Oliver Kraff1, Jenni Schulz2, Markus W May1,3, Tom WJ Scheenen1,2, Thomas B Meurs4, Sebastian Schmitter5, Jens M Theysohn6, and Harald H Quick1,3
1Erwin L Hahn Institute for MRI, University Duisburg-Essen, Essen, Germany, 2Medical Imaging, Radboud UMC, Nijmegen, Netherlands, 3High-Field and Hybrid MR Imaging, University Hospital Essen, Essen, Germany, 4Tesla Dynamic Coils, Zaltbommel, Netherlands, 5Physikalisch-Technische Bundesanstalt (PTB), Braunschweig und Berlin, Germany, 6Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany
Synopsis
Keywords: Joints, RF Pulse Design & Fields, UHF
First experiences with direct signal control
with variable excitation and refocusing (DiSCoVER) for T2-weighted TSE imaging
in 7 Tesla musculoskeletal (MSK) applications of hip, shoulder, and ankle are
presented. Images were compared to static RF shimming aiming at a homogeneous
excitation by using the MR system’s framework as well as optimizing for a B
1+-efficient
shim in an offline calculation. DiSCoVER yielded satisfying results within the
ROI defined for signal optimization but showed more pronounced signal dropouts
outside the ROI compared to static RF shimming. Workflow improvements were
particularly noted for DiSCoVER as it calculates pTx scale factors under SAR
constraints.
Introduction
Turbo-spin
echo (TSE) sequences are a challenge at ultra-high fields (UHF) as they rely on
large flip angles (FA), and hence are not only prone to B1+
inhomogeneities, but also suffer from local specific absorption rate
limitations. Body and musculoskeletal (MSK) imaging applications at 7T are
known to be heavily impaired by radiofrequency (RF) inhomogeneities and enhanced
tissue absorption, leading to strong FA variations and limited penetration
depth, especially in large cross-sections.1 Recently, direct signal control with variable
excitation and refocusing (DiSCoVER) has demonstrated improvements in image
quality compared to static RF shimming for TSE brain imaging on 7T parallel
transmission (pTx) MR systems.2-4 In this study, we report on first experiences
in using the DiSCoVER technique outside of the brain for MSK imaging cases of
the hip, ankle, and shoulder at 7T.Methods
Imaging was performed on a 7T pTx system (Magnetom
Terra, Siemens Healthcare, Germany) using an 8-channel body array
(Tesla Dynamic Coils, The Netherlands) for hip imaging (Fig. 1A) and an
8-channel coil with an additional 7-channel receive array for ankle and
shoulder imaging (Fig. 1B).5 For hip imaging one male,
and for ankle/shoulder imaging two female subjects were included after
obtaining informed consent. Specifications of the T2-weighted TSE sequence were
as follows: TR=4500ms, TE=30ms, nominal FA=180°, TF=5, BW=200Hz/px.
Spatial resolution/acquisition time varied between 0.4x0.4x3mm3 /
4:30min (ankle/shoulder) and 1x1x3mm3 / 2:40min (hip), respectively.
For ankle imaging an additional set of T2-w TSE was acquired with SPAIR fat
saturation. After acquiring images with the DiSCoVER optimization method two
more datasets were generated with static RF shimming for comparison. First, the
MR system’s integrated RF shimming algorithm was used which adjusts both
amplitudes and phases of the RF coil automatically for a homogeneous excitation
within a region of interest (ROI). Second, a phase-only optimization for a B1+-efficient
excitation within the ROI was performed offline in MATLAB R2020a (MathWorks, MA) based on exported relative B1+ maps.6,7 The phase settings with equal
amplitudes were then manually updated on the scanner console. ROIs were placed
over the joints to be imaged and kept identical within one examination. Hip
imaging started with unilateral optimization of the right joint only and followed
by the optimization for both joints. Images were evaluated qualitatively
regarding artifacts and homogeneity over the full cross section. For a
quantitative evaluation contrast ratios (CR) were calculated between cartilage
and bone, as well as between bone and Achilles tendon in the ankle.Results
Unilateral imaging over the right hip joint
yielded comparable results for all acquisitions with a good excitation over the
ROI but a strong signal dropout at the contralateral joint (Fig. 2). An
inhomogeneity at the gluteal muscle adjacent to the right femoral head was more
pronounced with DiSCoVER compared to static RF shimming. On the other hand,
DiSCoVER showed a 10% higher CR between femoral cartilage/bone and allowed
18 slices to be acquired within SAR guidelines compared to static RF shimming
with 15 (homogeneous shim) and 14 slices (B1+-efficient
shim), respectively.
Bilateral hip imaging (Fig. 3) showed a similar
excitation at the right joint but inferior quality at the left joint for all
three scans. Notably, static RF shimming outperformed DiSCoVER at the left
joint due to better delineation of the posterior part of the femoral cartilage.
In addition, the central signal dropout is much less pronounced. CR measured
for both hip joints was 10% better for DiSCoVER and B1+-efficient
shim compared the homogenous shim. Time-averaged RF power varied between 19 and
22W for the hip exams.
For imaging smaller joints homogeneous image quality without signal
dropouts was obtained in the ankle with all three techniques (Fig. 4). However, some blurring was visible in the calcaneus for the
B1+-efficient shim. CR between cartilage/bone
was 12-16% better for DiSCoVER and the B1+-efficient shim
compared to the homogeneous shim. A similar tendency was observed for CR
between bone/tendon as well for the acquisitions with SPAIR fat saturation.
Here, the SPAIR images obtained with DiSCoVER appeared sharper and less noisy
compared to static RF shimming. Time-averaged RF power was 3.4W and yielded no
limitations in spatial coverage.
MR imaging of the shoulder joint (Fig. 5) yielded again
stronger signal loss adjacent to the joint for DiSCoVER, both ventrally in the
deltoid muscle and dorsal in the infraspinatus muscle and scapula. CR between
cartilage/bone as well as between cartilage/fluid were similar for
acquisitions with DiSCoVER and static RF shimming. Time-averaged RF
power was 11.7W and yielded no limitations in spatial coverage.Discussion and Conclusion
Overall, the DiSCoVER signal
optimization technique for TSE imaging is very well integrated in the MR system
and yields satisfying image quality in first applications outside the brain. Workflow
improvements were particularly noted for hip imaging as DiSCoVER calculates pTx
scale factors for excitation and refocusing pulses under SAR constraints. For
static RF shimming, the MR system’s lookahead SAR prediction demanded a notable
decrease in spatial coverage. Limitations with DiSCoVER were observed in
regions adjacent to the ROI. Future work will focus on protocol optimization w.r.t.
chemical shift artefacts and other TSE-variants like PD fat saturation and STIR,
as well as on reproducibility in more subjects with varying body mass index.Acknowledgements
The body
array coil used in this work was constructed with support from the European
Fund for Regional Development EFRO OP-2014-2023-Oost [PROJ-01009].
Turbo spin-echo imaging with DiSCoVER signal homogenization was kindly provided by Siemens Healthineers as a work-in-progress sequence.
References
1. Kraff O, Quick HH. 7T: Physics,
safety, and potential clinical applications. JMRI 2017;46(6):1573-1589.
2. Beqiri A, et al. Whole-brain 3D FLAIR at 7T using direct
signal control. MRM 2018;80(4):1533-1545.
3. Malik SJ, et al. Direct signal control of the steady-state response of
3D-FSE sequences. MRM 2015;73(3):951-963.
4. Tomi-Tricot R, et al.
Fully integrated scanner implamentation of direct signal control for 2D
T2-weighted TSE at ultra-high fields. Proc. 29th Annual
Meeting ISMRM 2021; 621.
5. Rietsch SHG,
et al. An 8-channel transceiver 7-channel receive RF
coil setup for high SNR ultrahigh-field MRI of the shoulder at 7T. Med Phys 2017;44(12):6195-6208.
6. Van de Moortele P, et al.
Very fast multi channel B1 calibration at high field in the small flip angle
regime. Proc. 17th Annual Meeting ISMRM 2009; 367.
7. Schmitter S, et al. Cerebral TOF angiography at 7T: Impact of
B1+ shimming with a 16-channel transceiver array. MRM 2014;71(3):966-977.