field was estimated in
a volunteer and compared with those of the TFL method (Figure 2). As seen on
magnified subplots, the $$$B_1^+$$$ field near the metal is better resolved with our technique.
The effect of $$$B_1^+$$$ shimming on the
visualization of the bone-metal interface is shown in Figure 3. $$$B_1^+$$$ shimming in a
bone-metal containing ROI provides the optimal shim parameters, which were used
for subsequent imaging. Compared with the standard 1.5 and 3.0 T images, the visibility
of the bone-metal interface is significantly improved after $$$B_1^+$$$ shimming.Conclusion
The proposed $$$B_1^+$$$ mapping technique
provides promising results at the bone-metal interface invisible to other
mapping sequences. Our initial results suggest that patient-specific local $$$B_1^+$$$ shimming is clinically
feasible and can reduce $$$B_1^+$$$-related
artifacts surrounding the metal.Acknowledgements
No acknowledgement found.References
- Khodarahmi I, Nittka M,
Fritz J. Leaps in Technology:
Advanced MR Imaging after Total Hip Arthroplasty. Semin Musculoskelet Radiol.
2017;21(5):604-15.
- Bachschmidt TJ, Kohler M,
Nistler J, Geppert C, Jakob PM, Nittka M. Polarized Multichannel Transmit MRI
to Reduce Shading near Metal Implants. Magn Reson Med. 2016;75(1):217–226.
- Khodarahmi I, Keerthivasan
MB, Fritz J. Turbo-spin echo based B1+ mapping in the presence of metallic
hardware. 31st Annual Meeting of the International Society for Magnetic
Resonance in Medicine: International Society for Magnetic Resonance in
Medicine; 2022.
- Chung S, Kim D, Breton E,
and Axel L. Rapid B1+ mapping using a preconditioning RF pulse with turboflash
readout. Magn Reson Med. 2010;64(2):439-446.